Electrochemical sensor based on multi-walled carbon nanotubes for imidacloprid determination

被引:20
|
作者
Bruzaca, Evellin E. S. [1 ]
de Oliveira, Raissa C. [1 ]
Duarte, Mateus S. S. [1 ]
Sousa, Camila P. [1 ]
Morais, Simone [2 ]
Correia, Adriana N. [1 ]
de Lima-Neto, Pedro [1 ]
机构
[1] Univ Fed Ceara, Ctr Ciencias, Dept Quim Analit & Fis Quim, Campus Pici, BR-60440900 Fortaleza, CE, Brazil
[2] Inst Politecn Porto, Inst Super Engn Porto, REQUIMTE LAQV, R Dr Antonio Bernardino de Almeida 431, P-4200072 Porto, Portugal
关键词
SQUARE-WAVE VOLTAMMETRY; GRAPHENE OXIDE; NEONICOTINOID INSECTICIDES; LIQUID-CHROMATOGRAPHY; PERFORMANCE; ELECTRODE; WATERS; EXPOSURE; POLYMER; SHRIMP;
D O I
10.1039/d1ay00198a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and robust sensor (fMWCNT-Nafion (R)(0.5%)/GCE) for determination of imidacloprid (IMI), a widely used neonicotinoid, was developed using a glassy carbon electrode (GCE) modified with functionalized multi-walled carbon nanotubes (fMWCNT) and Nafion (R). The obtained data suggest that IMI reduction is an irreversible process, due to the reduction of the nitro group to hydroxylamine derivatives, with the participation of two protons and four electrons, and a charge transfer coefficient of 0.141. The optimized square-wave voltammetric conditions were: McIlvaine buffer at pH 6.0, 0.5% of Nafion (R) in the fMWCNT suspension, -0.6 V and 180 s as accumulation potential and time, respectively. A linearity in the range of 2.00 x 10(-7) to 1.77 x 10(-6) mol L-1 IMI, with the values of limit of detection and limit of quantification were equal to 3.74 x 10(-8) mol L-1 and 1.25 x 10(-7) mol L-1, respectively. Repeatability and reproducibility displayed relative standard deviations lower than 5%. Recovery tests performed in tap water, melon, and shrimp yielded mean values of 94 +/- 6%, 97 +/- 10% and 93 +/- 10%, respectively. Moreover, several inorganic and organic compounds did not significantly interfere (0.6 to 4.5%) on the IMI signal, proving the selectivity and applicability of the developed sensor for IMI detection in complex samples.
引用
收藏
页码:2124 / 2136
页数:13
相关论文
共 50 条
  • [1] Fabrication of an electrochemical sensor based on magnetic multi-walled carbon nanotubes for the determination of ciprofloxacin
    Bagheri, H.
    Khoshsafar, H.
    Amidi, S.
    Ardakani, Y. Hosseinzadeh
    [J]. ANALYTICAL METHODS, 2016, 8 (16) : 3383 - 3390
  • [2] Electrochemical Sensor Based on Multi-Walled Carbon Nanotubes/Gold Nanoparticles Composite for Determination of Melamine
    Qu, Jianying
    Du, Xueping
    Kang, Shiping
    Lou, Tongfang
    [J]. ASIAN JOURNAL OF CHEMISTRY, 2013, 25 (18) : 10147 - 10149
  • [3] Electrochemical Sensor Based on Multi-walled Carbon Nanotubes and Cobalt Phthalocyanine Composite for Pyridoxine Determination
    Porto, Lais Sales
    da Silva, Daniela Nunes
    Silva, Maria Cristina
    Pereira, Arnaldo Cesar
    [J]. ELECTROANALYSIS, 2019, 31 (05) : 820 - 828
  • [4] Electrochemical chlorine sensor with multi-walled carbon nanotubes as electrocatalysts
    Sun, Gengzhi
    Liu, Shiwei
    Hua, Kaifeng
    Lv, Xiangyu
    Huang, Li
    Wang, Yujiang
    [J]. ELECTROCHEMISTRY COMMUNICATIONS, 2007, 9 (09) : 2436 - 2440
  • [5] Construction and Properties of the Electrochemical Bisphenol A Sensor Based on Polyoxometalates and Multi-walled Carbon Nanotubes
    Li Na
    Zhang Cong
    Ren Ju-Jie
    Cui Min
    Zhao Hai-Yan
    Zhang Hong-Yue
    Chen Shi-Xu
    Han Hong-Yan
    Zhang Huan
    [J]. CHINESE JOURNAL OF INORGANIC CHEMISTRY, 2022, 38 (08) : 1633 - 1642
  • [6] Novel bismuth/multi-walled carbon nanotubes-based electrochemical sensor for the determination of neuroprotective drug cilostazol
    Rajeev Jain
    Ramkishor Sharma
    [J]. Journal of Applied Electrochemistry, 2012, 42 : 341 - 348
  • [7] Molecularly imprinted electrochemical sensor based on multi-walled carbon nanotubes for specific recognition and determination of chloramphenicol in milk
    Geng, Lingjun
    Huang, Jingcheng
    Zhai, Hongguo
    Shen, Zheng
    Han, Jie
    Yu, Yanyang
    Fang, Honggang
    Li, Falan
    Sun, Xia
    Guo, Yemin
    [J]. MICROCHEMICAL JOURNAL, 2022, 182
  • [8] Sensitive and selective molecularly imprinted electrochemical sensor based on multi-walled carbon nanotubes for doxycycline hyclate determination
    Zeyu Xu
    Xingxing Jiang
    Shuping Liu
    Minghui Yang
    [J]. Chinese Chemical Letters, 2020, 31 (01) : 185 - 188
  • [9] A novel electrochemical sensor based on Ag nanoparticles decorated multi-walled carbon nanotubes for applied determination of nitrite
    Wan, Yue
    Zheng, Yi Fan
    Wan, Hua Tao
    Yin, Hao Yong
    Song, Xu Chun
    [J]. FOOD CONTROL, 2017, 73 : 1507 - 1513
  • [10] A simple and efficient electrochemical sensor for determination of gallic acid based on multi-walled carbon nanotubes with carboxyl functionalization
    Zhu G.
    Hu N.
    Guo M.
    Liu Y.
    Ran Q.
    Zhao H.
    [J]. Materials Research Innovations, 2022, 26 (06): : 389 - 396