On-line column preconcentration for the determination of cobalt in sea water by flow-injection chemiluminescence detection

被引:0
|
作者
Hirata, S
Aihara, M
Hashimoto, Y
Mallika, GV
机构
[1] KINKI UNIV,HIGASHI KU,HIROSHIMA 72919,JAPAN
[2] CEYLON INST SCI & IND RES,COLOMBO 7,SRI LANKA
来源
关键词
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow-injection method for the determination of dissolved cobalt(II) in sea water has been studied based on a combination of column preconcentration using 8-quinolinol immobilized on silica gel, fluoride containing metal alkoxide glass (8HQ-MAF) and chemiluminescence detection with a gallic acid-hydrogen peroxide system. Co(II) is selectively recovered from an acidified sample with 8-quinolinol immobilized on silica gel. After elution with dilute hydrochloric acid the resultant eluent is mixed with the reagent solutions, heated to 60 degrees C and then introduced into the CL cell. The analysis time including the 2-min sample load was 8 min per sea water sample with a corresponding detection limit of 0.62 ngl(-1) (3 sigma). The average standard deviation calculated for 10 replicate measurements of artificial sea water samples with a concentration of 10 ng l(-1) cobalt was +/-2.1%. The method has been tested with the standard reference sea waters NASS and CASS.