Initial assessment of image quality for low-dose PET: evaluation of lesion detectability

被引:22
|
作者
Schaefferkoetter, Joshua D. [1 ,2 ]
Yan, Jianhua [1 ]
Townsend, David W. [1 ,2 ]
Conti, Maurizio [3 ]
机构
[1] A STAR NUS, Clin Imaging Res Ctr, Singapore, Singapore
[2] Natl Univ Singapore Hosp, Dept Diagnost Radiol, Singapore 117548, Singapore
[3] Siemens Healthcare Mol Imaging, Knoxville, TN 37919 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2015年 / 60卷 / 14期
基金
英国医学研究理事会;
关键词
low-dose PET; image quality; lesion detection; lung screening; CELL LUNG-CANCER; PATHOLOGICAL RESPONSE; COMPUTED-TOMOGRAPHY; RECONSTRUCTION; MODEL; MANAGEMENT; CHEMOTHERAPY; PERFORMANCE; OBSERVER; BENEFITS;
D O I
10.1088/0031-9155/60/14/5543
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In the context of investigating the potential of low-dose PET imaging for screening applications, we developed methods to assess small lesion detectability as a function of the number of counts in the scan. We present here our methods and preliminary validation using tuberculosis cases. FDG-PET data from seventeen patients presenting diffuse hyper-metabolic lung lesions were selected for the study, to include a wide range of lesion sizes and contrasts. Reduced doses were simulated by randomly discarding events in the PET list mode, and ten realizations at each simulated dose were generated and reconstructed. The data were grouped into 9 categories determined by the number of included true events, from >40 M to <250 k counts. The images reconstructed from the original full statistical set were used to identify lung lesions, and each was, at every simulated dose, quantified by 6 parameters: lesion metabolic volume, lesion-to-background contrast, mean lesion tracer uptake, standard deviation of activity measurements (across realizations), lesion signal-to-noise ratio (SNR), and Hotelling observer SNR. Additionally, a lesion-detection task including 550 images was presented to several experienced image readers for qualitative assessment. Human observer performances were ranked using receiver operating characteristic analysis. The observer results were correlated with the lesion image measurements and used to train mathematical observer models. Absolute sensitivities and specificities of the human observers, as well as the area under the ROC curve, showed clustering and performance similarities among images produced from 5 million or greater counts. The results presented here are from a clinically realistic but highly constrained experiment, and more work is needed to validate these findings with a larger patient population.
引用
收藏
页码:5543 / 5556
页数:14
相关论文
共 50 条
  • [1] Image quality evaluation of real low-dose breast PET
    Satoh, Yoko
    Imai, Masamichi
    Ikegawa, Chihiro
    Onishi, Hiroshi
    [J]. JAPANESE JOURNAL OF RADIOLOGY, 2022, 40 (11) : 1186 - 1193
  • [2] Image quality evaluation of real low-dose breast PET
    Yoko Satoh
    Masamichi Imai
    Chihiro Ikegawa
    Hiroshi Onishi
    [J]. Japanese Journal of Radiology, 2022, 40 : 1186 - 1193
  • [3] Assessment of Image Quality and Lesion Detectability With Digital PET/CT System
    Delcroix, Olivier
    Bourhis, David
    Keromnes, Nathalie
    Robin, Philippe
    Le Roux, Pierre-Yves
    Abgral, Ronan
    Salaun, Pierre-Yves
    Querellou, Solene
    [J]. FRONTIERS IN MEDICINE, 2021, 8
  • [4] Image quality and lesion detectability in low-dose pediatric 18F-FDG scans using total-body PET/CT
    Yu-Mo Zhao
    Ying-He Li
    Tao Chen
    Wei-Guang Zhang
    Lin-Hao Wang
    Jiatai Feng
    Chenwei Li
    Xu Zhang
    Wei Fan
    Ying-Ying Hu
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2021, 48 : 3378 - 3385
  • [5] Image quality and lesion detectability in low-dose pediatric 18F-FDG scans using total-body PET/CT
    Zhao, Yu-Mo
    Li, Ying-He
    Chen, Tao
    Zhang, Wei-Guang
    Wang, Lin-Hao
    Feng, Jiatai
    Li, Chenwei
    Zhang, Xu
    Fan, Wei
    Hu, Ying-Ying
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (11) : 3378 - 3385
  • [6] Quantitative Accuracy and Lesion Detectability of Low-Dose FDG-PET for Lung Cancer Screening
    Schaefferkoetter, Josh
    Townsend, David
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [7] Effect of acquisition time on image quality in low-dose PET scanning
    Tanaka, A.
    Kubo, N.
    Akagi, H.
    Harada, T.
    Kameya, T.
    Kawai, Y.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S488 - S488
  • [8] Quantitative Accuracy and Lesion Detectability of Low-Dose 18F-FDG PET for Lung Cancer Screening
    Schaefferkoetter, Joshua D.
    Yan, Jianhua
    Sjoholm, Therese
    Townsend, David W.
    Conti, Maurizio
    Tam, John Kit Chung
    Soo, Ross A.
    Tham, Ivan
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2017, 58 (03) : 399 - 405
  • [9] Low-dose computed tomography perceptual image quality assessment
    Lee, Wonkyeong
    Wagner, Fabian
    Galdran, Adrian
    Shi, Yongyi
    Xia, Wenjun
    Wang, Ge
    Mou, Xuanqin
    Ahamed, Md. Atik
    Imran, Abdullah Al Zubaer
    Oh, Ji Eun
    Kim, Kyungsang
    Baek, Jong Tak
    Lee, Dongheon
    Hong, Boohwi
    Tempelman, Philip
    Lyu, Donghang
    Kuiper, Adrian
    van Blokland, Lars
    Calisto, Maria Baldeon
    Hsieh, Scott
    Han, Minah
    Baek, Jongduk
    Maier, Andreas
    Wang, Adam
    Gold, Garry Evan
    Choi, Jang-Hwan
    [J]. Medical Image Analysis, 2025, 99
  • [10] Effects of CNN-based PET image denoising on image quality and lesion detectability
    Gnesin, S.
    Medici, S.
    Jreige, M.
    Conti, M.
    Schaefferkoetter, J.
    Prior, J. O.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S731 - S731