Amperometric sensor for monitoring of dissolved carbon dioxide in seawater

被引:16
|
作者
Ishiji, T
Chipman, DW
Takahashi, T
Takahashi, K
机构
[1] Riken Keiki Co Ltd, Itabashi Ku, Tokyo 1748744, Japan
[2] Columbia Univ, Lamont Doherty Geol Observ, Palisades, NY 10964 USA
[3] RIKEN, Inst Phys & Chem Res, Wako, Saitama 3510198, Japan
关键词
dissolved carbon dioxide; amperometric sensor; seawater; ruthenium oxide;
D O I
10.1016/S0925-4005(01)00580-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Characteristics of an amperometric CO2 sensor have been investigated for the measurement of CO2 concentration in air and CO2 partial pressure in seawater. Use of RuO2 for the gas permeable pH-dependent electrode, in place of PtO, leads to improved sensitivity. The sensor showed good current response for a wide range of CO2 concentrations in air. For seawater, the partial pressure of CO2 can be measured continuously with the sensor. CO2 from the sample solution permeates into the sensor electrolyte through the membrane in a similar manner as with gas sensing. The CO2 concentration or partial pressure can be detected as a current of pH dependent RuO2 reduction induced by hydrogen ions supplied from CO2. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:265 / 269
页数:5
相关论文
共 50 条
  • [1] A Fiber-Integrated CRDS Sensor for In-Situ Measurement of Dissolved Carbon Dioxide in Seawater
    Hu, Mai
    Chen, Bing
    Yao, Lu
    Yang, Chenguang
    Chen, Xiang
    Kan, Ruifeng
    [J]. SENSORS, 2021, 21 (19)
  • [2] Miniature mid-infrared photoacoustic gas sensor for detecting dissolved carbon dioxide in seawater
    Wang, Zhengzhi
    Zhang, Yajie
    Huang, Xijie
    Wang, Heng
    Zhao, Jikuan
    Li, Chenxi
    Zhu, Ming
    Chen, Ke
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2024, 405
  • [3] An optical sensor for the in situ monitoring of dissolved carbon dioxide in microbial fermentations.
    Spokane, RB
    Pitts, GE
    Wu, P
    Cordonnier, MJ
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1999, 217 : U188 - U189
  • [4] PIEZOELECTRIC SENSOR FOR DISSOLVED CARBON-DIOXIDE
    FOGLEMAN, WW
    SHUMAN, MS
    [J]. ANALYTICAL LETTERS, 1976, 9 (08) : 751 - 765
  • [5] Electrochemical sensor for dissolved carbon dioxide measurement
    Yang, S
    Wang, M
    [J]. JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2002, 149 (01) : H28 - H32
  • [6] AN AMPEROMETRIC SENSOR FOR CARBON-DIOXIDE BASED ON IMMOBILIZED BACTERIA UTILIZING CARBON-DIOXIDE
    SUZUKI, H
    TAMIYA, E
    KARUBE, I
    [J]. ANALYTICA CHIMICA ACTA, 1987, 199 : 85 - 91
  • [7] Development of a NASICON-based amperometric carbon dioxide sensor
    Yang, YB
    Liu, CC
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2000, 62 (01) : 30 - 34
  • [8] AMPEROMETRIC SENSOR FOR CARBON-DIOXIDE - DESIGN, CHARACTERISTICS, AND PERFORMANCE
    EVANS, J
    PLETCHER, D
    WARBURTON, PRG
    GIBBS, TK
    [J]. ANALYTICAL CHEMISTRY, 1989, 61 (06) : 577 - 580
  • [9] Ratiometric fluorescence-based dissolved carbon dioxide sensor for use in environmental monitoring applications
    Dorota Wencel
    John P. Moore
    Niall Stevenson
    Colette McDonagh
    [J]. Analytical and Bioanalytical Chemistry, 2010, 398 : 1899 - 1907
  • [10] A novel miniaturized sensor for carbon dioxide dissolved in liquids
    Fasching, R
    Keplinger, F
    Hanreich, G
    Jobst, G
    Urban, G
    Kohl, F
    Chabicovsky, R
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2001, 78 (1-3) : 291 - 297