Reproducibility and agreement of four anterior segment-optical coherence tomography devices for anterior chamber angle measurements

被引:8
|
作者
Pan, Xiaojing [1 ]
Maram, Jyotsna [2 ]
Nittala, Muneeswar G. [2 ]
Francis, Brian A. [3 ,4 ]
Chopra, Vikas [3 ,4 ]
Sadda, Srinivas R. [3 ,4 ]
机构
[1] Shandong First Med Univ & Shandong Acad Med Sci, Qingdao Eye Hosp, Shandong Eye Inst, Qingdao, Shandong, Peoples R China
[2] Univ Calif Los Angeles, Doheny Image Reading Ctr, Doheny Eye Inst, Los Angeles, CA USA
[3] Doheny Eye Inst, Doheny Image Reading Ctr, 1355 San Pablo St, Los Angeles, CA 90033 USA
[4] Univ Calif Los Angeles, David Geffen Sch Med, Dept Ophthalmol, Los Angeles, CA 90095 USA
关键词
Anterior chamber angle; AS-OCTs; Reproducibility; Agreement; Metrics; SCHWALBES LINE; OCT;
D O I
10.1007/s00417-020-04648-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the reproducibility and agreement of anterior chamber angle (ACA) parameters and metrics obtained by four different anterior segment-optical coherence tomography (AS-OCT) devices. Methods In this prospective study, 30 eyes from 15 normal subjects underwent anterior segment angle scanning using the Spectralis, Cirrus, and Optovue spectral domain optical coherence tomography (SD-OCT), as well as the Visante time-domain optical coherence tomography (TD-OCT). For each eye, the scan line was performed perpendicularly on the inferior (270 degrees) angle, and the inferior ACA image was acquired 2 times. Inter-instrument and intra-instrument, as well as inter-observer and intra-observer reproducibility of anterior chamber angle metrics, Schwalbe's line (SL) to scleral spur (SS) distance (TM-Span), angle opening distance (AOD), and trabecular iris space area (TISA) measurements, were evaluated by intraclass correlation coefficients (ICCs) and Bland-Altman plots with limits of agreement (LoA). Results For this cohort of 30 eyes of 15 normal subjects, the mean TM-Span, AOD, and TISA were 0.966 +/- 0.198 mm, 0.750 +/- 0.205 mm, and 0.286 +/- 0.090 mm(2) from the Spectralis; 0.929 +/- 0.113 mm, 0.717 +/- 0.120 mm, and 0.267 +/- 0.095 mm(2) from the Cirrus; 0.923 +/- 0.191 mm, 0.683 +/- 0.161 mm, and 0.265 +/- 0.072 mm(2) from the Optovue; and 0.970 +/- 0.070 mm, 0.705 +/- 0.150 mm, and 0.279 +/- 0.065 mm(2) from the Visante. The intra-instrument (ICCs > 0.838), intra-grader (ICCs > 0.910), and inter-grader (ICCs > 0.869) agreement were good. Agreement between the four instruments was also good with ICCs from 0.901 to 0.967 for TM-Span, 0.887 to 0.941 for AOD, and 0.923 to 0.961 for TISA. Conclusions Consistent and reproducible ACA measurements could be obtained from multiple AS-OCT devices including both SD- and TD-OCT instruments. These findings have relevance when considering multiple imaging devices in future studies.
引用
收藏
页码:1475 / 1481
页数:7
相关论文
共 50 条
  • [1] Reproducibility and agreement of four anterior segment-optical coherence tomography devices for anterior chamber angle measurements
    Xiaojing Pan
    Jyotsna Maram
    Muneeswar G. Nittala
    Brian A. Francis
    Vikas Chopra
    Srinivas R. Sadda
    [J]. Graefe's Archive for Clinical and Experimental Ophthalmology, 2020, 258 : 1475 - 1481
  • [2] Reproducibility of Anterior Chamber Angle Measurements with Anterior Segment Optical Coherence Tomography
    Tan, Annelie N.
    Sauren, Loes D. C.
    de Brabander, John
    Berendschot, Tos T. J. M.
    Passos, Valeria Lima
    Webers, Carroll A. B.
    Nuijts, Rudy M. M. A.
    Beckers, Henny J. M.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (05) : 2095 - 2099
  • [3] Reproducibility of anterior chamber angle measurements obtained with anterior segment optical coherence tomography
    Radhakrishnan, Sunita
    See, Jovina
    Smith, Scott D.
    Nolan, Winifred P.
    Ce, Zheng
    Friedman, David S.
    Huang, David
    Li, Yan
    Aung, Tin
    Chew, Paul T. K.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (08) : 3683 - 3688
  • [4] Reproducibility of Anterior Chamber Angle Measurements With Anterior Segment Coherence Tomography
    De Brabander, J.
    Tan, A. N.
    Berendschot, T. T. J. M.
    Sauren, L. D. C.
    Nekeman, D.
    Passos, V. Lima
    Webers, C. A. B.
    Beckers, H. J. M.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [5] Evaluation of Scanning Protocols for Imaging the Anterior Chamber Angle With Anterior Segment-Optical Coherence Tomography
    Khor, Wei-Boon
    Sakata, Lisandro M.
    Friedman, David S.
    Narayanaswamy, Arun
    Lavanya, Raghavan
    Perera, Shamira A.
    Aung, Tin
    [J]. JOURNAL OF GLAUCOMA, 2010, 19 (06) : 365 - 368
  • [6] Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography
    Li, Haitao
    Leung, Christopher Kai Shun
    Cheung, Carol Yim Lui
    Wong, Lee
    Pang, Chi Pui
    Weinreb, Robert Neal
    Lam, Dennis Shun Chiu
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (11) : 1490 - 1492
  • [7] Reproducibility and Agreement Between 2 Spectral Domain Optical Coherence Tomography Devices for Anterior Chamber Angle Measurements
    Marion, Kenneth M.
    Maram, Jyotsna
    Pan, Xiaojing
    Dastiridou, Anna
    Zhang, ZhouYuan
    Ho, Alex
    Francis, Brian A.
    Sadda, Srinivas R.
    Chopra, Vikas
    [J]. JOURNAL OF GLAUCOMA, 2015, 24 (09) : 642 - 646
  • [8] Characteristics and reproducibility of anterior chamber angle assessment by anterior-segment optical coherence tomography
    Kim, Dong Y.
    Sung, Kyung R.
    Kang, Sung Y.
    Cho, Jung W.
    Lee, Kyoung S.
    Park, Seong B.
    Kim, Soon T.
    Kook, Michael S.
    [J]. ACTA OPHTHALMOLOGICA, 2011, 89 (05) : 435 - 441
  • [9] Repeatability and reproducibility of pachymetric mapping with visante anterior segment-optical coherence tomography
    Mohamed, Shaheeda
    Lee, Gary K. Y.
    Rao, Srinivas K.
    Wong, Amy L.
    Cheng, Arthur C. K.
    Li, Emmy Y. M.
    Chi, Stanley C. C.
    Lam, Dennis S. C.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (12) : 5499 - 5504
  • [10] Evaluation of the Narrow Anterior Chamber Angle by Anterior Segment Optical Coherence Tomography
    Lan, Jie
    Sun, Dapeng
    Alabdulrasool, Khatoon
    Yusuf, Hassan Ebrahim
    Zhang, Lu
    [J]. OPHTHALMIC RESEARCH, 2017, 57 (02) : 87 - 91