Correction of patient motion in cone-beam CT using 3D-2D registration

被引:26
|
作者
Ouadah, S. [1 ]
Jacobson, M. [1 ]
Stayman, J. W. [1 ]
Ehtiati, T. [2 ]
Weiss, C. [1 ]
Siewerdsen, J. H. [1 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Siemens Med Solutions USA Inc, Imaging & Therapy Syst, Hoffman Estates, IL 60192 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2017年 / 62卷 / 23期
基金
美国国家卫生研究院;
关键词
3D-2D registration; C-arm; cone-beam CT; image-guided interventions; motion correction; image quality; motion artifact; IMAGE REGISTRATION; COMPENSATION; CALIBRATION; REDUCTION;
D O I
10.1088/1361-6560/aa9254
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Cone-beam CT (CBCT) is increasingly common in guidance of interventional procedures, but can be subject to artifacts arising from patient motion during fairly long (similar to 5-60 s) scan times. We present a fiducial-free method to mitigate motion artifacts using 3D-2D image registration that simultaneously corrects residual errors in the intrinsic and extrinsic parameters of geometric calibration. The 3D-2D registration process registers each projection to a prior 3D image by maximizing gradient orientation using the covariance matrix adaptationevolution strategy optimizer. The resulting rigid transforms are applied to the system projection matrices, and a 3D image is reconstructed via model-based iterative reconstruction. Phantom experiments were conducted using a Zeego robotic C-arm to image a head phantom undergoing 5-15 cm translations and 5-15 degrees rotations. To further test the algorithm, clinical images were acquired with a CBCT head scanner in which long scan times were susceptible to significant patient motion. CBCT images were reconstructed using a penalized likelihood objective function. For phantom studies the structural similarity (SSIM) between motion-free and motion-corrected images was > 0.995, with significant improvement (p < 0.001) compared to the SSIM values of uncorrected images. Additionally, motion-corrected images exhibited a point-spread function with full-width at half maximum comparable to that of the motion-free reference image. Qualitative comparison of the motion-corrupted and motion-corrected clinical images demonstrated a significant improvement in image quality after motion correction. This indicates that the 3D-2D registration method could provide a useful approach to motion artifact correction under assumptions of local rigidity, as in the head, pelvis, and extremities. The method is highly parallelizable, and the automatic correction of residual geometric calibration errors provides added benefit that could be valuable in routine use.
引用
收藏
页码:8813 / 8831
页数:19
相关论文
共 50 条
  • [1] Correction of Patient Motion in C-Arm Cone-Beam CT Using 3D-2D Registration
    Ouadah, S.
    Jacobson, M.
    Stayman, J. W.
    Ehtiati, T.
    Siewerdsen, J. H.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3792 - 3793
  • [2] Self-calibration of cone-beam CT geometry using 3D-2D image registration
    Ouadah, S.
    Stayman, J. W.
    Gang, G. J.
    Ehtiati, T.
    Siewerdsen, J. H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (07): : 2613 - 2632
  • [3] Motion-Compensated Targeting in Pulmonary Interventions Using Cone-Beam CT and Locally Rigid/Globally Deformable 3D-2D Registration
    Vijayan, Rohan C.
    Sheth, Niral
    Mekki, Lina
    Lu, Alexander
    Uneri, Ali
    Sisniega, Alejandro
    Maggaragia, Jessica
    Kleinszig, Gerhard
    Vogt, Sebastian
    Thiboutot, Jeffrey
    Lee, Hans
    Yarmus, Lonny
    Siewerdsen, Jeffrey H.
    [J]. MEDICAL IMAGING 2022: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2022, 12034
  • [4] An unsupervised 2D-3D deformable registration network (2D3D-RegNet) for cone-beam CT estimation
    Zhang, You
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (07):
  • [5] 3D-2D spatiotemporal registration for sports motion analysis
    Wang, Ruixuan
    Leow, Wee Kheng
    Leong, Hon Wai
    [J]. 2008 IEEE CONFERENCE ON COMPUTER VISION AND PATTERN RECOGNITION, VOLS 1-12, 2008, : 1444 - 1451
  • [6] Correction of motion artifacts in cone-beam CT using a patient-specific respiratory motion model
    Zhang, Qinghui
    Hu, Yu-Chi
    Liu, Fenghong
    Goodman, Karyn
    Rosenzweig, Kenneth E.
    Mageras, Gig S.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 2901 - 2909
  • [7] Respiratory Motion Correction of Cone-Beam CT in Abdomen Using a Patient-Specific Motion Model
    Zhang, Q.
    Hu, Y. C.
    Kriminski, S.
    Goodman, K.
    Rosenzweig, K. E.
    Mageras, G. S.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [8] 2D/3D Registration for Motion Compensated Reconstruction in Cone-Beam CT of Knees Under Weight-Bearing Condition
    Berger, M.
    Mueller, K.
    Choi, J. -H.
    Aichert, A.
    Maier, A.
    Fahrig, R.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 54 - 57
  • [9] Self-Calibration of Cone-Beam CT Geometry Using 3D-2D Image Registration: Development and Application to Tasked-Based Imaging with a Robotic C-Arm
    Ouadah, S.
    Stayman, J. W.
    Gang, G.
    Uneri, A.
    Ehtiati, T.
    Siewerdsen, J. H.
    [J]. MEDICAL IMAGING 2015: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2015, 9415
  • [10] CT to Cone-Beam CT Deformable Registration With Simultaneous Intensity Correction
    Zhen, X.
    Yan, H.
    Gu, X.
    Zhou, L.
    Jia, X.
    Jiang, S.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3960 - 3960