Optimization based trajectory planning for real-time 6DoF robotic patient motion compensation systems

被引:32
|
作者
Liu, Xinmin [1 ]
Wiersma, Rodney D. [1 ]
机构
[1] Univ Chicago, Dept Radiat & Cellular Oncol, Chicago, IL 60637 USA
来源
PLOS ONE | 2019年 / 14卷 / 01期
关键词
VOLUMETRIC MODULATED ARC; ORGAN MOTION; QUALITY-ASSURANCE; TRACKING SYSTEM; TREATMENT COUCH; TUMOR TRACKING; RADIOTHERAPY; THERAPY; CANCER; HEAD;
D O I
10.1371/journal.pone.0210385
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose Robotic stabilization of a therapeutic radiation beam with respect to a dynamically moving tumor target can be accomplished either by moving the radiation source, the patient, or both. As the treatment beam is on during this process, the primary goal is to minimize exposure of normal tissue to radiation as much as possible when moving the target back to the desired position. Due to the complex mechanical structure of 6 degree-of-freedom (6DoF) robots, it is not intuitive as to what 6 dimensional (6D) correction trajectory is optimal in achieving such a goal. With proportional-integrative-derivative (PID) and other controls, the potential exists that the controller may generate a trajectory that is highly curved, slow, or suboptimal in that it leads to unnecessary exposure of healthy tissue to radiation. This work investigates a novel feedback planning method that takes into account a robot's mechanical joint structure, patient safety tolerances, and other system constraints, and performs real-time optimization to search the entire 6D trajectory space in each time cycle so it can respond with an optimal 6D correction trajectory. Methods Computer simulations were created for two 6DoF robotic patient support systems: a Stewart-Gough platform for moving a patient's head in frameless maskless stereotactic radiosurgery, and a linear accelerator treatment table for moving a patient in prostate cancer radiation therapy. Motion planning was formulated as an optimization problem and solved at real-time speeds using the L-BFGS algorithm. Three planning methods were investigated, moving the platform as fast as possible (platform-D), moving the target along a straight-line (target-S), and moving the target based on the fastest descent of position error (target-D). Both synthetic motion and prior recorded human motion were used as input data and output results were analyzed. Results For randomly generated 6D step-like and sinusoidal synthetic input motion, target-D planning demonstrated the smallest net trajectory error in all cases. On average, optimal planning was found to have a 45% smaller target trajectory error than platform-D control, and a 44% smaller target trajectory error than target-S planning. For patient head motion compensation, only target-D planning was able to maintain a <= 0.5mm and <= 0.5deg clinical tolerance objective for 100% of the treatment time. For prostate motion, both target-S planning and target-D planning outperformed platform-D control. Conclusions A general 6D target trajectory optimization framework for robotic patient motion compensation systems was investigated. The method was found to be flexible as it allows control over various performance requirements such as mechanical limits, velocities, acceleration, or other system control objectives.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] A General Patient Motion Compensation in Robotic Systems Using Optimization Based 6DoF Trajectory Planning
    Liu, X.
    Wiersma, R.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E644 - E644
  • [2] Towards frameless maskless SRS through real-time 6DoF robotic motion compensation
    Belcher, Andrew H.
    Liu, Xinmin
    Chmura, Steven
    Yenice, Kamil
    Wiersma, Rodney D.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (23): : 9054 - 9066
  • [3] Positioning Accuracy of a Robotic Motion Head Platform for Real-Time 6DoF Robotic Motion-Compensated SRS
    Farrey, K.
    Yenice, K.
    Ferjancic, P.
    Flores-Martinez, E.
    Wiersma, R.
    Denis, C.
    Kelly, J.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E362 - E362
  • [4] Towards Frameless Maskless SRS Through Real-Time 6DOF Robotic Head Motion Stabilization
    Belcher, A. H.
    Liu, X.
    Chmura, S.
    Yenice, K.
    Wiersma, R. D.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3062 - 3062
  • [5] Robotic Motion Compensation for Radiation Therapy: A 6DOF Phantom Study
    Belcher, A. H.
    Liu, X.
    Wiersma, R.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3896 - 3896
  • [6] Trajectory Optimization for a 6 DOF Robotic Arm Based on Reachability Time
    Mousa, Mahmoud A. A.
    Elgohr, Abdelrahman T.
    Khater, Hatem A.
    [J]. Annals of Emerging Technologies in Computing, 2024, 8 (01) : 22 - 35
  • [7] Real-time scalable 6DOF pose estimation for textureless objects
    Cao, Zhe
    Sheikh, Yaser
    Banerjee, Natasha Kholgade
    [J]. 2016 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), 2016, : 2441 - 2448
  • [8] Real-time jerk-minimization trajectory planning of robotic arm based on polynomial curve optimization
    Wu, Guanglei
    Zhang, Sida
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 2022, 236 (21) : 10852 - 10864
  • [9] A Portable Real-Time Image-Guided 6DoF Robotic Positioning System for Canine FLASH Trials
    Sakaamini, A.
    Liu, X.
    Wiersma, R.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E776 - E777
  • [10] ART-SLAM: Accurate Real-Time 6DoF LiDAR SLAM
    Frosi, Matteo
    Matteucci, Matteo
    [J]. IEEE ROBOTICS AND AUTOMATION LETTERS, 2022, 7 (02) : 2692 - 2699