New impedance bridge at LNE for the two terminal-pair AC resistor calibration up to 20 kHz

被引:0
|
作者
Ouameur, Mohamed [1 ]
Vasconcellos, Renata [1 ]
Agazar, Mohamed [1 ]
Thevenot, Olivier [1 ]
Patois, Emmanuel [1 ]
机构
[1] Lab Natl Metrol & Essais, Paris, France
关键词
impedance metrology; calibration; PXI modules; Wheatstone bridge; AC resistors; CURRENT SHUNTS;
D O I
10.1088/1681-7575/ac8485
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper presents the evolution of LNE's impedance bridge (Wheatstone bridge) used to calibrate AC resistors defined in the two terminal-pair configuration. A new automated Wheatstone bridge, based on four resistances, was developed to operate in the full complex plane for resistances ranging from 400 omega to 2 M omega and for frequencies up to 20 kHz. A commercial dual precision arbitrary waveform generator supplies the bridge and PXI modules achieve the balance. Phase and frequency synchronization solutions between the arbitrary waveform generator source and the PXI modules were used. The choice of these instruments allows the bridge to be easily automated and to benefit from recent progress in the development of digital-analog converters, in particular those with high frequency sampling rates and resolution. The bridge is mainly dedicated to realize the periodic calibration of LNE's standard resistors, used for routine customer calibrations. The traceability of LNE's standard resistors is ensured, firstly, by the calibration of a 1 k omega resistor by comparison to a calculable resistor up to 20 kHz. Then, this resistor is used as a reference element to ensure the traceability of all LNE's standard resistors using the new bridge. The validation was done by comparing the results obtained, for some resistors with very low drift, with those obtained during calibrations performed in 1983, 1995 and 2009. For resistors of 1 k omega in a 1:1 ratio, the expanded uncertainty of series resistance variation and argument are, respectively, less than 1.3 mu omega omega(-1) (k = 1) and 1.2 mu rad (k = 1), up to 20 kHz.
引用
收藏
页数:13
相关论文
共 18 条
  • [1] A new four terminal-pair bridge for impedance measurements at frequencies up to 1 MHz
    Awan, SA
    Kibble, BP
    Robinson, IA
    Giblin, SP
    [J]. 2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 172 - 173
  • [2] A Digital Four Terminal-Pair Impedance Bridge
    Ihlenfeld, W. G. Kurten
    Vasconcellos, R. T. B.
    [J]. 2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [3] A new four terminal-pair bridge for traceable impedance measurements at frequencies up to 1 MHz
    Awan, SA
    Kibble, BP
    Robinson, IA
    Giblin, SP
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2001, 50 (02) : 282 - 285
  • [4] A FOUR TERMINAL-PAIR COAXIAL IMPEDANCE BRIDGE CONSTRUCTED AT INMETRO
    de Barros e Vasconcellos, Renata
    Ogino, Luiz Macoto
    [J]. 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 396 - 397
  • [5] A reconfigurable four terminal-pair digitally assisted and fully digital impedance ratio bridge
    Kucera, Jan
    Kovac, Jakub
    [J]. 2017 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2017, : 674 - 679
  • [6] A Reconfigurable Four Terminal-Pair Digitally Assisted and Fully Digital Impedance Ratio Bridge
    Kucera, Jan
    Kovac, Jakub
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2018, 67 (05) : 1199 - 1206
  • [7] A NEW UNIVERSAL CALIBRATION METHOD FOR 4-PAIR TERMINAL-PAIR ADMITTANCE STANDARDS
    SUZUKI, K
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1991, 40 (02) : 420 - 422
  • [8] The Josephson two-terminal-pair impedance bridge
    Lee, Jinni
    Schurr, Juergen
    Nissila, Jaani
    Palafox, Luis
    Behr, Ralf
    [J]. METROLOGIA, 2010, 47 (04) : 453 - 459
  • [9] Experiences With a Two-Terminal-Pair Digital Impedance Bridge
    Callegaro, Luca
    D'Elia, Vincenzo
    Kampik, Marian
    Kim, Dan Bee
    Ortolano, Massimo
    Pourdanesh, Faranak
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2015, 64 (06) : 1460 - 1465
  • [10] Four Terminal Pair Digital Sampling Impedance Bridge up to 1 MHz
    Maslan, Stanislav
    Sira, Martin
    Skalicka, Tereza
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,