Backside wear in acetabular hip joint replacement

被引:17
|
作者
Braun, Steffen [1 ]
Sonntag, Robert [1 ]
Schroeder, Stefan [1 ]
Mueller, Ulrike [1 ]
Jaeger, Sebastian [1 ]
Gotterbarm, Tobias [1 ,2 ]
Kretzer, Jan Philippe [1 ]
机构
[1] Heidelberg Univ Hosp, Clin Orthoped & Trauma Surg, Lab Biomech & Implant Res, Schlierbacher Landstr 200a, D-69118 Heidelberg, Germany
[2] Kepler Univ Hosp, Dept Orthoped, Krankenhausstr 9, A-4021 Linz, Austria
关键词
Total hip replacement; Hip cup system; Micro-motions; Backside wear; Particle analysis; MOLECULAR-WEIGHT POLYETHYLENE; OSTEOLYSIS; PARTICLES; LINERS; SIZE; MACROPHAGES; PROGRESSION; SIMULATION; DEBRIS;
D O I
10.1016/j.actbio.2018.10.045
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Introduction: Besides head-insert articulation in hip joint replacements, micro-motions between the backside of assembled polyethylene acetabular liners and the metal cup may cause additional wear. Pelvic osteolysis frequently occurs in the region of screw holes, and cup loosening hints to clinically relevant amounts of polyethylene backside wear. It has yet to be confirmed whether backside wear particles differ in size and morphology compared to articulating wear. Previous methods have been limited to subjective assessment of backside surface damages without consideration of wear debris. The aim of this study was to develop and validate a method for quantitative in vitro measurements of polyethylene backside wear in artificial hip cups and to characterize these wear particles for the first time. Methods Titanium cup-systems (Plasmafit (R) Plus7, Aesculap, UHMWPE liner) were sinusoidally loaded (2.5 kN) and a torque of 5 Nm was simultaneously applied. The front and rear side of the cup were separated to isolate backside wear. After 2 x 10(6) cycles the surrounding fluid was filtered and a particle analysis was performed. Results Backside wear had a particles size of 64.1 +/- 1.9 nm and was verified as round and oval particles with partly rough outlines. An estimated total number of particles of 1.26 x 10(9) +/- 1.67 x 10(8) per 10(6) cycles was determined. Conclusion Backside wear was estimated to be several times lower than published values of articulating wear. However, polyethylene backside wear particles represented significantly smaller particles with partly roughened outlines than articulating wear particles and may therefore cause higher biological response in macrophage-mediated bone resorption compared to articulated particles. Statement of Significance Within this study, an analytical method for quantitative measuring polyethylene backside wear of artificial hip cups was successfully developed and validated for the first time. It could be shown that backside wear is still present, even in modern cup-systems. These findings can be further used for investigations of the osteolytic potential of polyethylene particles, for evaluating and improving new implant systems and to evaluate the effectiveness of screw hole plugs to prevent the particle migration to the acetabulum. (C) 2018 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:467 / 476
页数:10
相关论文
共 50 条
  • [1] Quantitative Measurements of Backside Wear in Acetabular Hip Joint Replacement: Conventional Polyethylene Versus Cross-Linked Polyethylene
    Braun, Steffen
    Jaeger, Sebastian
    Sonntag, Robert
    Schroeder, Stefan
    Kretzer, J. Philippe
    MATERIALS, 2020, 13 (08)
  • [2] Backside wear, particle migration and effectiveness of screw hole plugs in acetabular hip joint replacement with cross-linked polyethylene
    Braun, Steffen
    Vardag, Shirin
    Mueller, Ulrike
    Schroeder, Stefan
    Sonntag, Robert
    Bormann, Therese
    Gotterbarm, Tobias
    Kretzer, Jan Philippe
    ACTA BIOMATERIALIA, 2019, 97 : 239 - 246
  • [3] Evaluation of acetabular wear in hip joint prostheses
    Pietrabissa, R
    Raimondi, M
    Quaglini, V
    Contro, R
    COMPUTER METHODS IN BIOMECHANICS & BIOMEDICAL ENGINEERING - 2, 1998, : 131 - 138
  • [4] The effect of joint loading on acetabular wear measurement in total hip arthroplasty
    Martell, JM
    Leopold, SS
    Liu, XL
    JOURNAL OF ARTHROPLASTY, 2000, 15 (04): : 512 - 518
  • [5] PROBLEM OF ACETABULAR LOOSENING IN PATIENTS WITH TOTAL HIP-JOINT REPLACEMENT
    TAGER, KH
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1977, 87 (01) : 39 - 49
  • [6] (PROBLEMS OF FIXING ACETABULAR COMPONENT IN TOTAL REPLACEMENT OF HIP-JOINT)
    WITT, AN
    HACKENBROCH, MH
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1972, 110 (03): : 292 - +
  • [7] Hip joint acetabular dysplasia
    Wedge, JH
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 1997, 17 (02) : 141 - 142
  • [8] EXPERIENCE ON HIP-JOINT REPLACEMENT USING A SPECIALLY DESIGNED ACETABULAR CUP
    SCHOLZ, J
    GERBIG, W
    JOURNAL OF BIOMECHANICS, 1980, 13 (09) : 807 - 807
  • [9] ARTIFICIAL REPLACEMENT OF THE HIP-JOINT IN SEVERE COXARTHROSE SECONDARY TO ACETABULAR DYSPLASIA
    MAYER, G
    HARTSEIL, K
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1984, 103 (05) : 332 - 336
  • [10] HIP JOINT ACETABULAR PROSTHESIS
    MCBRIDE, E
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1960, 42 (02): : 356 - 357