Active vibration control of gradient coils to reduce acoustic noise of MRI systems

被引:22
|
作者
Roozen, N. B. [1 ,2 ]
Koevoets, A. H. [1 ]
den Hamer, A. J. [2 ]
机构
[1] Philips Appl Technol, Royal Philips Elect, NL-5656 AE Eindhoven, Netherlands
[2] Eindhoven Univ Technol, Dept Mech Engn, NL-5600 MB Eindhoven, Netherlands
关键词
acoustic noise; actuators; control systems; dynamics; feedforward systems; magnetic resonance imaging; piezoelectric transducers; vibration control;
D O I
10.1109/TMECH.2008.924111
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Lorentz-force-induced vibrations in MRI systems cause significant acoustic noise levels during scanning, the main acoustic noise source being the vibrating gradient coil. In this paper, a novel active vibration control technique is presented to reduce vibrations of the gradient coil, and hence, achieve a reduction of acoustic noise during scanning. The active vibration control technique uses seismic masses that are actuated by means of piezo actuators to create forces on the gradient coil counteracting its vibrations. Using four seismic mass actuators, a vibration reduction of 3-8 dB at resonance frequencies is achieved, giving an overall vibration reduction of 3 dB for a typical field gradient (FE)-echo planar imaging (EPI) gradient sequence, as substantiated by measurements. Using eight actuators, an overall vibration reduction of 5 dB is predicted for this sequence.
引用
收藏
页码:325 / 334
页数:10
相关论文
共 50 条
  • [1] Active acoustic control in gradient coils for MRI
    Mansfield, P
    Haywood, B
    Coxon, R
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (04) : 807 - 818
  • [2] ACTIVE ACOUSTIC SCREENING - DESIGN PRINCIPLES FOR QUIET GRADIENT COILS IN MRI
    MANSFIELD, P
    GLOVER, P
    BOWTELL, R
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (08) : 1021 - 1025
  • [3] Shaping and Timing Gradient Pulses to Reduce MRI Acoustic Noise
    Segbers, Marcel
    Sierra, Carlos V. Rizzo
    Duifhuis, Hendrikus
    Hoogduin, Johannes M.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (02) : 546 - 553
  • [4] ACTIVE ACOUSTIC SCREENING - REDUCTION OF NOISE IN GRADIENT COILS BY LORENTZ FORCE BALANCING
    MANSFIELD, P
    CHAPMAN, BLW
    BOWTELL, R
    GLOVER, P
    COXON, R
    HARVEY, PR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (02) : 276 - 281
  • [5] Active vibration and noise control by hybrid active acoustic panels
    Stöbener, U
    Gaul, L
    [J]. ZEITSCHRIFT FUR ANGEWANDTE MATHEMATIK UND MECHANIK, 2001, 81 : S45 - S48
  • [6] A simple theory for vibration of MRI gradient coils
    Tomasi, D
    Ernst, T
    [J]. BRAZILIAN JOURNAL OF PHYSICS, 2006, 36 (1A) : 34 - 39
  • [7] Active-passive gradient shielding for MRI acoustic noise reduction
    Edelstein, WA
    Kidane, TK
    Taracila, V
    Baig, TN
    Eagan, TP
    Cheng, YCN
    Brown, RW
    Mallick, JA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (05) : 1013 - 1017
  • [8] Adaptive active noise and vibration control systems
    Mal'tsev, AA
    Maslennikov, RO
    Khoryaev, AV
    Cherepennikov, VV
    [J]. ACOUSTICAL PHYSICS, 2005, 51 (02) : 195 - 208
  • [9] Adaptive active noise and vibration control systems
    A. A. Mal’tsev
    R. O. Maslennikov
    A. V. Khoryaev
    V. V. Cherepennikov
    [J]. Acoustical Physics, 2005, 51 : 195 - 208
  • [10] Vibration and acoustic noise characterization of a gradient coil insert in a 4 T MRI
    Mechefske, CK