Further elucidation of nanofluid thermal conductivity measurement using a transient hot-wire method apparatus

被引:7
|
作者
Yoo, Donghoon [1 ]
Lee, Joohyun [2 ]
Lee, Byeongchan [1 ]
Kwon, Suyong [2 ]
Koo, Junemo [1 ]
机构
[1] Kyung Hee Univ, Dept Mech Engn, Yongin 17104, South Korea
[2] Korea Res Inst Stand & Sci, Daejeon 34113, South Korea
关键词
transienthot-wiremethod; temperaturecoefficient of resistance (TCR); nanofluid thermal conductivity; power analysis; CONVECTIVE HEAT-TRANSFER; ENHANCEMENT;
D O I
10.1007/s00231-017-2144-y
中图分类号
O414.1 [热力学];
学科分类号
摘要
The Transient Hot-Wire Method (THWM) was developed to measure the absolute thermal conductivity of gases, liquids, melts, and solids with low uncertainty. The majority of nanofluid researchers used THWM to measure the thermal conductivity of test fluids. Several reasons have been suggested for the discrepancies in these types of measurements, including nanofluid generation, nanofluid stability, and measurement challenges. The details of the transient hot-wire method such as the test cell size, the temperature coefficient of resistance (TCR) and the sampling number are further investigated to improve the accuracy and consistency of the measurements of different researchers. It was observed that smaller test apparatuses were better because they can delay the onset of natural convection. TCR values of a coated platinum wire were measured and statistically analyzed to reduce the uncertainty in thermal conductivity measurements. For validation, ethylene glycol (EG) and water thermal conductivity were measured and analyzed in the temperature range between 280 and 310 K. Furthermore, a detailed statistical analysis was conducted for such measurements, and the results confirmed the minimum number of samples required to achieve the desired resolution and precision of the measurements. It is further proposed that researchers fully report the information related to their measurements to validate the measurements and to avoid future inconsistent nanofluid data.
引用
收藏
页码:415 / 424
页数:10
相关论文
共 50 条
  • [1] Further elucidation of nanofluid thermal conductivity measurement using a transient hot-wire method apparatus
    Donghoon Yoo
    Joohyun Lee
    Byeongchan Lee
    Suyong Kwon
    Junemo Koo
    [J]. Heat and Mass Transfer, 2018, 54 : 415 - 424
  • [2] Further examination of the transient hot-wire method for the simultaneous measurement of thermal conductivity and thermal diffusivity
    Watanabe, H
    [J]. METROLOGIA, 2002, 39 (01) : 65 - 81
  • [3] Measurement of the thermal conductivity of argon and methane:: a test of a transient hot-wire apparatus
    Pátek, J
    Klomfar, J
    [J]. FLUID PHASE EQUILIBRIA, 2002, 198 (01) : 147 - 163
  • [4] MEASUREMENT OF THE THERMAL-CONDUCTIVITY OF GASES BY THE TRANSIENT HOT-WIRE METHOD
    JOHNS, AI
    SCOTT, AC
    WATSON, JTR
    FERGUSON, D
    CLIFFORD, AA
    [J]. PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 1988, 325 (1585): : 295 - 356
  • [5] A TRANSIENT HOT-WIRE THERMAL-CONDUCTIVITY APPARATUS FOR FLUIDS
    RODER, HM
    [J]. JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1981, 86 (05): : 457 - 493
  • [7] A hot-wire method based thermal conductivity measurement apparatus for teaching purposes
    Alvarado, S.
    Marin, E.
    Juarez, A. G.
    Calderon, A.
    Ivanov, R.
    [J]. EUROPEAN JOURNAL OF PHYSICS, 2012, 33 (04) : 897 - 906
  • [8] Thermal conductivity measurements using the transient hot-wire method: a review
    Salim, S. G. R.
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2022, 33 (12)
  • [9] RAPID MEASUREMENT OF LIQUID THERMAL-CONDUCTIVITY BY TRANSIENT HOT-WIRE METHOD
    TRUMP, WN
    LUEBKE, HW
    FOWLER, L
    EMERY, EM
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1977, 48 (01): : 47 - 51
  • [10] Influence of insulation coating on thermal conductivity measurement by transient hot-wire method
    Yu, Wenhua
    Choi, Stephen U. -S.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (07):