Calibration methods of force sensors in the micro-Newton range

被引:6
|
作者
Nafari, Alexandra [1 ]
Ghavanini, Farzan Alavian
Bring, Martin
Svensson, Krister
Enoksson, Peter
机构
[1] Chalmers, Dept Microtechnol & Nanosci, Micro & Nanosyst Grp, SE-41296 Gothenburg, Sweden
[2] Nanofactory Instruments AB, SE-41258 Gothenburg, Sweden
[3] Karlstad Univ, Dept Phys, SE-65188 Karlstad, Sweden
关键词
D O I
10.1088/0960-1317/17/10/023
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A micromachined capacitive force sensor operating in the micro-Newton range has been calibrated using both dynamic and static methods. Both calibrations are non-destructive, accurate and traceable to Systeme International ( SI) fundamental units. The dynamic calibration is a differential mass loading resonant method where the resonance frequency with and without an added mass is measured. This gives enough information to compute the spring constant. In this paper, we evaluate the resonant mass loading method for more complex MEMS devices. Analytical calculations and finite element analysis have been performed to investigate the dynamic properties of the sensor, e. g. modal interference. The frequency response was measured with the third harmonic method where the third harmonic of the current through the sensor was measured. To detect and analyse the resonance mode of the structure during excitation, a scanning laser Doppler vibrometer was used. Two designs of a capacitive nanoindenter force sensor with flexure-type springs have been evaluated using these methods. The quality of the resonant calibration method has been tested using static mass loading in combination with transmission electron microscopy imaging of the sensor displacement. This shows that the resonant method can be extended to calibrate more complex structures than plain cantilevers. Both calibration methods used are traceable to SI fundamental units as they are based on masses weighed on a calibrated scale. The masses used do not need to be fixed or glued in any way, making the calibration non-destructive.
引用
收藏
页码:2102 / 2107
页数:6
相关论文
共 50 条
  • [1] Simulation and test for the micro-newton electromagnetic calibration force measurement
    Long, Jianfei
    Cheng, Ye
    Wang, Jiabin
    Cong, Linxiao
    Huang, Dan
    Xu, Luxiang
    Zheng, Yelong
    Guo, Ning
    MEASUREMENT, 2024, 235
  • [2] Design and functional analysis of a micro-Newton force generator
    Wang Y.
    Liu H.-L.
    Liu Z.-S.
    Hu S.-S.
    Guangxue Jingmi Gongcheng/Optics and Precision Engineering, 2018, 26 (10): : 2527 - 2535
  • [3] Precision electromagnetic calibration technique for micro-Newton thrust stands
    He, Zhen
    Wu, Jianjun
    Zhang, Daixian
    Lu, Gaofei
    Liu, Zejun
    Zhang, Rui
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (05):
  • [4] Micro-Newton Thrust Stand Calibration Using Electrostatic Lantern Rings
    Mu, Jianchao
    Cong, Linxiao
    Hu, Qiwei
    Qiao, Congfeng
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2023, 72
  • [5] A disturbance suppression micro-Newton force sensor based on shadow method
    Yang, Yong
    Zhao, Meirong
    Li, Dantong
    Tao, Moran
    Zhu, Chunyuan
    Zheng, Yelong
    Tian, Yu
    ISA TRANSACTIONS, 2023, 134 : 442 - 450
  • [6] Micro-Newton Thrust Stand Calibration Using Electrostatic Lantern Rings
    Mu, Jianchao
    Cong, Linxiao
    Hu, Qiwei
    Qiao, Congfeng
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2023, 72
  • [7] Upconverting Nanoparticles as Optical Sensors of Nano- to Micro-Newton Forces
    Lay, Alice
    Wang, Derek S.
    Wisser, Michael D.
    Mehlenbacher, Randy D.
    Lin, Yu
    Goodman, Miriam B.
    Mao, Wendy L.
    Dionne, Jennifer A.
    NANO LETTERS, 2017, 17 (07) : 4172 - 4177
  • [8] Micro-Newton Detection by Using Graphene-Paper Force Sensor
    Yadegari, Amir
    Omidi, Meisam
    Choolaei, Mohammadmehdi
    Haghiralsadat, F.
    Yazdian, F.
    28TH EUROPEAN CONFERENCE ON SOLID-STATE TRANSDUCERS (EUROSENSORS 2014), 2014, 87 : 967 - 970
  • [9] A gravity component-based calibration method for micro-newton thrust stand
    Zhu, Chunyuan
    Lu, Shixu
    Kang, Dan
    Chen, Congyun
    Xia, Ningning
    Zhao, Meirong
    Wang, Bin
    Zheng, Yelong
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2025, 36 (04)
  • [10] A Method of Obtaining Micro Newton Calibration Force
    Li Shipo
    Lv Xiang
    2021 12TH INTERNATIONAL CONFERENCE ON MECHANICAL AND AEROSPACE ENGINEERING (ICMAE), 2021, : 292 - 297