A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models

被引:111
|
作者
Jarry, G [1 ]
DeMarco, JJ
Beifuss, U
Cagnon, CH
McNitt-Gray, MF
机构
[1] McGill Univ, Med Phys Unit, Montreal Gen Hosp, Montreal, PQ, Canada
[2] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiat Oncol, Los Angeles, CA 90095 USA
[3] BrainLAG AG, Munich, Germany
[4] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiol, Los Angeles, CA 90095 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2003年 / 48卷 / 16期
关键词
D O I
10.1088/0031-9155/48/16/306
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this work is to develop and test a method to estimate the relative and absolute absorbed radiation dose from axial and spiral CT scans using a Monte Carlo approach. Initial testing was done in phantoms and preliminary results were obtained from a standard mathematical anthropomorphic model (MIRD V) and voxelized patient data. To accomplish this we have modified a general purpose Monte Carlo transport code (MCNP4B) to simulate the CT x-ray source and movement, and then to calculate absorbed radiation dose in desired objects. The movement of the source in either axial or spiral modes was modelled explicitly while the CT system components were modelled using published information about x-ray spectra as well as information provided by the manufacturer. Simulations were performed for single axial scans using the head and body computed tomography dose index (CTDI) polymethylmethacrylate phantoms at both central and peripheral positions for all available beam energies and slice thicknesses. For comparison, corresponding physical measurements of CTDI in phantom were made with an ion chamber. To obtain absolute dose values, simulations and measurements were performed in air at the scanner isocentre for each beam energy. To extend the verification, the CT scanner model was applied to the MIRD V model and compared with published results using similar technical factors. After verification of the model, the generalized source was simulated and applied to voxelized models of patient anatomy. The simulated and measured absolute dose data in phantom agreed to within 2% for the head phantom and within 4% for the body phantom at 120 and 140 kVp; this extends to 8% for the head and 9% for the body phantom across all available beam energies and positions. For the head phantom, the simulated and measured absolute dose data agree to within 2% across all slice thicknesses at 120 kVp. Our results in the MIRD phantom agree within 11% of all the different organ dose values published by the UK's ImPACT group for a scan using an equivalent scanner, kVp, collimation, pitch and mAs. The CT source model was shown to calculate both a relative and absolute radiation dose distribution throughout the entire volume in a patient-specific matrix geometry. Results of initial testing are promising and application to patient models was shown to be feasible.
引用
收藏
页码:2645 / 2663
页数:19
相关论文
共 50 条
  • [1] Monte Carlo Dose Estimation in Patient-Specific Volume Mesh From CT Images
    Morato, S.
    Miro, R.
    Juste, B.
    Verdu, G.
    Santos, A.
    2016 9TH INTERNATIONAL CONGRESS ON IMAGE AND SIGNAL PROCESSING, BIOMEDICAL ENGINEERING AND INFORMATICS (CISP-BMEI 2016), 2016, : 1590 - 1593
  • [2] Patient-specific Monte Carlo-based organ dose estimates in spiral CT via optical 3D body scanning and adaptation of a voxelized phantom dataset: proof-of-principle
    Maddaloni, Francesca Saveria
    Sarno, Antonio
    Mettivier, Giovanni
    Clemente, Stefania
    Oliviero, Caterina
    Ricciardi, Roberta
    Varallo, Antonio
    Russo, Paolo
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (08):
  • [3] Monte carlo calculation of radiation dose in CT examinations using phantom and patient tomographic models
    Salvadó, M
    López, M
    Morant, JJ
    Calzado, A
    RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 364 - 368
  • [4] PATIENT-SPECIFIC MONTE CARLO-BASED DOSE-KERNEL APPROACH FOR INVERSE PLANNING IN AFTERLOADING BRACHYTHERAPY
    D'Amours, Michel
    Pouliot, Jean
    Dagnault, Anne
    Verhaegen, Frank
    Beaulieu, Luc
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (05): : 1582 - 1589
  • [5] A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms
    DeMarco, JJ
    Cagnon, CH
    Cody, DD
    Stevens, DM
    McCollough, CH
    O'Daniel, J
    McNitt-Gray, MF
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17): : 3989 - 4004
  • [6] Accuracy of patient-specific CT organ doses from Monte Carlo simulations: influence of CT-based voxel models
    Verfaillie, Gwenny
    Rutten, Jeff
    DAsseler, Yves
    Bacher, Klaus
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (03) : 989 - 1000
  • [7] Monte Carlo methods to simulate radiation dose from spiral computed tomography: Phantom studies
    Beifuss, U
    Cagnon, CH
    Demarco, JJ
    McNitt-Gray, MF
    RADIOLOGY, 2000, 217 : 488 - 488
  • [8] A Monte-Carlo based method to estimate radiation dose from multidetector helical CT: Verification in anthropomorphic phantoms
    DeMarco, J
    Cagnon, C
    O'Daniel, J
    Cody, D
    McCollough, C
    McNitt-Gray, M
    MEDICAL PHYSICS, 2004, 31 (06) : 1841 - 1841
  • [9] Monte Carlo and phantom study of the radiation dose to the body from dedicated CT of the breast
    Sechopoulos, Ioannis
    Vedantham, Srinivasan
    Suryanarayanan, Sankararaman
    D'Orsi, Carl J.
    Karellas, Andrew
    RADIOLOGY, 2008, 247 (01) : 98 - 105
  • [10] A Monte Carlo-based tool for proton dose verification on patient CT data
    Szymanowski, H
    Nill, S
    Oelfke, U
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S319 - S319