摘要
Purpose: To develop the moving phantom for dosimetric and geometric evaluation of the 4D radiotherapy such as gated radiotherapy, breathing control radiotherapy and tumor tracking. Method and Materials: This system consisted of the software, the respiration monitoring mask (ReMM) and the moving phantom. Patient's respiration was measured with thermocouple installed in ReMM. The ReMM was connected directly to the operational amplifier. The respiratory signal was digitized by the data acquisition card in the computer. The program, which acquires and records respiratory signal and generates the digital signal to make the phantom simulate respiratory organ motion, was developed. The phantom moved by the servo motor, to which the program sent the digital signal. The respiration of three lung cancer patients, whose organ motion was greater than 5 mm, was monitored through ReMM and fluoroscopy simultaneously. While the phantom simulated patients' organ motion, its motion was monitored by RPM® to verify the effectiveness of the phantom. The discrepancies between the respiration curve and the organ motion and between the respiration and the phantom motion were estimated as standard deviations for each patient to evaluate the moving phantom. Results: Comparing with the curves of respiration measured by thermocouple and these of the organ motion measured by fluoroscope, the mean standard deviation of discrepancies was 9.68%. The standard deviation of the discrepancy between the respiratory curve and the organ motion was estimated at 8.52%. Conclusion: The patients did not complain about discomfort with the respiration monitoring mask. The phantom could simulate the organ motion according to the respiratory signal from ReMM. It is expected that the simulating phantom could be used to verify the 4D radiation therapy. © 2005, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:1921 / 1922
页数:2
相关论文
共 50 条
  • [1] A robotic 4D phantom for the radiotherapy of moving objects
    Baier, K.
    Wilbert, J.
    Richter, A.
    Guckenberger, M.
    Flentje, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 : 15 - 15
  • [2] Implementing 4D XCAT Phantom for 4D Radiotherapy Research
    Panta, R.
    Segars, W.
    Yin, F.
    Cai, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3686 - 3686
  • [3] A deformable phantom for 4D radiotherapy verification: Design and image registration evaluation
    Serban, Monica
    Heath, Emily
    Stroian, Gabriela
    Collins, D. Louis
    Seuntjens, Jan
    MEDICAL PHYSICS, 2008, 35 (03) : 1094 - 1102
  • [4] Establishing a framework to implement 4D XCAT Phantom for 4D radiotherapy research
    Panta, Raj K.
    Segars, Paul
    Yin, Fang-Fang
    Cai, Jing
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2012, 8 (04) : 565 - 570
  • [5] A deformable phantom for quality assurance in 4D radiotherapy
    Margeanu, M.
    Heath, E.
    Stroian, G.
    Seuntjens, J.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S80 - S81
  • [6] Evaluation of the Elekta Symmetry™ 4D IGRT system by using a moving lung phantom
    Hun-Joo Shin
    Shin-Wook Kim
    Chul Seung Kay
    Jae-Hyuk Seo
    Gi-Woong Lee
    Ki-Mun Kang
    Hong Seok Jang
    Young-nam Kang
    Journal of the Korean Physical Society, 2015, 67 : 260 - 263
  • [7] Evaluation of the Elekta Symmetry™ 4D IGRT System by Using a Moving Lung Phantom
    Shin, Hun-Joo
    Kim, Shin-Wook
    Kay, Chul Seung
    Seo, Jae-Hyuk
    Lee, Gi-Woong
    Kang, Ki-Mun
    Jang, Hong Seok
    Kang, Young-nam
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (01) : 260 - 263
  • [8] A novel deformable lung phantom for 4D radiotherapy verification
    Margeanu, M.
    Stroian, G.
    Heath, E.
    Seuntjens, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2649 - 2649
  • [9] 4D radiotherapy planning strategies for moving targets
    Ma, Y.
    Lee, L.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S616 - S616
  • [10] A 4D Multimodal Lung Phantom for Regmentation Evaluation
    Markerl, D.
    Levesque, I. R.
    El Naqa, I.
    MEDICAL PHYSICS, 2014, 41 (08) : 21 - 21