Characteristics of metal-plate/film detectors at therapy energies. I. Modulation transfer function

被引:16
|
作者
Falco, T
Fallone, BG
机构
[1] McGill Univ, Montreal Gen Hosp, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
[2] McGill Univ, Montreal Gen Hosp, Dept Phys, Montreal, PQ H3G 1A4, Canada
关键词
MTF; metal plates; portal film; therapy detectors;
D O I
10.1118/1.598436
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Measurements of modulation transfer function (MTF) for front and back metal-plate/film portal detectors are reported for the Cobalt-60 and 10 MV spectra. The detectors consist of a double-emulsion portal film secured between plates of Al, Cu, brass, or Pb with thicknesses varying from 0 to 4.81 mm. Secondary electrons produced within the front plate generate the main signal, but the MTF decreases with an increase in front plate thickness greater than the maximum range of electrons R-max because of photon scatter in the front plate. Because the decrease of MTF with backplate thickness ceases for backplate thickness greater than R-max, the MTF is influenced more by the backscatter electrons than the backscatter photons. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)01212-7].