Assessment of newly developed real-time human respiration carbon dioxide measurement device for management of asthma outside of hospital

被引:3
|
作者
Singh, Om Prakash [1 ]
Bin Ahmed, Ismail [2 ]
Malarvili, M. B. [1 ]
机构
[1] Univ Teknol Malaysia, Fac Biosci & Med Engn, Biosignal Proc Res Grp, Skudai 81310, Malaysia
[2] Univ Teknol Malaysia, Pengarah Pusat Kesihatan Mahasiswa, Skudai 81310, Malaysia
关键词
Asthma; capnography; early screening; CO2; measurement; CHILDREN; PREVALENCE; CAPNOGRAM; UTILITY;
D O I
10.3233/THC-181288
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
BACKGROUND: Assessment of asthma outside of the hospital using a patient independent device is highly in demand due to the limitation of existing devices, which are manual and unreliable if patients are not cooperative. OBJECTIVE: The study aims to verify the use of newly developed human respiration, carbon dioxide (CO2) measurement device for the management of asthma outside of the hospital. METHOD: The data were collected from 60 subjects aged between 18-35 years via convenience sampling method reported in UTM Health Center using the device. Furthermore, the data were normalized and analyzed using descriptive statistics, t-test, and area (A(z)) under receiver operating characteristic curve (ROC). RESULT: Findings revealed that the normalized mean values of end-tidal carbon dioxide (EtCO2), Hjorth Activity (HA), and respiratory rate (RR) were lower in asthmatic compared with healthy subjects with minimum deviation from the mean. In addition, each parameter was found to significantly differ statistically for asthma and non-asthma with p < 0.05. Furthermore, the A(z) shows the strong association for the screening of asthma and non-asthma with an average of 0.71 (95% CI: 0.57-0.83), 0.77 (95% CI: 0.64-0.90), and 0.83 (95% CI: 0.73-0.94) for RR, EtCO2, and HA, respectively. CONCLUSIONS: This study demonstrates that the newly developed handheld human respiration CO2 measurement device may possibly be used as an effort-independent asthma management method outside of the hospital.
引用
收藏
页码:785 / 794
页数:10
相关论文
共 22 条
  • [1] Real-time human respiration carbon dioxide measurement device for cardiorespiratory assessment
    Singh, Om Prakash
    Howe, Teo Aik
    Malarvili, M. B.
    JOURNAL OF BREATH RESEARCH, 2018, 12 (02)
  • [2] A new real-time and precision capnography for human respiration carbon dioxide concentration
    Yang, Jiachen
    Wang, Bin
    Xu, Ran
    Wang, Lei
    PERFUSION-UK, 2010, 25 (06): : 399 - 408
  • [3] Performance Characteristics of "Spotter," a Newly Developed Real-Time Wave Measurement Buoy
    Raghukumar, Kaustubha
    Chang, Grace
    Spada, Frank
    Jones, Craig
    Janssen, Tim
    Gans, Andrew
    JOURNAL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY, 2019, 36 (06) : 1127 - 1141
  • [4] Performance Characteristics of Newly Developed Real-Time Wave Measurement Buoy Using the Variometric Approach
    Xue, Chen
    Guo, Jingsong
    Jiang, Shumin
    Wang, Yanfeng
    Guo, Yanliang
    Li, Jie
    JOURNAL OF MARINE SCIENCE AND ENGINEERING, 2024, 12 (11)
  • [5] Real-time measurement of NO concentration in coronary sinus by a newly developed catheter-type NO sensor
    Neishi, Y.
    Mochizuki, S.
    Goto, M.
    Kume, T.
    Renan, S.
    Kawamoto, T.
    Akasaka, T.
    Yoshida, K.
    EUROPEAN HEART JOURNAL, 2005, 26 : 262 - 262
  • [6] Review of the Newly Developed, Mobile Optical Sensors for Real-Time Measurement of the Atmospheric Particulate Matter Concentration
    Molaie, Sama
    Lino, Paolo
    MICROMACHINES, 2021, 12 (04)
  • [7] Real-time bowel sound analysis using newly developed device in patients undergoing gastric surgery for gastric tumor
    Namikawa, Tsutomu
    Yamaguchi, Sachi
    Fujisawa, Kazune
    Ogawa, Maho
    Iwabu, Jun
    Munekage, Masaya
    Uemura, Sunao
    Maeda, Hiromichi
    Kitagawa, Hiroyuki
    Kobayashi, Michiya
    Matsuda, Kenichi
    Hanazaki, Kazuhiro
    JGH OPEN, 2021, 5 (04): : 454 - 458
  • [8] An integrated light management system with real-time light measurement and human perception
    Tsesmelis, T.
    Hasan, I
    Cristani, M.
    Del Bue, A.
    Galasso, F.
    LIGHTING RESEARCH & TECHNOLOGY, 2021, 53 (01) : 74 - 88
  • [9] Photoacoustic heterodyne breath sensor for real-time measurement of human exhaled carbon monoxide
    Li, Biao
    Feng, Chaofan
    Wu, Hongpeng
    Jia, Suotang
    Dong, Lei
    PHOTOACOUSTICS, 2022, 27
  • [10] Photoacoustic heterodyne breath sensor for real-time measurement of human exhaled carbon monoxide
    Li, Biao
    Feng, Chaofan
    Wu, Hongpeng
    Jia, Suotang
    Dong, Lei
    PHOTOACOUSTICS, 2022, 27