Comparison and quantitation of wear debris of failed total hip and total knee arthroplasty

被引:0
|
作者
Hirakawa, K
Bauer, TW
Stulberg, BN
Wilde, AH
机构
[1] CLEVELAND CLIN FDN, DEPT ANAT PATHOL, CLEVELAND, OH 44195 USA
[2] CLEVELAND CLIN FDN, DEPT ORTHOPED SURG, CLEVELAND, OH 44195 USA
[3] CLEVELAND CTR JOINT RECONSTRUCT, CLEVELAND, OH 44115 USA
来源
关键词
D O I
10.1002/(SICI)1097-4636(199606)31:2<257::AID-JBM13>3.0.CO;2-I
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To compare the physical properties of debris particles associated with failed total hip and total knee arthroplasty, we applied a recently developed assay to electronically characterize the size, number, and composition of debris particles isolated from tissues adjacent to failed implants. We identified 21 samples (from 20 patients) of hip synovia and 35 samples (from 32 patients) of knee tissues that had been obtained at the time of revision arthroplasty. There were 12 females and 9 males in the hip group, and 16 females and 19 males in the knee group. Primary arthroplasty was performed for osteoarthritis (OA, 15 cases) or rheumatoid arthritis (RA, 6 cases) in the hip, and for OA (23) or RA (12) in the knee. Patients ranged in age from 23 to 85 (mean 59 years) for total hip, and from 27 to 84 (mean 61 years) for total knee arthroplasty. Implantation duration was from 5 to 123 months (mean 37.8) for total hip, and from 11 to 123 months (mean 63.1) for total knee arthroplasty. All of the implants were composed of cobalt-chromium alloy articulating with ultrahigh-molecular-weight polyethylene. The number of particles smaller than 10 mu m ranged from 1.04 x 10(8)/g to 1.91 x 10(10)/g in the hip, and from 6.69 x 10(8)/g to 2.13 x 10(10)/g in the knee. Energy-dispersive X-ray spectroscopy and polarized light analysis showed both polyethylene and metal particles in most cases. The mean diameter of particles smaller than 10 mu m was 0.72 +/- 0.2 mu m in the hip, and 0.74 +/- 0.1 mu m in the knee. Evaluation of particles larger than 10 mu m showed a larger range of particle size in knee tissues (maximum 6.1 mm, mean 283 mu m), than in the hip tissues (maximum 826 mu m, mean 81 mu m) (p < 0.001). Very small particles are common in both groups, but it appears that a larger range of particle sizes is present adjacent to failed knee than to failed hip prostheses. The higher frequency of large particles in failed knee prostheses probably reflects the perceived higher rate of delamination and fragmentation of tibial and patellar compared to that of acetabular polyethylene. (C) 1996 John Wiley & Sons, Inc.
引用
收藏
页码:257 / 263
页数:7
相关论文
共 50 条
  • [1] Wear debris analysis for total hip and total knee arthroplasty
    Hirakawa, K
    Bauer, TW
    Stulberg, BN
    Yamaguchi, M
    Koshino, T
    Saito, T
    [J]. JOINT ARTHROPLASTY, 1999, : 36 - 45
  • [2] WEAR DEBRIS IN CEMENTED TOTAL HIP-ARTHROPLASTY
    HUO, MH
    SALVATI, EA
    BULY, RL
    [J]. ORTHOPEDICS, 1991, 14 (03) : 335 - 340
  • [3] Characterization and comparison of wear debris from failed total hip implants of different types
    Hirakawa, K
    Bauer, TW
    Stulberg, BN
    Wilde, AH
    Secic, M
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (08): : 1235 - 1243
  • [4] Sciatic neuropathy secondary to total hip arthroplasty wear debris
    Fischer, SR
    Christ, DJ
    Roehr, BA
    [J]. JOURNAL OF ARTHROPLASTY, 1999, 14 (06): : 771 - 774
  • [5] The inflammatory response to particulate wear debris in total hip arthroplasty
    Boynton, EL
    Henry, M
    Morton, J
    Waddell, JP
    [J]. CANADIAN JOURNAL OF SURGERY, 1995, 38 (06) : 507 - 515
  • [6] FAILED TOTAL HIP ARTHROPLASTY
    Monsef, Jad Bou
    Parekh, Amit
    Osmani, Feroz
    Gonzalez, Mark
    [J]. JBJS REVIEWS, 2018, 6 (11) : E3
  • [7] FAILED TOTAL KNEE ARTHROPLASTY
    Monsef, Jad Bou
    Schraut, Nicholas
    Gonzalez, Mark
    [J]. JBJS REVIEWS, 2014, 2 (12)
  • [8] Effects of UHMWPE wear debris generated from total hip and total knee replacements
    Hulbert, SF
    Megremis, SJ
    [J]. PROCEEDINGS OF THE 1996 FIFTEENTH SOUTHERN BIOMEDICAL ENGINEERING CONFERENCE, 1996, : 201 - 206
  • [9] Comparison of Total Hip Arthroplasty after Two Types of Failed Hip Preserving Procedures with Primary Total Hip Arthroplasty
    Zuo, Wei
    Ma, Jin-hui
    Cui, Wei
    Guo, Wan-shou
    Sun, Wei
    [J]. ORTHOPAEDIC SURGERY, 2020, 12 (01) : 162 - 169
  • [10] Wear rate model for UHMWPE in total hip and knee arthroplasty
    Goswami, T.
    Alhassan, S.
    [J]. MATERIALS & DESIGN, 2008, 29 (02): : 289 - 296