First high-pressure XAFS results at the bending-magnet-based energy-dispersive XAFS beamline BL-8 at the Indus-2 synchrotron facility

被引:1
|
作者
Lahiri, Debdutta [1 ]
Dwivedi, Ashutosh [2 ]
Vasanthi, R. [1 ]
Jha, S. N. [2 ]
Garg, Nandini [1 ,3 ]
机构
[1] Bhabha Atom Res Ctr, High Pressure & Synchrotron Radiat Phys Div, Mumbai 400085, Maharashtra, India
[2] Bhabha Atom Res Ctr, Atom & Mol Phys Div, Mumbai 400085, Maharashtra, India
[3] Homi Bhabha Natl Inst, Training Sch Complex, Mumbai 400094, Maharashtra, India
关键词
high-pressure XAFS; energy-dispersive XAFS; RAY-ABSORPTION SPECTROSCOPY; XAS TEXAS FACILITY; DIAMOND-ANVIL CELL; RADIATION FACILITY; EXAFS; SUPERCONDUCTIVITY; TRANSITION; ALLOY; NB; CALIBRATION;
D O I
10.1107/S1600577520006098
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The static focusing optics of the existing energy-dispersive XAFS beamline BL-8 have been advantageously exploited to initiate diamond anvil cell based high-pressure XANES experiments at the Indus-2 synchrotron facility, India. In the framework of the limited photon statistics with the 2.5GeV bending-magnet source, limited focusing optics and 4mm-thick diamond windows of the sample cell, a (non-trivial) beamline alignment method for maximizing photon statistics at the sample position has been designed. Key strategies include the selection of a high X-ray energy edge, the truncation of the smallest achievable focal spot size to target size with a slit and optimization of the horizontal slit position for transmission of the desired energy band. A motor-scanning program for precise sample centering has been developed. These details are presented with rationalization for every step. With these strategies, Nb K-edge XANES spectra for Nb2O5 under high pressure (0-16.9GPa) have been generated, reproducing the reported spectra for Nb2O5 under ambient conditions and high pressure. These first HPXANES results are reported in this paper. The scope of extending good data quality to the EXAFS range in the future is addressed. This work should inspire and guide future high-pressure XAFS experiments with comparable infrastructure.
引用
收藏
页码:988 / 998
页数:11
相关论文
共 4 条
  • [1] First results from the XMCD facility at the Energy-Dispersive EXAFS beamline of the Indus-2 synchrotron source
    Patra, N.
    Sachan, U. G. P. S.
    SundarRajan, S.
    Malhotra, Sanjay
    Harad, Vijay
    Agarwal, Ankur
    Divedi, Ashutosh
    Jha, S. N.
    Bhattacharyya, D.
    JOURNAL OF SYNCHROTRON RADIATION, 2019, 26 : 445 - 449
  • [2] First results from the XMCD facility at the Energy-Dispersive EXAFS beamline of the indus-2 synchrotron source
    Atomic and Molecular Physics Division, Bhabha Atomic Research Centre, Mumbai
    400 085, India
    不详
    400 094, India
    不详
    400 085, India
    J. Synchrotron Radiat., 2019, 2 (445-449):
  • [3] First phase commissioning of high pressure XAFS setup at ED-XAFS beamline, Indus-2 synchrotron radiation source, India
    Ramanan N.
    Kumar A.
    Rajput P.
    Thankarajan K.
    Bhattacharyya D.
    Jha S.N.
    Lahiri D.
    J. Opt., 2 (182-194): : 182 - 194
  • [4] A horizontal two-axis diffractometer for high-energy X-ray diffraction using synchrotron radiation on bending magnet beamline BL04B2 at SPring-8
    Kohara, S
    Suzuya, K
    Kashihara, Y
    Matsumoto, N
    Umesaki, N
    Sakai, I
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 467 : 1030 - 1033