Trends in cup position utilizing computer-assisted navigation during total hip arthroplasty: A retrospective observational study

被引:0
|
作者
Mays, Rachel R. [1 ]
Benson, Jessica R. [1 ]
Muir, Jeffrey M. [1 ]
Meftah, Morteza [2 ]
机构
[1] Intellijoint Surg, Dept Clin Res, Waterloo, ON, Canada
[2] NYU Langone Hlth, NYU Langone Orthoped Hosp, Div Adult Reconstruct Surg, New York, NY 10003 USA
来源
CURRENT ORTHOPAEDIC PRACTICE | 2020年 / 31卷 / 03期
关键词
total hip arthroscopy; acetabular cup; safe zone; computer-assisted navigation system; ACETABULAR COMPONENT POSITION; SAFE ZONE; DISLOCATION; REPLACEMENT; PLACEMENT; ACCURACY; REVISION; RISK; WEAR; EPIDEMIOLOGY;
D O I
10.1097/BCO.0000000000000869
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Proper positioning of the acetabular cup deters dislocation after total hip arthroplasty (THA) and is therefore a key focus for orthopaedic surgeons. The concept of a safe zone for acetabular component placement remains widely utilized in contemporary THA practice; however, components positioned in this safe zone still dislocate. The purpose of this study was to characterize trends in acetabular cup positioning and to determine if the surgical approaches used affected the rate at which the components were placed in the safe zones. Methods: This retrospective, observational study reviewed acetabular cup position in 1236 patients who underwent THA using computer-assisted navigation. Outcomes included acetabular cup position measurements derived from a surgical navigation device and surgical approach. Results: The overall mean cup position of all recorded cases was 21.8 degrees (+/- 7.7 degrees, 95% CI=6.7 degrees, 36.9 degrees) of anteversion and 40.9 degrees (+/- 6.5 degrees, 95% CI=28.1 degrees, 53.7 degrees) of inclination. For anteversion and inclination, 65.5% (809/1236) of acetabular cup components were within the Lewinnek safe zone, and 58.4% (722/1236) were within the Callanan safe zone. Acetabular cups were placed at a mean of 6.8 degrees of anteversion (posterior/lateral approach: 7.0 degrees, anterior approach: 5.6 degrees) higher than the Lewinnek and Callanan safe zones, whereas inclination was positioned 0.9 degrees higher than the reported Lewinnek safe zone and 3.4 degrees higher than the Callanan safe zone. Conclusions: Our data show that while most acetabular cups were placed within the traditional safe zones, the mean anteversion orientation is considerably higher than those suggested by the Lewinnek and Callanan safe zones. The implications of this observation warrant further investigation.
引用
收藏
页码:211 / 217
页数:7
相关论文
共 50 条
  • [1] Accuracy of computer-assisted cup placement in total hip arthroplasty
    Jolles, BM
    Genoud, P
    Hoffmeyer, P
    [J]. CARS 2001: COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2001, 1230 : 301 - 305
  • [2] Clinical Validation of Computer-Assisted Navigation in Total Hip Arthroplasty
    Beaumont, Eric
    Beaumont, Pierre
    Odermat, Daniel
    Fontaine, Isabelle
    Jansen, Herbert
    Prince, Francois
    [J]. ADVANCES IN ORTHOPEDICS, 2011, 2011
  • [3] Limitations of Imageless Computer-Assisted Navigation for Total Hip Arthroplasty
    Lin, Fang
    Lim, Dohyung
    Wixson, Richard L.
    Milos, Steven
    Hendrix, Ronald W.
    Makhsous, Mohsen
    [J]. JOURNAL OF ARTHROPLASTY, 2011, 26 (04): : 596 - 605
  • [4] Total Hip Arthroplasty Using Imageless Computer-Assisted Hip Navigation A Prospective Randomized Study
    Lass, Richard
    Kubista, Bernd
    Olischar, Boris
    Frantal, Sophie
    Windhager, Reinhard
    Giurea, Alexander
    [J]. JOURNAL OF ARTHROPLASTY, 2014, 29 (04): : 786 - 791
  • [5] Reliability and Correlation of the Force-PRO Device and Computer-Assisted Navigation System for Measurement of Acetabular Cup Position in Total Hip Arthroplasty
    Vijittrakarnrung, Chaiyanun
    Manupibul, Udomporn
    Charoensuk, Warakorn
    Wongsak, Siwadol
    Sa-Ngasoongsong, Paphon
    [J]. SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2020, 36
  • [6] Total hip arthroplasty in osteopetrosis using computer-assisted fluoroscopic navigation
    Egawa, H
    Nakano, S
    Hamada, D
    Sato, R
    Yasui, N
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (08): : 1074 - 1077
  • [7] Evaluating Alternate Registration Planes for Imageless, Computer-Assisted Navigation During Total Hip Arthroplasty
    Vigdorchik, Jonathan M.
    Sculco, Peter K.
    Inglis, Allan E.
    Schwarzkopf, Ran
    Muir, Jeffrey M.
    [J]. JOURNAL OF ARTHROPLASTY, 2021, 36 (10): : 3527 - 3533
  • [8] Computer-Assisted Navigation in Total Knee Arthroplasty
    Mathew, Kevin K.
    DeGouveia, William
    Marchand, Kevin B.
    Ehiorobo, Joseph O.
    Tarazi, John M.
    Sodhi, Nipun
    Salem, Hytham S.
    Mont, Michael A.
    [J]. SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2020, 36
  • [9] Computer-assisted navigation in total knee arthroplasty
    de Zwart, P. M.
    Weise, K.
    Ochs, U.
    [J]. TRAUMA UND BERUFSKRANKHEIT, 2010, 12 (02) : 107 - 112
  • [10] Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement
    Jolles, BM
    Genoud, P
    Hoffmeyer, P
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (426) : 174 - 179