Three-dimensional power Doppler imaging: A phantom study to quantify vessel stenosis

被引:58
|
作者
Guo, ZY
Fenster, A
机构
[1] JOHN P ROBARTS RES INST, IMAGING RES LABS, LONDON, ON N6A 5K8, CANADA
[2] UNIV WESTERN ONTARIO, DEPT MED BIOPHYS, LONDON, ON, CANADA
来源
ULTRASOUND IN MEDICINE AND BIOLOGY | 1996年 / 22卷 / 08期
关键词
power Doppler imaging; three-dimensional imaging; vessel stenosis;
D O I
10.1016/S0301-5629(96)00125-1
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This study investigated whether three-dimensional (3D) power Doppler imaging can be used to quantify arterial stenosis and its potential as an alternative to x-ray angiography. Three-dimensional power Doppler images of in vitro stenotic vessels were generated under different hemodynamic conditions with a 3D power Doppler imaging system, This system includes: a Macintosh Quadra 840AV computer used to perform 3D imaging acquisition, reconstruction and display; a computer-controlled motor-driven translation assembly used to move the transducer; and an ATL Ultramark 9 HDI ultrasound system, Three vascular- and tissue-mimicking phantoms containing three wall-less stenotic vessels with area reduction of 80%, 50% and 30% were imaged with different flow rates under both steady and pulsatile flow conditions and with different Doppler angles under steady flow condition, With the use of the blood mimic, experimental results demonstrated that power Doppler imaging is nearly independent on flow velocity and Doppler angle, It was also demonstrated that 3D power Doppler imaging can produce nonpulsatile angiographic-like 3D images of the flow field, The stenotic vessels were quantified with an overall accuracy of 8.3% of the vessel area and an overall precision of 7% of the vessel area under the conditions described in this paper, It is believed that 3D power Doppler imaging can be used to quantify arterial stenosis, and in some applications it could be an alternative to x-ray angiography. Copyright (C) 1996 World Federation for Ultrasound in Medicine & Biology
引用
收藏
页码:1059 / 1069
页数:11
相关论文
共 50 条
  • [1] Abdominal three-dimensional power Doppler imaging
    von Herbay, A
    Häussinger, D
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 2001, 20 (02) : 151 - 157
  • [2] Three-dimensional power Doppler imaging of the Fallopian tube
    Sladkevicius, P
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 1999, 13 (04) : 287 - 287
  • [3] Three-dimensional color power Doppler as an imaging technique in cancer
    Shi, Huijie
    Wang, Hongling
    Wang, Ying
    Zhao, Hui
    Xv, Na
    Liu, Feng
    Peng, Xiangwen
    [J]. INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2016, 9 (06): : 9781 - 9790
  • [4] Three-dimensional power Doppler imaging of vertebrobasilar circulation in adults
    Koga, M
    Kimura, K
    Yasaka, M
    Otsubo, R
    Hasegawa, Y
    Minematsu, K
    Yamaguchi, T
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 1999, 20 (05) : 943 - 944
  • [5] Manufacture of Flow Phantom with Stenosis and Imaging Evaluation of Power Doppler
    Park, Hee-Young
    Bae, Jong-Rim
    Kim, Jeong-Koo
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF KOREA, 2009, 28 (08): : 732 - 739
  • [6] Evaluation of the volume vascularisation index using three-dimensional Power Doppler ultrasonography: A phantom study.
    Wijman, MJ
    Struijk, PC
    Wladimiroff, JW
    Steegers, EA
    [J]. JOURNAL OF THE SOCIETY FOR GYNECOLOGIC INVESTIGATION, 2004, 11 (02) : 223A - 223A
  • [7] Three-dimensional power Doppler in study of embryonic vasculogenesis
    Kurjak, A
    Hafner, T
    Kupesic, S
    Kostovic, L
    [J]. JOURNAL OF PERINATAL MEDICINE, 2002, 30 (01) : 18 - 25
  • [8] Three-dimensional phantom imaging for photoacoustic mammography
    Kharine, A
    Manohar, S
    Steenbergen, W
    [J]. JOURNAL DE PHYSIQUE IV, 2005, 125 : 713 - 715
  • [9] Three-dimensional power Doppler angiography
    Guo, ZY
    Durand, LG
    Holdsworth, D
    Fenster, A
    [J]. PHYSIOLOGY AND FUNCTION FROM MULTIDIMENSIONAL IMAGES - MEDICAL IMAGING 1997, 1997, 3033 : 215 - 224
  • [10] Three-dimensional power Doppler imaging in the assessment of Fallopian tube patency
    Sladkevicius, P
    Ojha, K
    Campbell, S
    Nargund, G
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2000, 16 (07) : 644 - 647