Oblique needle segmentation and tracking for 3D TRUS guided prostate brachytherapy

被引:25
|
作者
Wei, ZP
Gardi, L
Downey, DB
Fenster, A
机构
[1] John P Robarts Res Inst, Imaging Res Labs, London, ON N6A 5K8, Canada
[2] Univ Western Ontario, Biomed Engn Program, London, ON N6A 5B9, Canada
[3] Univ Western Ontario, Dept Radiol & Nucl Med, London, ON N6A 5A5, Canada
基金
加拿大健康研究院;
关键词
ultrasound-guided brachytherapy; 3D ultrasound-guided brachytherapy; prostate cancer; prostate brachytherapy; needle segmentation;
D O I
10.1118/1.2011108
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An algorithm was developed in order to segment and track brachytherapy needles inserted along oblique trajectories. Three-dimensional (3D) transrectal ultrasound (TRUS) images of the rigid rod simulating the needle inserted into the tissue-mimicking agar and chicken breast phantoms were obtained to test the accuracy of the algorithm under ideal conditions. Because the robot possesses high positioning and angulation accuracies, we used the robot as a "gold standard," and compared the results of algorithm segmentation to the values measured by the robot. Our testing results showed that the accuracy of the needle segmentation algorithm depends on the needle insertion distance into the 3D TRUS image and the angulations with respect to the TRUS transducer, e.g., at a 10 degrees insertion anglulation in agar phantoms, the error of the algorithm in determining the needle tip position was less than I mm when the insertion distance was greater than 15 mm. Near real-time needle tracking was achieved by scanning a small volume containing the needle. Our tests also showed that, the segmentation time was less than 60 ms, and the scanning time was less than 1.2 s, when the insertion distance into the 3D TRUS image was less than 55 mm. In our needle tracking tests in chicken breast phantoms, the errors in determining the needle orientation were less than 2 degrees in robot yaw and 0.7 degrees in robot pitch orientations, for up to 20 degrees needle insertion angles with the TRUS transducer in the horizontal plane when the needle insertion distance was greater than 15 mm. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:2928 / 2941
页数:14
相关论文
共 50 条
  • [1] Oblique needle segmentation for 3D trus-guided robot-aided transperineal prostate brachytherapy
    Wei, ZP
    Gardi, L
    Downey, DB
    Fenster, A
    [J]. 2004 2ND IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1 AND 2, 2004, : 960 - 963
  • [2] Automatic Shape-Based 3D Level Set Segmentation for Needle Tracking in 3D TRUS Guided Prostate Brachytherapy
    Yan, P.
    Chao, K.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [3] 3D TRUS image segmentation in prostate brachytherapy
    Ding, Mingyue
    Gardi, Lori
    Wei, Zhouping
    Fenster, Aaron
    [J]. 2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 7170 - 7173
  • [4] 3D TRUS guided robot assisted prostate brachytherapy
    Wei, ZP
    Ding, MY
    Downey, D
    Fenster, A
    [J]. MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2005, PT 2, 2005, 3750 : 17 - 24
  • [5] AUTOMATIC SHAPE-BASED LEVEL SET SEGMENTATION FOR NEEDLE TRACKING IN 3-D TRUS-GUIDED PROSTATE BRACHYTHERAPY
    Yan, Ping
    Cheeseborough, John C., III
    Chao, K. S. Clifford
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2012, 38 (09): : 1626 - 1636
  • [6] Needle segmentation using 3D Hough transform in 3D TRUS guided prostate transperineal therapy
    Qiu, Wu
    Ming Yuchi
    Ding, Mingyue
    Tessier, David
    Fenster, Aaron
    [J]. MEDICAL PHYSICS, 2013, 40 (04)
  • [7] Robotic aided 3D TRUS guided intraoperative prostate brachytherapy
    Wei, ZP
    Wan, G
    Gardi, L
    Downey, DB
    Fenster, A
    [J]. MEDICAL IMAGING 2004: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2004, 5367 : 361 - 370
  • [8] Needle steering for 3D TRUS-guided robot-aided brachytherapy
    Wan, G
    Wei, ZP
    Gardi, L
    Downey, D
    Fenster, A
    [J]. CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 1325 - 1325
  • [9] Segmentation of implanted radioactive seeds in 3D trus images for intraoperative prostate brachytherapy
    Wei, Zhouping
    Gardi, Lori
    Downey, Donal B.
    Fenster, Aaron
    [J]. 2006 3RD IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1-3, 2006, : 876 - +
  • [10] Adaptive Metal Artifacts Correction in 3D TRUS Image Guided Prostate Brachytherapy
    Yan, P.