A case of appropriate inappropriate device therapy: Hyperkalemia-induced ventricular oversensing

被引:4
|
作者
Oudit, Gavin Y. [1 ]
Cameron, Doug [1 ]
Harris, Louise [1 ]
机构
[1] Univ Toronto, Toronto Gen Hosp, Univ Hlth Network, Dept Med, Div Cardiol, Toronto, ON, Canada
关键词
Electrogram fractionation; Hyperkalemia; Inappropriate shock;
D O I
10.1016/S0828-282X(08)70592-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The present case describes a patient who received inappropriate, but potentially life-saving, therapy from her implantable cardioverter defibrillator (ICD) in the setting of acute hyperkalemia (plasma potassium concentration = 8 mM). Hyperkalemia was associated with the development of a slow sinusoidal ventricular tachycardia, at a rate of 100 beats/min to 125 beats/min (610 ms to 480 ms) in a patient who is pacemaker-dependent. There was associated fractionation of the ICD electrogram and T wave oversensing, leading to ventricular oversensing with resultant detection in the ventricular fibrillation rate zone. This was followed by shock therapy, even though the ventricular tachycardia rate was below the programmed detection rate of the ICD. The subsequent emergency treatment of the hyperkalemia normalized the electrogram, corrected the ventricular oversensing and arrhythmia, and restored rate-adaptive single-chamber ventricular pacing.
引用
收藏
页码:E16 / E18
页数:3
相关论文
共 50 条
  • [1] Hyperkalemia-Induced T Wave Oversensing in an Implantable Cardioverter Defibrillator
    Gonzalez, Ariel
    Franqui, Hilton
    Lopez, Jose
    Banchs, Hector
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (02)
  • [2] A case of inappropriate implantable cardioverter defibrillator therapy induced by T-wave oversensing due to hyperkalemia
    Aoki, Kosuke
    Okajima, Katsunori
    Kiuchi, Kunihiko
    Yokoi, Kiminobu
    Teranishi, Jin
    Shimane, Akira
    JOURNAL OF ARRHYTHMIA, 2015, 31 (06) : 395 - 397
  • [3] Hyperkalemia-induced Brugada phenocopy: a systematic review of case reports
    Alnajjar, Asmaa Zakria
    Ibrahim, Afnan Ismail
    Ellebedy, Mohamed
    CLINICAL RESEARCH IN CARDIOLOGY, 2025,
  • [4] Hyperkalemia induced T wave oversensing leading to loss of biventricular pacing and inappropriate ICD shocks
    Koul, AK
    Keller, S
    Clancy, JF
    Lampert, R
    Batsford, WP
    Rosenfeld, LE
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (05): : 681 - 683
  • [5] Hyperkalemia-induced conversion of permanent atrial fibrillation to normal sinus rhythm: a case report
    Romeo, Emanuele
    D'Alto, Michele
    Santarpia, Giuseppe
    Di Nardo, Giangiacomo
    Sarubbi, Berardo
    Russo, Maria Giovanna
    Calabro, Raffaele
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2011, 12 (09) : 678 - 680
  • [6] Shock-induced ventricular oversensing due to seat plug damage: A potential mechanism of inappropriate device therapies in implantable cardioverter-defibrillators
    Cheung, JW
    Iwai, S
    Lerman, BB
    Mittal, S
    HEART RHYTHM, 2005, 2 (12) : 1371 - 1375
  • [7] HYPERKALEMIA-INDUCED CONVERSION OF CHRONIC ATRIAL-FIBRILLATION TO NORMAL SINUS RHYTHM - CASE-REPORT
    MATHEW, PK
    ANGIOLOGY, 1979, 30 (02) : 143 - 146
  • [8] A Rare type of Ventricular Oversensing in ICD Therapy-Inappropriate ICD Shock Delivery Due to Triple Counting
    Guenther, Michael
    Rauwolf, Thomas P.
    Bock, Manja
    Strasser, Ruth H.
    Braun, Martin U.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2010, 33 (02): : e17 - e19
  • [9] PREDICTORS OF VENTRICULAR ARRHYTHMIA AND APPROPRIATE DEVICE THERAPY IN TRANSTHYRETIN CARDIAC AMYLOIDOSIS
    Patel, Rahul
    Steinberg, Rebecca S.
    Rim, Austin
    Yalamanchili, Sreeram
    Brown, Matthew T.
    Bhatt, Kunal
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 81 (08) : 643 - 643
  • [10] FATAL CASE OF INAPPROPRIATE ADH SECRETION INDUCED BY CYCLOPHOSPHAMIDE THERAPY
    HARLOW, PJ
    DECLERCK, YA
    SHORE, NA
    ORTEGA, JA
    CARRANZA, A
    HEUSER, E
    CANCER, 1979, 44 (03) : 896 - 898