Evaluation of the new red cell parameters on Beckman Coulter DxH800 in distinguishing iron deficiency anaemia from thalassaemia trait

被引:22
|
作者
Ng, E. H. Y. [1 ]
Leung, J. H. W. [1 ]
Lau, Y. S. [1 ]
Ma, E. S. K. [1 ]
机构
[1] Hong Kong Sanat & Hosp, Dept Pathol, Happy Valley, Hong Kong, Peoples R China
关键词
Red blood cell size factor; low haemoglobin density; microcytic anaemia factor; thalassaemia carrier; iron deficiency anaemia; RETICULOCYTE HEMOGLOBIN EQUIVALENT; SYSMEX XE 5000; DIFFERENTIAL-DIAGNOSIS; AUTOMATED MEASUREMENT; CLINICAL UTILITY; INDEXES; SUBPOPULATIONS; DENSITY; MARKER;
D O I
10.1111/ijlh.12262
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionThe new red blood cell (RBC) parameters such as reticulocyte haemoglobin content and percentage of hypochromic red cells or equivalent, although useful in the laboratory assessment of iron deficiency anaemia (IDA), are confounded by thalassaemia trait (TT). We aim to evaluate the new red cell parameters on the Beckman Coulter DxH800 in distinguishing between IDA and TT. MethodsA total of 246 normal subjects, 102 patients with IDA and 115 subjects with TT were accrued for the study. The parameters studied were red blood cell size factor (RSF), low haemoglobin density (LHD%), microcytic anaemia factor (MAF), standard deviation of conductivity of the nonreticulocyte population (SD-C-NRET) and unghosted cell (UGC). Comparison between groups was performed by Student's t-test, and the diagnostic performance was determined by receiver operating characteristic (ROC) curve analysis. ResultsBoth the LHD% and RSF were significantly higher in IDA than TT, whereas MAF and SD-C-NRET were significantly lower. The SD-C-NRET showed the best diagnostic performance as a single parameter. A formula, [(RBC+Hb)x(HCT+SD-C-NRET)]/RDW-SD, was devised to distinguish between IDA and TT. With a cut-off value of 23, the area under the curve (AUC) was 0.995 (95% CI of 0.99-1.00), the sensitivity was 97%, and the specificity was 99.1%. ConclusionsThe new RBC parameters on Beckman Coulter DxH800 provide useful information in distinguishing between IDA and TT, which is important for clinical decision-making and for streamlining laboratory testing. A new formula is devised that performs better than other discriminant functions in the literature.
引用
收藏
页码:199 / 207
页数:9
相关论文
共 50 条
  • [1] EVALUATION OF NEWRED CELL PARAMETERS ON BECKMAN COULTER DXH800IN THE DETECTION OF IRON DEFICIENCY AND THALASSAEMIA CARRIER
    Ng, Esda H. Y.
    Leung, Junior H. W.
    Lau, Y. S.
    Ma, Edmond S. K.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2013, 35 : 56 - 56
  • [2] EVALUATION STUDY OF DXH500 HEMATOLOGY ANALYZER AND COMPARISON WITH BECKMAN COULTER DXH800
    Nieto, Javier
    Tejedor, Xavier
    Torres, Silvia
    Morales-Indiano, Cristian
    Martinez, Alicia
    Cabrera, Cristina
    Angels Sala, Ma
    Jimenez, Orlando
    Antonia Llopis, Ma
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 : 85 - 85
  • [3] Evaluation of the Beckman-Coulter DxH800 for Use in a Cord Blood Bank Setting
    Johnson, Phillip
    [J]. TRANSFUSION, 2017, 57 (07) : 7A - 8A
  • [4] FLOW CYTOMETRIC VALIDATION OF CORRECTED BECKMAN COULTER DXH800 WHITE CELL COUNTS
    Andriolo, Kerrie
    Naumovski, Jovanco
    Stapleton, Jennifer
    Crispin, Philip
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 : 15 - 15
  • [5] SCREENING FOR SEPSIS WITH CELL POPULATION DATA ON BECKMAN COULTER DXH800 HEMATOLOGY ANALYSER
    Shaikh, Nahid Anis
    Omar, Zulfa
    Ali, Akbar
    Douglas, Melvyn
    Daniel, Kashif
    Belal, Sana
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 95 - 95
  • [6] SCREENING FOR MALARIA WITH CELL POPULATION DATA ON BECKMAN COULTER DXH800 HEMATOLOGY ANALYZER.
    Shaikh, Nahid Anis
    Saheb, Akbar Ali Usman
    Douglas, Melvyn
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 94 - 95
  • [7] THE ROLE OF CELL POPULATION DATA ON BECKMAN COULTER DXH-800 IN DIFFERENTIAL DIAGNOSIS OF IRON DEFICIENCY ANEMIA AND THALASSEMIA TRAIT.
    Morales-Indiano, Cristian
    Rodriguez, Ines
    Jimenez, Orlando
    Serrando, Maite
    Tejedor, Xavier
    Martinez, Alicia
    Iglesias, Eva
    Sala, Angels
    Duenas, Ana
    Milla, Fuensanta
    Llopis, Maantonia
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 75 - 76
  • [8] DETECTING IRON DEFICIENCY IN PREGNANCY WITH UNICEL DXH800 LHD AND MAF PARAMETERS
    Casolari, Barbara
    Cenci, Anna Maria
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 109 - 109
  • [9] Discriminant Red Cell Indices for Microcytic Hypochromic Anaemia in Distinguishing Beta Thalassaemia Trait and Iron Deficiency Anaemia: A Systematic Review
    Sain, Ankita
    Bhake, Arvind
    Agrawal, Anil
    Thomas, Sophia
    [J]. JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2021, 15 (01)
  • [10] Evaluation of the acceptable measurement range of nucleated red blood cell on Beckman Coulter UniCel DxH 800 analyser
    Zhu, W. -G.
    Pan, F. -G.
    Zhang, Z. -H.
    Zhang, J.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 (05) : E87 - E89