Effect of magnetic field fluctuation on ultra-low field MRI measurements in the unshielded laboratory environment

被引:4
|
作者
Liu, Chao [1 ,3 ,4 ]
Chang, Baolin [1 ,3 ,4 ]
Qiu, Longqing [1 ,3 ,4 ]
Dong, Hui [1 ,3 ,4 ]
Qiu, Yang [1 ,3 ,4 ]
Zhang, Yi [2 ,3 ,4 ]
Krause, Hans-Joachim [2 ,3 ,4 ]
Offenhaeusser, Andreas [2 ,3 ,4 ]
Xie, Xiaoming [1 ,3 ,4 ]
机构
[1] Chinese Acad Sci, SIMIT, State Key Lab Funct Mat Informat, Shanghai 200050, Peoples R China
[2] Forschungszentrum Julich, Peter Grunberg Inst PGI 8, D-52425 Julich, Germany
[3] Collaborat CAS Shanghai, Joint Res Lab Superconduct & Bioelect, Shanghai 200050, Peoples R China
[4] FZJ, D-52425 Julich, Germany
基金
中国国家自然科学基金; 上海市自然科学基金;
关键词
dc SQUID; Ultra-low field MRI; Magnetic field fluctuation; MICROTESLA MRI; HUMAN BRAIN;
D O I
10.1016/j.jmr.2015.04.014
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Magnetic field fluctuations in our unshielded urban laboratory can reach hundreds of nT in the noisy daytime and is only a few nT in the quiet midnight. The field fluctuation causes the Larmor frequency f(L) to drift randomly for several Hz during the unshielded ultra-low field (ULF) nuclear magnetic resonance (NMR) and magnetic resonance imaging (MRI) measurements, thus seriously spoiling the averaging effect and causing imaging artifacts. By using an active compensation (AC) technique based on the spatial correlation of the low-frequency magnetic field fluctuation, the field fluctuation can be suppressed to tens of nT, which is a moderate situation between the noisy daytime and the quiet midnight. In this paper, the effect of the field fluctuation on ULF MRI measurements was investigated. The 1D and 2D MRI signals of a water phantom were measured using a second-order low-T-c superconducting quantum interference device (SQUID) in three fluctuation cases: severe fluctuation (noisy daytime), moderate fluctuation (daytime with AC) and minute fluctuation (quiet midnight) when different gradient fields were applied. When the active compensation is applied or when the frequency encoding gradient field G(x) reaches a sufficiently strong value in our measurements, the image artifacts become invisible in all three fluctuation cases. Therefore it is feasible to perform ULF-MRI measurements in unshielded urban environment without imaging artifacts originating from magnetic fluctuations by using the active compensation technique and/or strong gradient fields. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:8 / 14
页数:7
相关论文
共 50 条
  • [1] Ultra-low field magnetic resonance imaging detection with gradient tensor compensation in urban unshielded environment
    Dong, Hui
    Qiu, Longqing
    Shi, Wen
    Chang, Baolin
    Qiu, Yang
    Xu, Lu
    Liu, Chao
    Zhang, Yi
    Krause, Hans-Joachim
    Offenhaeusser, Andreas
    Xie, Xiaoming
    [J]. APPLIED PHYSICS LETTERS, 2013, 102 (10)
  • [2] Adaptive suppression of power line interference in ultra-low field magnetic resonance imaging in an unshielded environment
    Huang, Xiaolei
    Dong, Hui
    Qiu, Yang
    Li, Bo
    Tao, Quan
    Zhang, Yi
    Krause, Hans-Joachim
    Offenhaeusser, Andreas
    Xie, Xiaoming
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2018, 286 : 52 - 59
  • [3] Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment
    Matlashov, Andrei
    Magnelind, Per
    Newman, Shaun
    Sandin, Henrik
    Urbaitis, Algis
    Volegov, Petr
    Espy, Michelle
    [J]. 2015 15TH INTERNATIONAL SUPERCONDUCTIVE ELECTRONICS CONFERENCE (ISEC), 2015,
  • [4] Quantum oscillation measurements in high magnetic field and ultra-low temperature
    王瀑
    李岗
    雒建林
    [J]. Chinese Physics B, 2018, (07) : 91 - 95
  • [5] Quantum oscillation measurements in high magnetic field and ultra-low temperature
    Wang, Pu
    Li, Gang
    Luo, Jian-Lin
    [J]. CHINESE PHYSICS B, 2018, 27 (07)
  • [6] The Effect of Low Frequency External Field Disturbance on the SQUID Based Ultra-Low Field NMR Measurements
    Jin, Y. R.
    Wang, N.
    Li, S.
    Tian, Y.
    Ren, Y. F.
    Wu, Y. L.
    Deng, H.
    Chen, Y. F.
    Li, J.
    Tian, H. Y.
    Chen, G. H.
    Zheng, D. N.
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2011, 21 (03) : 518 - 521
  • [7] Integrated LTS gradiometer SQUIDs for measurements of magnetic field distributions in an unshielded environment
    Stolz, R
    Schultze, V
    Zakosarenko, VM
    Fritzsch, L
    Meyer, HG
    [J]. CZECHOSLOVAK JOURNAL OF PHYSICS, 1996, 46 : 2769 - 2770
  • [8] SQUIDs for Magnetic Resonance Imaging at Ultra-low Magnetic Field
    Matlashov, A. N.
    Zotev, V. S.
    Kraus, R. H., Jr.
    Sandin, H.
    Urbaitis, A. V.
    Volegov, P. L.
    Espy, M. A.
    [J]. PIERS 2009 MOSCOW VOLS I AND II, PROCEEDINGS, 2009, : 802 - 806
  • [9] ULTRA-LOW FIELD MAGNETIC STUDIES OF CERAMIC SUPERCONDUCTORS
    GOUGH, CE
    KEENE, MN
    MEE, C
    RAE, AIM
    ABELL, SJ
    [J]. PHYSICA C, 1988, 153 : 1501 - 1502
  • [10] Ultra-low noise magnetic field for quantum gases
    Xu, Xiao-Tian
    Wang, Zong-Yao
    Jiao, Rui-Heng
    Yi, Chang-Rui
    Sun, Wei
    Chen, Shuai
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (05):