Correction for oral contrast artifacts in CT attenuation-corrected PET images obtained by combined PET/CT

被引:0
|
作者
Nehmeh, SA
Erdi, YE
Kalaigian, H
Kolbert, KS
Pan, TS
Yeung, H
Squire, O
Sinha, A
Larson, SM
Humm, JL
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
[2] Gen Elect Med Syst, Appl Sci Lab, Waukesha, WI USA
[3] Mem Sloan Kettering Canc Ctr, Dept Radiol, New York, NY 10021 USA
关键词
PET/CT contrast artifacts;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Recent studies have shown increased artifacts in CT attenuation-corrected (CTAC) PET images acquired with oral contrast agents because of misclassification of contrast as bone. We have developed an algorithm, segmented contrast correction (SCC), to properly transform CT numbers in the contrast regions from CT energies (40-140 keV) to PET energy at 511 keV. Methods: A bilinear transformation, equivalent to that supplied by the PET/CT scanner manufacturer, for the conversion of linear attenuation coefficients of normal tissues from CT to PET energies was optimized for BaSO4 contrast agent. This transformation was validated by comparison with the linear attenuation coefficients measured for BaSO4 at concentrations ranging from 0% to 80% at 511 keV for PET transmission images acquired with Ge-68 rod sources. In the CT images, the contrast regions were contoured to exclude bony structures and then segmented on the basis of a minimum threshold CT number (300 Hounsfield units). The CT number in each pixel identified with contrast was transformed into the corresponding effective bone CT number to produce the correct attenuation coefficient when the data were translated by the manufacturer software into PET energy during the process of CT attenuation correction. CT images were then used for attenuation correction of PET emission data. The algorithm was validated with a phantom in which a lesion was simulated within a volume of BaSO4 contrast and in the presence of a human vertebral bony structure. Regions of interest in the lesion, bone, and contrast on emission PET images reconstructed with and without the SCC algorithm were analyzed. The results were compared with those for images obtained with Ge-68-based transmission attenuation-corrected PET. Results: The SCC algorithm was able to correct for contrast artifacts in CTAC PET images. In the phantom studies, the use of SCC resulted in an approximate 32% reduction in the apparent activity concentration in the lesion compared with data obtained from PET images without SCC and a <7.6% reduction compared with data obtained from Ge-68-based attenuation-corrected PET images. In one clinical study, maximum standardized uptake value (SUVmax) measurements for the lesion, bladder, and bowel were, respectively, 14.52, 13.63, and 13.34 g/mL in CTAC PET images, 59.45, 26.71, and 37.22 g/mL in Ge-68-based attenuation-corrected PET images, and 11.05, 6.66, and 6.33 g/mL in CTAC PET images with SCC. Conclusion: Correction of oral contrast artifacts in PET images obtained by combined PET/CT yielded more accurate quantitation of the lesion and other, normal structures. The algorithm was tested in a clinical case, in which SUVmax measurements showed discrepancies of 2%, 1.3%, and 5% between Ge-68-based attenuation-corrected PET images and CTAC PET images with SCC for the lesion, bladder, and bowel, respectively. These values correspond to 6.5%, 62%, and 66% differences between CTAC-based measurements and Ge-68-based ones.
引用
收藏
页码:1940 / 1944
页数:5
相关论文
共 50 条
  • [1] Segmented contrast correction (SCC): An algorithm to correct for oral contrast artifacts in the CT attenuation corrected PET images.
    Nehmeh, SA
    Erdi, YE
    Kalaigian, H
    Pan, T
    Yeung, HW
    Larson, SM
    Ling, CC
    Humm, JL
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 269P - 270P
  • [2] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Medhat M. Osman
    Christian Cohade
    Yuji Nakamoto
    Richard L. Wahl
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 603 - 606
  • [3] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT.
    Osman, MM
    Cohade, C
    Wahl, RL
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 305P - 305P
  • [4] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Osman, MM
    Cohade, C
    Nakamoto, Y
    Wahl, RL
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (04) : 603 - 606
  • [5] PET recognition of pulmonary metastases on PET/CT imaging: impact of attenuation-corrected and non-attenuation-corrected PET images
    Reinhardt, MJ
    Wiethoelter, N
    Matthies, A
    Joe, AY
    Strunk, H
    Jaeger, U
    Biersack, HJ
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 (02) : 134 - 139
  • [6] PET recognition of pulmonary metastases on PET/CT imaging: impact of attenuation-corrected and non-attenuation-corrected PET images
    Michael J. Reinhardt
    Nicole Wiethoelter
    Alexander Matthies
    Alexius Y. Joe
    Holger Strunk
    Ursula Jaeger
    Hans-Juergen Biersack
    European Journal of Nuclear Medicine and Molecular Imaging, 2006, 33 : 134 - 139
  • [7] Comparison of PET image quality between germanium transmission attenuation-corrected images and combined PET/CT images.
    Yoshikawa, K
    Sagou, K
    Tomemori, T
    Tamura, K
    Yamada, M
    Suzuki, K
    Tanada, S
    Tsujii, H
    Murata, H
    Sasaki, Y
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 378P - 379P
  • [8] Standardised uptake values from PET/CT images: comparison with conventional attenuation-corrected PET
    Souvatzoglou, M.
    Ziegler, S. I.
    Martinez, M. J.
    Busch, R.
    Dzewas, G.
    Schwaiger, M.
    Bengel, F.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 (03) : 405 - 412
  • [9] Standardised uptake values from PET/CT images: comparison with conventional attenuation-corrected PET
    M. Souvatzoglou
    S. I. Ziegler
    M. J. Martinez
    R. Busch
    G. Dzewas
    M. Schwaiger
    F. Bengel
    European Journal of Nuclear Medicine and Molecular Imaging, 2007, 34 : 405 - 412
  • [10] Is non-attenuation-corrected PET inferior to body attenuation-corrected PET or PET/CT in lung cancer?
    Maintas, Dimitris
    Houzard, Claire
    Ksyar, Rachid
    Mognetti, Thomas
    Maintas, Catherine
    Scheiber, Christian
    Itti, Roland
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (02): : 167 - 170