Calibration of laminar flow meters for process gases

被引:11
|
作者
Wright, John D. [1 ]
Cobu, Thiago [2 ]
Berg, Robert F. [1 ]
Moldover, Michael R. [1 ]
机构
[1] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
[2] Inst Pesquisas Tecnol, Sao Paulo, Brazil
关键词
Laminar flow meter; Calibration; Gas species; Gas properties; Rate of rise; Dynamic flow standard; Gravimetric flow standard; Process gases; CAPILLARY;
D O I
10.1016/j.flowmeasinst.2011.08.007
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
We calibrated three models of commercially-manufactured, laminar flow meters (LFMs) at four pressures (100, 200, 300, and 400 kPa) with five gases (N-2, Ar, He, CO2, and SF6) over a 10:1 flow range using NIST's primary flow standards as references. We combined three items: (1) the calibration data acquired with N-2, (2) gas-property data from NIST's database REFPROP 9.0, and (3) a physical model for each LFM that accounts for the effects of viscosity, entrance and exit effects, gas expansion, gas non-ideality, and slip. This combination predicted the calibrations for the flow of Ar, He, CO2, and SF6 with a maximum error of 0.8% for Reynolds numbers Re < 500. Under these conditions, the present LFM model allows prediction of calibration results for other gases with approximately three times more accuracy than conventional approaches that plot the flow coefficient as a function of the viscosity coefficient or Re. We represented the calibration data for SF6 in the range 500 < Re < 2000 by adding an empirical quadratic function to the model for one of the LFMs. Published by Elsevier Ltd
引用
收藏
页码:8 / 14
页数:7
相关论文
共 50 条
  • [1] The calibration of flow meters for the measurement of insecticide gases
    Craig, LC
    Richardson, CH
    [J]. JOURNAL OF ECONOMIC ENTOMOLOGY, 1930, 23 : 988 - 991
  • [2] Dynamic calibration of laminar flow sensor for gases
    Funaki, T
    Sengoku, K
    Kawashima, K
    Kagawa, T
    [J]. SICE 2004 ANNUAL CONFERENCE, VOLS 1-3, 2004, : 2437 - 2442
  • [3] CALIBRATION OF PEAK FLOW METERS
    FISHER, J
    SHAW, A
    [J]. THORAX, 1980, 35 (11) : 880 - 880
  • [4] CALIBRATION OF PEAK FLOW METERS
    FISHER, J
    SHAW, A
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1981, 53 (07) : 777 - 778
  • [5] RELATIVE CALIBRATION OF FLOW METERS WITH AN INTERFEROMETER
    KURLAPOV, LI
    [J]. INDUSTRIAL LABORATORY, 1967, 33 (01): : 149 - &
  • [6] Magnetic fluids in flow meters network for gases
    Popa, NC
    Siblini, A
    Jorat, L
    [J]. INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2004, 19 (1-4) : 509 - 514
  • [7] A new method for the calibration of peak flow meters
    Bushman, JA
    Calvert, JR
    Stockwell, J
    [J]. MEDICAL ENGINEERING & PHYSICS, 1997, 19 (04) : 359 - 365
  • [8] MEASUREMENT AND CONTROL TECHNOLOGY Calibration of flow meters
    Bolf, N.
    [J]. KEMIJA U INDUSTRIJI-JOURNAL OF CHEMISTS AND CHEMICAL ENGINEERS, 2016, 65 (7-8): : 449 - 452
  • [9] A NEW SYSTEM FOR THE CALIBRATION OF GAS FLOW METERS
    Pavlovic, Berislav
    Kozmar, Hrvoje
    Sunic, Miljenko
    [J]. TRANSACTIONS OF FAMENA, 2009, 33 (01) : 37 - 46
  • [10] CALIBRATION OF SOME WRIGHT PEAK FLOW METERS
    FISHER, J
    SHAW, A
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1980, 52 (04) : 461 - 464