Accuracy and efficiency of an infrared based positioning and tracking system for patient set-up and monitoring in image guided radiotherapy

被引:6
|
作者
Jia, Jing [1 ,2 ,3 ]
Xu, Gongming [2 ,3 ]
Pei, Xi [2 ,3 ]
Cao, Ruifen [2 ,3 ]
Hu, Liqin [1 ,2 ,3 ]
Wu, Yican [1 ,2 ,3 ]
机构
[1] Univ Sci & Technol China, Hefei 230027, Anhui, Peoples R China
[2] Chinese Acad Sci, Inst Nucl Energy Safety Technol, Key Lab Neutron & Radiat Safety, Hefei 230031, Anhui, Peoples R China
[3] Engn Technol Res Ctr Accurate Radiotherapy Anhui, Hefei 230031, Anhui, Peoples R China
基金
中国国家自然科学基金;
关键词
Infrared positioning; Respiratory control; Patient setup; QUALITY-ASSURANCE;
D O I
10.1016/j.infrared.2015.01.001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An infrared based positioning and tracking (IPT) system was introduced and its accuracy and efficiency for patient setup and monitoring were tested for daily radiotherapy treatment. The IPT system consists of a pair of floor mounted infrared stereoscopic cameras, passive infrared markers and tools used for acquiring localization information as well as a custom controlled software which can perform the positioning and tracking functions. The evaluation of IPT system characteristics was conducted based on the AAPM 147 task report. Experiments on spatial drift and reproducibility as well as static and dynamic localization accuracy were carried out to test the efficiency of the IPT system. Measurements of known translational (up to 55.0 mm) set-up errors in three dimensions have been performed on a calibration phantom. The accuracy of positioning was evaluated on an anthropomorphic phantom with five markers attached to the surface; the precision of the tracking ability was investigated through a sinusoidal motion platform. For the monitoring of the respiration, three volunteers contributed to the breathing testing in real time. The spatial drift of the IPT system was 0.65 mm within 60 min to be stable. The reproducibility of position variations were between 0.01 and 0.04 mm. The standard deviation of static marker localization was 0.26 mm. The repositioning accuracy was 0.19 mm, 0.29 mm, and 0.53 mm in the left/right (L/R), superior/inferior (S/I) and anterior/posterior (A/P) directions, respectively. The measured dynamic accuracy was 0.57 mm and discrepancies measured for the respiratory motion tracking was better than 1 mm. The overall positioning accuracy of the IPT system was within 2 mm. In conclusion, the IPT system is an accurate and effective tool for assisting patient positioning in the treatment room. The characteristics of the IPT system can successfully meet the needs for real time external marker tracking and patient positioning as well as respiration monitoring during image guided radiotherapy treatments. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:26 / 31
页数:6
相关论文
共 50 条
  • [1] Accuracy and efficiency of an infrared based positioning and tracking system for image-guided intervention
    Chen, Yunxia
    Zhang, Jie
    Ge, Yun
    Zhang, Bingqing
    Chai, Lei
    Cai, Jing
    Yan, Jing
    Liu, Baorui
    2016 IEEE 10TH INTERNATIONAL CONFERENCE ON NANO/MOLECULAR MEDICINE AND ENGINEERING (NANOMED), 2016, : 39 - 43
  • [2] Patient set-up in radiotherapy with video-based positioning system
    Tao, S
    Wu, A
    Wu, Y
    Chen, Y
    Zhang, J
    CLINICAL ONCOLOGY, 2006, 18 (04) : 363 - 366
  • [3] Dosimetric Validation of Surface Guided Patient Set-Up for Breast Radiotherapy
    Gopal, A.
    Lasio, G.
    Zhang, B.
    Xu, H.
    Chen, S.
    Mcavoy, S.
    Decesaris, C.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [4] Positioning accuracy of a new image-guided radiotherapy system
    Miyabe, Yuki
    Sawada, Akira
    Takayama, Kenji
    Kaneko, Shuji
    Mizowaki, Takashi
    Kokubo, Masaki
    Hiraoka, Masahiro
    MEDICAL PHYSICS, 2011, 38 (05) : 2535 - 2541
  • [5] IMAGE-GUIDED STEREOTACTIC BODY RADIOTHERAPY FOR LUNG TUMORS USING BODYLOC WITH TOMOTHERAPY: CLINICAL IMPLEMENTATION AND SET-UP ACCURACY
    Zhou, Jining
    Uhl, Barry
    Dewitt, Kelly
    Young, Mark
    Taylor, Brian
    Fei, Ding-Yu
    Lo, Yeh-Chi
    MEDICAL DOSIMETRY, 2010, 35 (01) : 12 - 18
  • [6] Positioning accuracy using an image-guided patient localization system
    Yan, H
    Yin, F
    Kim, J
    MEDICAL PHYSICS, 2002, 29 (06) : 1241 - 1241
  • [7] Set-up uncertainty during breast radiotherapy Image-guided radiotherapy for patients with initial extensive variation
    Yang, D. S.
    Yoon, W. S.
    Chung, S. Y.
    Lee, J. A.
    Lee, S.
    Park, Y. J.
    Kim, C. Y.
    Son, G. S.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2013, 189 (04) : 315 - 320
  • [8] Image-guided set-up verification for thoracic radiotherapy: A comparison of two immobilization methods
    Dahele, M.
    Tirona, R.
    Skinner, M.
    Lightstone, A.
    Ung, Y.
    Hunter, M.
    Cheung, P.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S59 - S59
  • [9] An Evaluation of Translational Set-Up Errors to Define An Institutional Protocol for Set-Up Margins in the Treatment of Cancer Cervix Using Image Guided Radiotherapy (IGRT)
    Rani, G.
    Kaur, H.
    Ninan, J.
    Chandrasenan, P.
    Abdussameeh, M. P.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [10] Study of total positioning accuracy of a newly developed image-guided radiotherapy system
    Kaneko, S.
    Takayama, K.
    Nagano, K.
    Narita, Y.
    Kokubo, M.
    Takahashi, K.
    Mizowaki, T.
    Hiraoka, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2533 - 2533