Detecting intake of radionuclides: In vivo screening measurements with conventional radiation protection instruments

被引:4
|
作者
Medici, S. [1 ,2 ,3 ]
Carbonez, P. [1 ,4 ]
Damet, J. [2 ,3 ,4 ]
Bochud, F. [2 ,3 ]
Bailat, C. [2 ,3 ]
Pitzschke, A. [2 ,3 ]
机构
[1] CERN, European Org Nucl Res, Geneva, Switzerland
[2] Lausanne Univ Hosp, Inst Radiat Phys, Lausanne, Switzerland
[3] Lausanne Univ, Lausanne, Switzerland
[4] Univ Otago, Dept Radiol, Christchurch, New Zealand
关键词
Intake of radionuclides; Occupational exposure; In vivo screening measurement; Instrument characterisation; Monitoring programme; ASSESS INTERNAL CONTAMINATION; CALIBRATION; DOSIMETRY; PHANTOM; I-131;
D O I
10.1016/j.radmeas.2019.01.022
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Internal contamination monitoring of a large number of workers is challenging, especially when dealing with various radionuclides. This work presents an approach for a pragmatic internal monitoring programme at CERN that is compatible with the Swiss regulation and features quick in vivo screening measurements. These are intended to be performed by the workers themselves using conventional radiation protection instruments. The programme ensures that the yearly committed effective dose does not exceed 1 mSv. This is done by defining screening intervals and coordinating the workers activity according to the minimum detectable committed effective dose associated to each screening. We elaborated a procedure to determine whether a radiation protection instrument is suitable to perform screening measurements and ensure the compliance with the annual E-50 <= 1 mSv limit. Seven conventional radiation protection instruments were characterised with reference sources of Co-57, Ba-133, Cs-137 and Co-60. For these instruments, values ranging between tens and hundreds of mu Sv were found for the minimum detectable committed effective dose when performing the screening measurement the day after the intake. The approach provides local radiation protection officers with basic tools to efficiently set up an internal monitoring programme.
引用
收藏
页码:126 / 132
页数:7
相关论文
共 7 条
  • [1] In vivo screening measurements with common radiation protection instruments
    Medici, Siria
    Desorgher, Laurent
    Carbonez, Pierre
    Damet, Jerome
    Bochud, Francois
    Pitzschke, Andreas
    12TH INTERNATIONAL CONFERENCE ON THE HEALTH EFFECTS OF INCORPORATED RADIONUCLIDES (HEIR 2018), 2019, 14
  • [2] Ultraviolet radiation protection by clothing:: Comparison of in vivo and in vitro measurements
    Gies, HP
    Roy, CR
    Holmes, G
    RADIATION PROTECTION DOSIMETRY, 2000, 91 (1-3) : 247 - 250
  • [3] Basic Characteristics of Radiation Protection Instruments on the Market: Assurance of the Quality of Measurements in Correspondence with International Standards
    Magyarov, A.
    Kratko, A.
    Blinou, U.
    Bury, D.
    2013 3RD INTERNATIONAL CONFERENCE ON ADVANCEMENTS IN NUCLEAR INSTRUMENTATION, MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA), 2013,
  • [4] Combined beta/gamma-radiation and bremsstrahlung monitor for measurements of incorporated radionuclides: The Sr-90 in-vivo counter
    Wahl, W
    Maushart, R
    Konig, K
    HornungLauxmann, L
    Burkart, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 369 (2-3): : 693 - 697
  • [5] Combined beta/gamma-radiation and bremsstrahlung monitor for measurements of incorporated radionuclides: The 90Sr in-vivo counter
    Wahl, W.
    Maushart, R.
    Konig, K.
    Hornung-Lauxmann, L.
    Burkart, W.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1996, 369 (2-3): : 693 - 697
  • [6] Combined beta/gamma-radiation and bremsstrahlung monitor for measurements of incorporated radionuclides: The 90Sr in-vivo counter
    Wahl, W.
    Maushart, R.
    Konig, K.
    Hornung-Lauxmann, L.
    Burkart, W.
    Nuclear Instruments & Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1996, 369 (2-3):
  • [7] VITAMINS AS RADIOPROTECTORS IN-VIVO .2. PROTECTION BY VITAMIN-A AND SOYBEAN OIL AGAINST RADIATION-DAMAGE CAUSED BY INTERNAL RADIONUCLIDES
    HARAPANHALLI, RS
    NARRA, VR
    YAGHMAI, V
    AZURE, MT
    GODDU, SM
    HOWELL, RW
    RAO, DV
    RADIATION RESEARCH, 1994, 139 (01) : 115 - 122