Analysis of a Unique Postmarket Surveillance Dataset That a Glucose Test-Strip Demonstrates no Evidence of Interference and Robust Clinical Accuracy Irrespective of the Prescription Medication Status of a Large Cohort of Patients With Diabetes

被引:5
|
作者
Grady, Mike [1 ]
Cameron, Hilary [1 ]
Phillips, Stuart [1 ]
Smith, Gillian [1 ]
Setford, Steven [1 ]
机构
[1] LifeScan Scotland Ltd, Beechwood Pk North, Inverness IV2 3ED, Scotland
来源
关键词
medications; accuracy; blood glucose monitor; self-monitoring of blood glucose;
D O I
10.1177/1932296819873053
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Despite a marked increase in polypharmacy in patients with diabetes there have been no thorough evaluations of the impact of polypharmacy on the accuracy of any current blood glucose monitoring (BGM) system. This study evaluated the accuracy of a BGM test-strip with respect to polypharmacy using a large clinical registry dataset. Methods: Medication profiles were analyzed for 830 subjects (334 with type 1 [T1D] and 496 with type 2 diabetes [T2D]) attending three hospitals. Blood samples were analyzed to determine clinical accuracy of the BGM test-strip compared to a laboratory comparator. Results: Across the 830 subjects, 473 different medications (41 diabetes and 432 nondiabetes) were recorded. Patients took on average 6.5 (n = 1-23) individual medications and 4 (n = 1-11) unique classes of medication. Clinical accuracy to EN ISO 15197:2015 criteria was met irrespective of increasing average number of individual medication, categorized from 1 to 4, 5 to 8, 9 to 12, and >12 taken per subject (97.7%, 98.4%, 98.1%, and 98.5%, respectively). Clinical accuracy to EN ISO 15197:2015 criteria was also met across 15 classes of medication using the combined dataset (98.1%; 13 003/13 253). Surveillance error grid analysis showed 98.8% (13 079/13 232) of readings presented no clinical risk. No individual class or combination of medication classes impacted clinical accuracy of the BGM test-strip. Conclusions: This comprehensive analysis for this specific test-strip platform demonstrated no evidence of interference and robust clinical accuracy of this test strip, irrespective of the prescription medication status of patients with diabetes.
引用
收藏
页码:82 / 90
页数:9
相关论文
共 1 条
  • [1] Post-Market Surveillance of a Blood Glucose Test Strip Demonstrates No Evidence of Interference on Clinical Accuracy in a Large Cohort of People with Type 1 or Type 2 Diabetes
    Phillips, Stuart
    Setford, Steven
    Grady, Mike
    Liu, Zuifang
    Cameron, Hilary
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2023, 17 (01): : 141 - 151