Quality of ECG monitoring with a miniature ECG recorder

被引:6
|
作者
Janata, Andreas [1 ]
Lemmert, Miguel E. [2 ]
Russell, James K. [3 ]
Gehman, Stacy [3 ]
Fleischhackl, Roman [1 ]
Robak, Oliver [1 ]
Pernicka, Elisabeth [4 ]
Sterz, Fritz [1 ]
Gorgels, Anton P. M. [2 ]
机构
[1] Med Univ Vienna, Dept Emergency Med, Vienna, Austria
[2] Univ Limburg, Acad Hosp Maastricht, Dept Cardiol, Maastricht, Netherlands
[3] Philips Med Syst, Seattle, WA USA
[4] Med Univ Vienna, Core Unit Med Stat & Informat, Vienna, Austria
来源
关键词
arrhythmia monitoring; electrocardiogram; new technology;
D O I
10.1111/j.1540-8159.2008.01070.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Long-term monitoring of patients at risk of arrhythmias would benefit from a miniaturized device. This study evaluated the quality of electrocardiogram (ECG) signals recorded by a miniaturized ECG recorder. Methods: ECG data were concurrently recorded with an investigational device (Philips Medical Systems, Seattle, WA, USA) and a Holter recorder from patients with cardiac-related symptoms at the emergency department of a tertiary care university hospital. The device was attached in one of four locations (L1: below left clavicle, L2: midsternal, L3: below left breast, L4: left anterior axillary line in 5th intercostal space). Selected ECG strips were analyzed for atrial activity, QRS width, and basic rhythm by two physicians. Patients were divided into groups based on their number of strips (A 2-5, B 6-9, C 10-21) for descriptive presentation of the data. For statistical analyses, nonlinear mixed-effects models were used. Results: In 90 patients, 574 pairs of strips were analyzed. In L1 (n = 37), agreement between the device and the Holter recorder on presence of P wave was 58-80% in groups A-C and in L2 (n = 28):94-98%, L3 (n = 12):30-72%, and L4 (n = 13):0-70%; on PR interval in L1:58-80%, L2:94-98%, L3:30-62%, and L4:0-70%; on rhythm in L1:56-79%, L2:94-98%, L3:30-62%, and L4:0-70%; on QRS width in L1:86-100%, L2:88-98%, L3:100%, and L4:96-100%. The agreement of L2 in the parameters P wave, PR interval, and rhythm diagnosis was better than in the other locations (P < 0.01). Conclusions: A miniature ECG monitor provided accurate assessment of atrial beats and rhythm diagnosis at a midsternal location and of QRS width at all locations.
引用
收藏
页码:676 / 684
页数:9
相关论文
共 50 条
  • [1] Wearable ECG Recorder with Acceleration Sensors for Monitoring Daily Stress
    Okada, Yoshio
    Yoto, Tsuyoshi Yi
    Suzuki, Taka-aki
    Sakuragawa, Satoshi
    Sakakibara, Hiroyuki
    Shimoi, Kayoko
    Sugiura, Toshifumi
    [J]. JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2013, 33 (04) : 420 - 426
  • [2] ECG Quality Measures in Telecare Monitoring
    Redmond, Stephen J.
    Lovell, Nigel H.
    Basilakis, Jim
    Celler, Branko G.
    [J]. 2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 2869 - +
  • [3] An Algorithm for Evaluating the ECG Signal Quality in 12 Lead ECG Monitoring System
    Zhang, Yue
    Hou, Zhongjie
    [J]. PROCEEDINGS OF 2015 6TH IEEE INTERNATIONAL CONFERENCE ON SOFTWARE ENGINEERING AND SERVICE SCIENCE, 2015, : 453 - 456
  • [4] LONG TIME ECG RECORDER WITH ANALYSING
    YangZhenye
    ChenHonglin
    PengLiangrui
    ChenEnke
    LiLinghua
    LinJiarui
    [J]. Chinese Journal of Biomedical Engineering, 1995, (02) : 91 - 92
  • [5] ECG RECORDER DISTRIBUTION AND HOSPITAL UTILIZATION
    HASLAM, KR
    [J]. MEDICAL INSTRUMENTATION, 1982, 16 (02): : 111 - 113
  • [6] A PORTABLE ECG RECORDER FOR SHIPBOARD USE
    RYACK, BL
    [J]. MILITARY MEDICINE, 1989, 154 (04) : 184 - 185
  • [7] Design of portable ECG recorder with USB storage
    Ho, C. S.
    Chiang, T. K.
    Lin, C. H.
    Lin, P. Y.
    Cheng, J. L.
    Ho, S. H.
    [J]. EDSSC: 2007 IEEE INTERNATIONAL CONFERENCE ON ELECTRON DEVICES AND SOLID-STATE CIRCUITS, VOLS 1 AND 2, PROCEEDINGS, 2007, : 1095 - +
  • [8] UNUSUAL ECG ARTIFACT - RESULTS OF A FAULTY RECORDER
    AGARWAL, SK
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1979, 242 (07): : 617 - 618
  • [9] A Palm Pilot based pocket ECG recorder
    Zou, YN
    Guo, ZY
    [J]. 2000 IEEE EMBS INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY APPLICATIONS IN BIOMEDICINE, PROCEEDINGS, 2000, : 110 - 112
  • [10] ECG Monitoring in an Airplane Seat: Appraising the Signal Quality
    Schumm, Johannes
    Arnrich, Bert
    Troester, Gerhard
    [J]. IEEE PERVASIVE COMPUTING, 2012, 11 (04) : 28 - 34