Application of Dual-Energy Spectral Computed Tomography in Bone Mineral Density Measurement: Phantom and Clinical Research

被引:6
|
作者
Wang, Mingyue [1 ]
Wu, Yan [1 ]
Zhou, Yue [1 ]
Dong, Junqiang [1 ]
Hu, Shenshen [1 ]
Hou, Ping [1 ]
Gao, Jianbo [1 ]
机构
[1] Zhengzhou Univ, Affiliated Hosp 1, Dept Radiol, Zhengzhou, Peoples R China
关键词
dual-energy spectral computed tomography; quantitative computed tomography; bone mineral density; European spine phantom; spine; MATERIAL DECOMPOSITION; TRABECULAR BONE; CT; OSTEOPOROSIS; RECONSTRUCTION; ACCURACY; SPINE; FAT;
D O I
10.2147/IJGM.S381857
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Early detection and treatment of osteoporosis through bone mineral density (BMD) measurement could aid in the prevention of osteoporosis-related fractures. We aimed to assess the parameter of dual-energy spectral CT (DesCT) consistency with BMD determination using quantitative computed tomography (QCT), thereby establishing a basis for further DesCT application for BMD determination. Methods: We subjected the European spine phantom, which contains three vertebral bodies (V1, V2, and V3), to DesCT with different radiation doses. The basis material pairs were hydroxyapatite (water), calcium (water), and hydroxyapatite (fat). Additionally, the medical records of 152 patients who underwent QCT and DesCT for chest scans in a two-month period were reviewed to measure BMD values. Results: No significant differences were found in the basis pair values of the V1, V2, or V3 vertebrae under different radiation doses in the phantom; in particular, the hydroxyapatite (water), hydroxyapatite (fat), relative error values of V1, V2, and V3 under different radiation doses were not significantly different (all p > 0.05). For patients, the hydroxyapatite (water), hydroxyapatite (fat), and hydroxyapatite (average) values measured by DesCT had a significant correlation with BMD measured by QCT Among 242 vertebrae (152 T12 and 90 L1 vertebrae), there was no significant difference between the BMD measured by QCT and the HAP (average) measured by DesCT (p = 0.071). The interclass correlation coefficient (ICC) value was 0.925 between the HAP (average) and HAP (average) with DesCT and BMD measured by QCT (p < 0.001). Bland-Altman diagram indicated that both measurements were in good agreement. Discussion: We showed that BMD values measured by DesCT were stable and repeatable under different radiation doses. DesCT and QCT measurements of human BMD were highly correlated. Thus, DesCT-based BMD assessment of the spine in a clinical setting could be considered feasible.
引用
收藏
页码:6887 / 6896
页数:10
相关论文
共 50 条
  • [1] The accuracy of bone mineral density measurement using dual-energy spectral CT and quantitative CT: a comparative phantom study
    Li, X.
    Li, J.
    Jiao, X.
    Jia, X.
    Zhang, X.
    Fan, G.
    Yang, J.
    Guo, J.
    CLINICAL RADIOLOGY, 2020, 75 (04) : 320.e9 - 320.e15
  • [3] The new fast kilovoltage-switching dual-energy computed tomography for measuring bone mineral density
    Wang, Mingyue
    Wu, Yan
    Zhou, Yue
    Dong, Junqiang
    Hou, Ping
    Gao, Jianbo
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (02) : 801 - 811
  • [4] Application of Dual-Energy Spectral Computed Tomography to Thoracic Oncology Imaging
    Kim, Cherry
    Kim, Wooil
    Park, Sung-Joon
    Lee, Young Hen
    Hwang, Sung Ho
    Yong, Hwan Seok
    Oh, Yu-Whan
    Kang, Eun-Young
    Lee, Ki Yeol
    KOREAN JOURNAL OF RADIOLOGY, 2020, 21 (07) : 838 - 850
  • [5] Spectral Signal Density of Carotid Plaque Using Dual-Energy Computed Tomography
    Reynoso, Exequiel
    Rodriguez-Granillo, Gaston A.
    Capunay, Carlos
    Deviggiano, Alejandro
    Meli, Francisco
    Carrascosa, Patricia
    JOURNAL OF NEUROIMAGING, 2017, 27 (05) : 511 - 516
  • [6] Clinical Application of Dual-Energy Spectral Computed Tomography in Detecting Cholesterol Gallstones From Surrounding Bile
    Yang, Chuang-bo
    Zhang, Shuang
    Jia, Yong-jun
    Duan, Hai-feng
    Ma, Guang-ming
    Zhang, Xi-rong
    Yu, Yong
    He, Tai-ping
    ACADEMIC RADIOLOGY, 2017, 24 (04) : 478 - 482
  • [7] Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment
    He, Tian
    Qian, Xiaojun
    Zhai, Renyou
    Yang, Zongtao
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (06) : 981 - 985
  • [8] Computation of Calcium Score With Dual-Energy Computed Tomography: A Phantom Study
    Kumar, Vidhya
    Min, James K.
    He, Xin
    Raman, Subha V.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2017, 41 (01) : 156 - 158
  • [9] MEASUREMENT OF TRABECULAR BONE-MINERAL BY DUAL ENERGY COMPUTED-TOMOGRAPHY
    ADAMS, JE
    CHEN, SZ
    ADAMS, PH
    ISHERWOOD, I
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1982, 6 (03) : 601 - 607
  • [10] A phantom for evaluating bone mineral density of the hand by dual-energy x-ray absorptiometry
    Steel, SA
    Swann, P
    Langley, G
    Langton, CM
    PHYSIOLOGICAL MEASUREMENT, 1997, 18 (03) : 233 - 240