Mechanical Properties of Open-Cell, Self-Expandable Shape Memory Alloy Carotid Stents

被引:28
|
作者
Carnelli, Davide [1 ]
Pennati, Giancarlo [1 ]
Villa, Tomaso [1 ]
Baglioni, Luigi [1 ]
Reimers, Bernhard [2 ]
Migliavacca, Francesco [1 ]
机构
[1] Politecn Milan, Dept Struct Engn, Lab Biol Struct Mech, I-20133 Milan, Italy
[2] Civ Hosp, Dept Cardiol, Mirano, Italy
关键词
Carotid stents; Bending stiffness; Radial stiffness; Shape memory alloy;
D O I
10.1111/j.1525-1594.2010.01018.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mechanical properties of carotid stents are important for the selection of the implantable device. The efficacy of cell design is usually investigated on the basis of clinical follow-ups; however, for the specific carotid application the relative importance of the cell design on the stenting outcome is debated. Ideally, a carotid stent should have at the same time low bending stiffness (i.e., high flexibility) to correctly conform to the vessel curvatures in the stenotic region and high stiffness to sustain the vessel wall in the radial direction. In this study, experimental tests on six carotid open-cell design self-expandable stents were carried out to assess bending and radial stiffness. A four-point bending test was adopted to measure the bending stiffness. Radial stiffness was obtained by measuring the stent resistance to local compression in a V-shaped support. Geometrical features, such as the link positioning and their length, were examined as a possible determinant of the mechanical properties of the tested stents. The tested stents showed very different geometrical features and mechanical properties despite similar material and a common open-cell structure consisting in rings connected by roughly straight links. In particular, different strategies are evident regarding the link position. Three distinct behaviors were observed: stents with both low radial and bending stiffness, moderate bending and high radial stiffness, and high bending and moderate radial stiffness. In general, a large link length allows the stent to be more flexible under bending condition, whereas the mutual links positioning plays a key role in the radial stiffness of these devices. Furthermore, for a fixed strut layout, a correlation was found between the strut cross area, as well as metal-to-artery ratio and the resultant radial and bending stiffness.
引用
收藏
页码:74 / U152
页数:7
相关论文
共 50 条
  • [1] Rapidly self-expandable polymeric Stents with a shape-memory property
    Chen, Mei-Chin
    Tsai, Hung-Wen
    Chang, Yen
    Lai, Wei-Yun
    Mi, Fwu-Long
    Liu, Chin-Tang
    Wong, Hen-Sheng
    Sung, Hsing-Wen
    BIOMACROMOLECULES, 2007, 8 (09) : 2774 - 2780
  • [2] A computational study on vascular damage caused by shape memory alloy self-expandable and balloon-expandable stents in a stenosed artery
    Rouhani, Fatemeh
    Fereidoonnezhad, Behrooz
    Zakerzadeh, Mohammad Reza
    Baghani, Mostafa
    JOURNAL OF INTELLIGENT MATERIAL SYSTEMS AND STRUCTURES, 2019, 30 (20) : 3113 - 3123
  • [3] Biocompatible Shape Memory Blend for Self-Expandable Stents with Potential Biomedical Applications
    Zheng, Yu
    Li, Ying
    Hu, Xue
    Shen, Jiabin
    Guo, Shaoyun
    ACS APPLIED MATERIALS & INTERFACES, 2017, 9 (16) : 13988 - 13998
  • [4] Mechanical properties and degradation process of biliary self-expandable biodegradable stents
    Kwon, Chang-Il
    Son, Jun Sik
    Kim, Kyu Seok
    Moon, Jong Pil
    Park, Sehwan
    Jeon, Jinkyung
    Kim, Gwangil
    Choi, Sung Hoon
    Ko, Kwang Hyun
    Jeong, Seok
    Lee, Don Haeng
    DIGESTIVE ENDOSCOPY, 2021, 33 (07) : 1158 - 1169
  • [5] Understanding the mechanical properties of self-expandable stents: A key to successful product development
    Yoshino, Daisuke
    Inoue, Katsumi
    Narita, Yukihito
    STROJNISKI VESTNIK-JOURNAL OF MECHANICAL ENGINEERING, 2008, 54 (06): : 471 - 485
  • [6] Biodegradable shape-memory block co-polymers for fast self-expandable stents
    Xue, Liang
    Dai, Shiyao
    Li, Zhi
    BIOMATERIALS, 2010, 31 (32) : 8132 - 8140
  • [7] Analytical models for predicting mechanical properties of self-expandable metal stents with cover membrane
    Moon, T
    Hong, D
    Chun, HJ
    Hyun, JH
    Lee, KB
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 163 - 166
  • [8] Can we develop self-expandable metallic stents without consideration of mechanical properties?
    Isayama, Hiroyuki
    Nakai, Yousuke
    Kogure, Hirofumi
    Hamada, Tsuyoshi
    Yamamoto, Natsuyo
    Koike, Kazuhiko
    ENDOSCOPY, 2014, 46 (08) : 715 - 715
  • [9] Functional Performances of CuZnAl Shape Memory Alloy Open-Cell Foams
    Biffi, C. A.
    Casati, R.
    Bassani, P.
    Tuissi, A.
    JOURNAL OF MATERIALS ENGINEERING AND PERFORMANCE, 2018, 27 (04) : 1706 - 1713
  • [10] Functional Performances of CuZnAl Shape Memory Alloy Open-Cell Foams
    C. A. Biffi
    R. Casati
    P. Bassani
    A. Tuissi
    Journal of Materials Engineering and Performance, 2018, 27 : 1706 - 1713