Characterization of one-dimensional capacitive micromachined ultrasonic immersion transducer arrays

被引:126
|
作者
Jin, XC [1 ]
Oralkan, Ö
Degertekin, FL
Khuri-Yakub, BT
机构
[1] Stanford Univ, Edward L Ginzton Lab, Stanford, CA 94305 USA
[2] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Atlanta, GA 30332 USA
基金
美国国家科学基金会;
关键词
D O I
10.1109/58.920706
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In this paper, we report on the characterization of 1-D arrays of capacitive micromachined ultrasonic transducers (cMUT). A 275- x 5600-mum 1-D CMUT array element is experimentally characterized, and the results are found to be in agreement with theoretical predictions. As a receiver, the transducer has a 0.28-fm/root Hz displacement sensitivity, and, as a transmitter, it produces 5 kPa/V of output pressure at the transducer surface at 3 MHz with a DC bias of 35 V, The transducer has more than 100% fractional bandwidth around 3 MHz, which makes it suitable for ultrasound imaging. The radiation pattern of isolated single elements, as well as those of array elements are measured, and two major sources of acoustical cross talk are identified. A weakly dispersive non-leaky interface wave (Stoneley wave) is observed to be propagating at the silicon substrate-fluid interface at a speed close to the speed of sound in the fluid. This wave causes internal reflections, spurious resonance, and radiation from the edges of the silicon substrate. The large lateral component of the particle velocity generated by the membranes at the edge of the cMUT array elements is found to be the source of this interface wave. Lowest order Lamb waves in the silicon substrate are also found to contribute to the cross talk between elements. These waves are excited at the edges of individual vibrating membranes, where they are anchored to the substrate. and result in a narrowing of the beam profile of the array elements. Several methods, such as trench isolation and wafer thinning, are proposed and implemented to modify the acoustical cross coupling between array elements.
引用
收藏
页码:750 / 760
页数:11
相关论文
共 50 条
  • [1] Initial pulse-echo imaging results with one-dimensional capacitive micromachined ultrasonic transducer arrays
    Oralkan, Ö
    Jin, XC
    Kaviani, K
    Ergun, AS
    Degertekin, FL
    Karaman, M
    Khuri-Yakub, BT
    [J]. 2000 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2000, : 959 - 962
  • [2] One-dimensional capacitative micromachined ultrasonic transducer arrays for echographic probes
    Cianci, E
    Minotti, A
    Foglietti, V
    Caliano, G
    Pappalardo, M
    [J]. MICROELECTRONIC ENGINEERING, 2004, 73-4 : 502 - 507
  • [3] Micromachined capacitive ultrasonic immersion transducer for medical imaging
    Jin, XC
    Khuri-Yakub, BT
    Degertekin, FL
    Ladabaum, I
    Calmes, S
    [J]. PROCEEDINGS OF THE 20TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 20, PTS 1-6: BIOMEDICAL ENGINEERING TOWARDS THE YEAR 2000 AND BEYOND, 1998, 20 : 779 - 782
  • [4] Recent progress in capacitive micromachined ultrasonic immersion transducer array
    Jin, XC
    Cheng, CH
    Oralkan, Ö
    Calmes, S
    Degertekin, FL
    Khuri-Yakub, BT
    [J]. ISIC-99: 8TH INTERNATIONAL SYMPOSIUM ON INTEGRATED CIRCUITS, DEVICES & SYSTEMS, PROCEEDINGS, 1999, : 159 - 162
  • [5] Modeling and investigation of immersion based capacitive micromachined ultrasonic transducer
    Lalnunfeli, H.
    Maity, Reshmi
    Tiwari, R. C.
    Dutta, Shankar
    Maity, Niladri Pratap
    [J]. MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2024, 30 (10): : 1315 - 1324
  • [6] Design and characterization of capacitive micromachined ultrasonic transducer
    Ahn, Bongyoung
    Kim, Ki bok
    Park, Hae won
    Kim, Young joo
    Kwak, Yong Seok
    [J]. ADVANCED NONDESTRUCTIVE EVALUATION I, PTS 1 AND 2, PROCEEDINGS, 2006, 321-323 : 132 - 135
  • [7] Fabrication and characterization of capacitive micromachined ultrasonic transducer
    Kim, KB
    Ahn, BY
    Park, HW
    Kim, YJ
    [J]. ADVANCES IN FRACTURE AND STRENGTH, PTS 1- 4, 2005, 297-300 : 2225 - 2230
  • [8] One-dimensional equivalent circuit for ultrasonic transducer arrays
    Bybi, Bdelmajid
    Mouhat, Ouadia
    Garoum, Mohammed
    Drissi, Hilal
    Grondel, Sebastien
    [J]. APPLIED ACOUSTICS, 2019, 156 : 246 - 257
  • [9] Capacitive micromachined ultrasonic transducer arrays as tunable acoustic metamaterials
    Lani, Shane W.
    Rashid, M. Wasequr
    Hasler, Jennifer
    Sabra, Karim G.
    Degertekin, F. Levent
    [J]. APPLIED PHYSICS LETTERS, 2014, 104 (05)
  • [10] Medical imaging using capacitive micromachined ultrasonic transducer arrays
    Johnson, J
    Oralkan, Ö
    Demirci, U
    Ergun, S
    Karaman, M
    Khuri-Yakub, P
    [J]. ULTRASONICS, 2002, 40 (1-8) : 471 - 476