Progress in the development of large-area modular 64 x 64 CdZnTe imaging arrays for nuclear medicine

被引:34
|
作者
Matherson, KJ [1 ]
Barber, HB
Barrett, HH
Eskin, JD
Dereniak, EL
Marks, DG
Woolfenden, JM
Young, ET
Augustine, FL
机构
[1] Univ Arizona, Ctr Opt Sci, Tucson, AZ 85721 USA
[2] Univ Arizona, Div Nucl Med, Tucson, AZ 85724 USA
[3] Univ Arizona, Steward Observ, Tucson, AZ 85721 USA
[4] Augustine Engn, Encinitas, CA 92024 USA
关键词
D O I
10.1109/23.682407
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Previous efforts by our group have demonstrated the potential of hybrid semiconductor detector arrays for use in gamma-ray imaging applications. In this paper, we describe progress in the development of a prototype imaging system consisting of a 64 x 64-pixel CdZnTe detector array mated to a multiplexer readout circuit that was custom designed for our nuclear medicine application. The detector array consists of a 0.15 cm thick slab of CdZnTe which has a 64x64 array of 380 mu m square pixel electrodes on one side produced by photolithography; the other side has a continuous electrode biased at -150 V. Electrical connections between the detector electrodes and corresponding multiplexer bump pads are made with indium bump bonds. Although the CdZnTe detector arrays characterized in this paper are room-temperature devices, a slight amount of cooling is necessary to reduce thermally generated dark current in the detectors. Initial tests show that this prototype imager functions well with more than 90% of its pixels operating. The device is an excellent imager; phantom images have a spatial resolution of 1.5 mm, limited by the collimator bore.
引用
下载
收藏
页码:354 / 358
页数:5
相关论文
共 50 条
  • [1] Progress in the development of large-area modular 64 x 64 CdZnTe imaging arrays for nuclear medicine
    Matherson, KJ
    Barber, HB
    Barrett, HH
    Eskin, JD
    Dereniak, EL
    Marks, DG
    Woolfenden, JM
    Young, ET
    Augustine, FL
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 276 - 280
  • [2] Progress in the development of large-area modular 64 × 64 CdZnTe imaging arrays for nuclear medicine
    Matherson, K.J.
    Barber, H.B.
    Barrett, H.H.
    Eskin, J.D.
    Dereniak, E.L.
    Marks, D.G.
    Woolfenden, J.M.
    Young, E.T.
    Augustine, F.L.
    IEEE Transactions on Nuclear Science, 1998, 45 (3 pt 1): : 354 - 358
  • [3] Modular 64x64 CdZnTe arrays with multiplexer readout for high-resolution nuclear medicine imaging
    Woolfenden, JM
    Barber, HB
    Barrett, HH
    Dereniak, EL
    Eskin, JD
    Marks, DG
    Matherson, KJ
    Young, ET
    Augustine, FL
    INFRARED APPLICATIONS OF SEMICONDUCTORS II, 1998, 484 : 267 - 271
  • [4] Modular 64x64 CdZnTe arrays with multiplexer readout for high-resolution nuclear medicine imaging
    Woolfenden, JM
    Barber, HB
    Barrett, HH
    Dereniak, EL
    Eskin, JD
    Marks, DG
    Matherson, KJ
    Young, ET
    Augustine, FL
    SEMICONDUCTORS FOR ROOM-TEMPERATURE RADIATION DETECTOR APPLICATIONS II, 1997, 487 : 239 - 243
  • [5] Development of modular 64x64 CdZnTe imaging arrays with multiplexer readout.
    Woolfenden, JM
    Barber, HB
    Barrett, HH
    Dereniak, ES
    Eskin, JD
    Marks, DG
    Matherson, KJ
    Young, ET
    Augustine, FL
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05) : 536 - 536
  • [6] Development of a 64x64 CdZnTe array and associated readout integrated circuit for use in nuclear medicine
    Barber, HB
    Barrett, HH
    Augustine, FL
    Hamilton, WJ
    Apotovsky, BA
    Dereniak, EL
    Doty, FP
    Eskin, JD
    Garcia, JP
    Marks, DG
    Matherson, KJ
    Woolfenden, JM
    Young, ET
    JOURNAL OF ELECTRONIC MATERIALS, 1997, 26 (06) : 765 - 772
  • [7] Development of a 64 × 64 CdZnTe array and associated readout integrated circuit for use in nuclear medicine
    H. B. Barber
    H. H. Barrett
    F. L. Augustine
    W. J. Hamilton
    B. A. Apotovsky
    E. L. Dereniak
    F. P. Doty
    J. D. Eskin
    J. P. Garcia
    D. G. Marks
    K. J. Matherson
    J. M. Woolfenden
    E. T. Young
    Journal of Electronic Materials, 1997, 26 : 765 - 772
  • [8] CdZnTe arrays for nuclear medicine imaging
    Barber, HB
    HARD X-RAY/GAMMA-RAY AND NEUTRON OPTICS, SENSORS, AND APPLICATIONS, 1996, 2859 : 26 - 28
  • [9] Fabrication and performance of large (64x64) arrays of integrated VCSELs and detectors
    Geib, KM
    Choquette, KD
    Serkland, DK
    Allerman, AA
    Hargett, TW
    VERTICAL-CAVITY SURFACE-EMITTING LASERS VI, 2002, 4649 : 152 - 157
  • [10] Progress in developing focal-plane-multiplexer readout for large CdZnTe arrays for nuclear medicine applications
    Barber, HB
    Marks, DG
    Apotovsky, BA
    Augustine, FL
    Barrett, HH
    Butler, JF
    Dereniak, EL
    Doty, FP
    Eskin, JD
    Hamilton, WJ
    Matherson, KJ
    Venzon, JE
    Woolfenden, JM
    Young, ET
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2): : 262 - 265