Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers

被引:20
|
作者
Bechstedt, Maren [1 ]
Gustafson, Jonathan A. [2 ]
Mell, Steven P. [2 ]
Guehrs, Julian [1 ]
Morlock, Michael M. [1 ]
Levine, Brett R. [2 ]
Lundberg, Hannah J. [2 ]
机构
[1] TUHH Hamburg Univ Technol, Inst Biomech, D-21073 Hamburg, Germany
[2] Rush Univ, Dept Orthoped Surg, Med Ctr, 1611 W Harrison St Suite 201, Chicago, IL 60612 USA
关键词
Total hip arthroplasty; Modularity; Surface topography; Finite element analysis; Contact mechanics; FINITE-ELEMENT; FRETTING WEAR; MORSE TAPER; MECHANICAL RESPONSE; REALISTIC FORCES; INTERFACE; STRENGTH; REPLACEMENTS; PREDICTION; IMPLANTS;
D O I
10.1016/j.jbiomech.2020.109689
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Implant failure due to fretting-corrosion of head-neck modular junctions is a rising problem in total hip arthroplasty. Fretting-corrosion initiates when micromotion leads to metal release; however, factors leading to micromotion, such as microgrooves on the stem taper, are not fully understood. The purpose of this study is to describe a finite element analysis technique to determine head-neck contact mechanics and investigate the effect of stem taper microgroove height during head-neck assembly. Two-dimensional axisymmetric finite element models were created. Models were created for a ceramic femoral head and a CoCrMo femoral head against Ti6AI4V stem tapers and compared to available data from prior experiments. Stem taper microgroove height was investigated with a generic 12/14 model. Head-neck assembly was performed to four maximum loads (500 N, 2000 N, 4000 N, 8000 N). For the stem taper coupled with the ceramic head, the number of microgrooves in contact and plastically deformed differed by 2.5 microgrooves (4%) and 6.5 microgrooves (11%), respectively, between the finite element models and experiment. For the stem taper coupled with the CoCrMo head, all microgrooves were in contact after all assembly loads in the finite element model due to an almost identical conical angle between the taper surfaces. In the experiments, all grooves were only in contact for the 8000 N assembly load. Contact area, plastic (permanent) deformation, and contact pressure increased with increasing assembly loads and deeper microgrooves. The described modeling technique can be used to investigate the relationship between implant design factors, allowing for optimal microgroove design within material couples. (C) 2020 Elsevier Ltd. All rights reserved.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Head-neck taper corrosion following total hip arthroplasty with Stryker Meridian stem
    Patel, Sonika
    Talmo, Carl T.
    Nandi, Sumon
    HIP INTERNATIONAL, 2016, 26 (06) : E49 - E51
  • [2] Geometric Variations of Modular Head-Stem Taper Junctions of Total Hip Replacements
    Wade, A.
    Beadling, A. R.
    Neville, A.
    De Villiers, D.
    Cullum, C. J.
    Collins, S.
    Bryant, M. G.
    MEDICAL ENGINEERING & PHYSICS, 2020, 83 : 34 - 47
  • [3] Finite element analysis of the head-neck taper interface of modular hip prostheses
    Dyrkacz, R. M. R.
    Brandt, J. M.
    Morrison, J. B.
    O' Brien, S. T.
    Ojo, O. A.
    Turgeon, T. R.
    Wyss, U. P.
    TRIBOLOGY INTERNATIONAL, 2015, 91 : 206 - 213
  • [4] Head-Neck Taper Corrosion in Hip Arthroplasty
    Hussenbocus, S.
    Kosuge, D.
    Solomon, L. B.
    Howie, D. W.
    Oskouei, R. H.
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [5] Does Surface Topography Play a Role in Taper Damage in Head-neck Modular Junctions?
    Pourzal, Robin
    Hall, Deborah J.
    Ha, Nguyen Q.
    Urban, Robert M.
    Levine, Brett R.
    Jacobs, Joshua J.
    Lundberg, Hannah J.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (10) : 2232 - 2242
  • [6] A comparison study of stem taper material loss at similar and mixed metal head-neck taper junctions
    Langton, D. J.
    Sidaginamale, R. P.
    Joyce, T. J.
    Meek, R. D.
    Bowsher, J. G.
    Deehan, D.
    Nargol, A. V. F.
    Holland, J. P.
    BONE & JOINT JOURNAL, 2017, 99B (10): : 1304 - 1312
  • [7] CORR InsightsA®: Does Surface Topography Play a Role in Taper Damage in Head-neck Modular Junctions?
    Fernandez-Fairen, Mariano
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (10) : 2243 - 2245
  • [8] AN ANALYSIS OF THE HEAD-NECK TAPER INTERFACE IN RETRIEVED HIP PROSTHESES
    LIEBERMAN, JR
    RIMNAC, CM
    GARVIN, KL
    KLEIN, RW
    SALVATI, EA
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1994, (300) : 162 - 167
  • [9] Spontaneous reduction in both dislocated head and dissociation of head-neck in a modular total hip arthroplasty
    Raviraj, A.
    Anand, Ashish
    Chakravarthy, Murali
    Pai, Sanjay
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2012, 22 : 127 - 130
  • [10] Contact Mechanics and Plastic Deformation at the Local Surface Topography Level After Assembly of Modular Head-Neck Junctions in Modern Total Hip Replacement Devices
    Lundberg, Hannah J.
    Ha, Nguyen Q.
    Hall, Deborah J.
    Urban, Robert M.
    Levine, Brett R.
    Pourzal, Robin
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 59 - 82