Traceable radiance source for spectroradiometer calibration

被引:0
|
作者
van den Berg, Steven [1 ]
Dekker, Paul [1 ]
van der Laan, Tessel [1 ]
Otter, Gerard [2 ]
Dijkhuizen, Niels [2 ]
Gur, Bilgehan [2 ]
Brinkers, Sanneke [2 ]
机构
[1] VSL, Thijsseweg 11, NL-2629 JA Delft, Netherlands
[2] TNO, Stieltjesweg 1, NL-2628 CK Delft, Netherlands
基金
欧盟地平线“2020”;
关键词
Radiance measurement; detector calibration; metrology; traceability; Earth observation;
D O I
10.1117/12.2536143
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
SI traceable calibration of satellite instruments like spectroradiometers get increasingly important in space-based observation for various reasons. First, requirements on measurements in terms of accuracy, dynamic range and wavelength range get increasingly demanding. Second, long term observations require SI traceability for comparability of data records that have been acquired over a long time interval (decades) and with different instruments. In this contribution we outline an approach for radiance calibration of spectroradiometers with a wavelength-tunable radiance source. The radiance source is based on an integrating sphere and a widely tunable laser (UVN-SWIR) and will be calibrated against a set of reference detector standards. By offering the radiance source as a stimulus to the spectroradiometer under test, the instrument can be calibrated for spectral radiance response. Potentially this traceability route could lead to a reduced measurement uncertainty compared to the traditional FEL-lamp based method. Furthermore, other instrument characteristics can also be derived from the same set of measurements, like stray-light effects and the instrument response function.
引用
下载
收藏
页数:9
相关论文
共 50 条
  • [1] Spectroradiometer Calibration for Radiance Transfer Measurements
    Rammeloo, Clemens
    Baumgartner, Andreas
    SENSORS, 2023, 23 (04)
  • [2] Calibration of a CubeSat spectroradiometer with a narrow-band widely tunable radiance source
    van den Berg, Steven
    Dekker, Paul
    Otter, Gerard
    Pascoa, Marcela Pelica
    Dijkhuizen, Niels
    APPLIED OPTICS, 2021, 60 (07) : 1995 - 2002
  • [3] Freeform reflector light source used for space traceable spectral radiance calibration on the solar reflected band
    Jiang, Ye
    Tian, Jiewen
    Fang, Wei
    Hu, Denghui
    Ye, Xin
    OPTICS EXPRESS, 2023, 31 (05) : 8049 - 8067
  • [4] Spectral radiance responsivity calibration of ultraviolet remote sensing spectroradiometer in space using integrating sphere
    Wang, Shu-Rong
    Xing, Jin
    Li, Fu-Tian
    Guangxue Jingmi Gongcheng/Optics and Precision Engineering, 2006, 14 (02): : 185 - 190
  • [5] A SPECTROSCOPIC BLACKBODY RADIANCE CALIBRATION PROCEDURE WITH INHERENT SOURCE TEMPERATURE CALIBRATION
    TANK, V
    INFRARED PHYSICS, 1989, 29 (2-4): : 209 - 210
  • [6] Comparison of spectral radiance calibrations of spectroradiometer for ultraviolet space remote sensing using three calibration techniques
    Xing, Jin
    Wang, Shu-Rong
    Li, Fu-Tian
    Zhongguo Jiguang/Chinese Journal of Lasers, 2006, 33 (04): : 509 - 515
  • [7] CALIBRATION OF A FIELD SPECTRORADIOMETER
    LEAMER, RW
    LAROCCA, AJ
    OPTICAL ENGINEERING, 1973, 12 (04) : 124 - 130
  • [8] MEASURING RADIANCE CHARACTERISTICS OF SOIL WITH A FIELD SPECTRORADIOMETER
    CIPRA, JE
    STONER, ER
    BAUMGARDNER, MF
    MACDONAL.RB
    SOIL SCIENCE SOCIETY OF AMERICA PROCEEDINGS, 1971, 35 (06): : 1014 - +
  • [9] Radiance uniformity measurement of a miniaturized integrating sphere light source for microscope radiance responsivity calibration
    Fu, Yangting
    Liu, Xiangliang
    He, Yingwei
    Liu, Wende
    Xu, Nan
    Lin, Yandong
    Gan, Haiyong
    AOPC 2020: OPTICAL SENSING AND IMAGING TECHNOLOGY, 2020, 11567
  • [10] Study on the Characteristics of Radiance Calibration Using Nonuniformity Extended Source
    Wang Jian-wei
    Huang Min
    Xiangli Bin
    Tu Xiao-long
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33 (07) : 2002 - 2005