Surface dosimetry for oblique tangential photon beams: A Monte Carlo simulation study

被引:22
|
作者
Chow, James C. L. [1 ,2 ,3 ,4 ]
机构
[1] Univ Toronto, Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
[3] Univ Waterloo, Dept Phys, Waterloo, ON N2L 3G1, Canada
[4] Ryerson Univ, Dept Phys, Toronto, ON M5B 2K3, Canada
关键词
surface dose; beam obliquity; Monte Carlo simulation; tangential beam;
D O I
10.1118/1.2818956
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effect of beam obliquity on the surface relative dose profiles for the tangential photon beams was studied. The 6 and 15 MV photon beams with 4 x 4 and 10 x 10 cm(2) field sizes produced by a Varian 21 EX linear accelerator were used. Phase-space models of the photon beams were created using Monte Carlo simulations based on the EGSnrc code, and were verified using film measurements. The relative dose profiles in the phantom skin, at 2 mm depth from the surface of the half-phantom geometry, or HPG, were calculated for increasing gantry angles from 270 to 280 deg clockwise. Relative dose profiles of a full phantom enclosing the whole tangential beam (full phantom geometry, or FPG) were also calculated using Monte Carlo simulation as a control for comparison. The results showed that, although the relative dose profiles in the phantom skin did not change significantly with an oblique beam using a FPG, the surface relative depth dose was increased for the HPG. In the HPG, with 6 MV photon beams and field size= 10 x 10 cm(2), when the beam angle, starting from 270 deg, was increased from 1 to 3 deg, the relative depth doses in the phantom skin were increased from 68% to 79% at 10 cm depth. This increase in dose was slightly larger than the dose from 15 MV photon beams with the same field size and beam angles, where the relative depth doses in phantom skin were increased from 81% to 87% at 10 cm depth. A parameter called the percent depth dose (PDD) ratio, defined as the relative depth dose from the HPG to the relative depth dose from the FPG at a given depth along the phantom skin, was used to evaluate the effect of the phantom-air interface. It is found that the PDD ratio increased significantly when the beam angle was changed from zero to 1-3 degrees. Moreover, the PDD ratio, for a given field size, experienced a greater increase for 6 MV than for 15 MV. For the same photon beam energy, the PDD ratio increased more with a 4 x 4 cm(2) field compared to 10 x 10 cm(2). The results in this study will be useful for physicists and dosimetrists to predict the surface relative dose variations when using clinical tangential-like photon beams in radiation therapy. (c) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:70 / 76
页数:7
相关论文
共 50 条
  • [1] Dosimetry of oblique tangential photon beams calculated by superposition/convolution algorithms: a Monte Carlo evaluation
    Chow, James C. L.
    Jiang, Runqing
    Leung, Michael K. K.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (01): : 108 - 121
  • [2] Monte Carlo Simulation On the Percentage Depth Dose Ratio for the Surface Dosimetry Using Tangential Photon Beams
    Chow, J.
    Grigorov, G.
    MEDICAL PHYSICS, 2008, 35 (06) : 2807 - +
  • [3] Mucosal dosimetry on unflattened photon beams: a Monte Carlo phantom study
    Chow, James C. L.
    Owrangi, Amir M.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (01):
  • [4] Monte Carlo physical dosimetry for small photon beams
    Perucha, M
    Sánchez-Doblado, F
    Rincón, M
    Leal, A
    Núnez, L
    Arráns, R
    Carrasco, E
    Sánchez-Calzado, JA
    Errazquin, L
    ADVANCED MONTE CARLO FOR RADIATION PHYSICS, PARTICLE TRANSPORT SIMULATION AND APPLICATIONS, 2001, : 555 - 559
  • [5] Surface Dosimetry in the Presence of a Bone in Skin Radiotherapy: A Monte Carlo Study Using KV Photon Beams
    Chow, J.
    Owrangi, A.
    Gorov, G. Gri
    MEDICAL PHYSICS, 2013, 40 (06)
  • [6] Monte Carlo Simulation On Surface Dose in Preclinical Irradiation Using Monoenergetic Photon Beams
    Vuong, A.
    Chow, J.
    MEDICAL PHYSICS, 2016, 43 (06) : 3620 - 3620
  • [7] Determining superficial dosimetry for the internal canthus from the Monte Carlo simulation of kV photon and MeV electron beams
    B. E. Currie
    Australasian Physics & Engineering Sciences in Medicine, 2009, 32 : 68 - 80
  • [8] Determining superficial dosimetry for the internal canthus from the Monte Carlo simulation of kV photon and MeV electron beams
    Currie, B. E.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (02) : 68 - 80
  • [9] Interface dosimetry: measurements and Monte Carlo simulations of low-energy photon beams
    Das, IJ
    Kassaee, A
    Verhaegen, F
    Moskvin, VP
    RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) : 593 - 595
  • [10] Monte Carlo study of conversion factors for ionization chamber dosimetry in solid slab phantoms for MV photon beams
    Park, Dong-wook
    Lee, Jai-ki
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 69 (04) : 647 - 652