Hand-held nerve conduction device in carpal tunnel syndrome: A prospective study

被引:2
|
作者
Tan, S. Veronica [1 ]
Sandford, Fiona [2 ]
Stevenson, Mark [1 ]
Probert, Sara [3 ]
Sanders, Sue [1 ]
Mills, Kerry R. [4 ]
Koutroumanidis, Michaelis [1 ]
机构
[1] Guys & St Thomas Fdn NHS Trust, St Thomas Hosp, Dept Neurol & Clin Neurophysiol, London, England
[2] Guys & St Thomas Fdn NHS Trust, St Thomas Hosp, Dept Physiotherapy, London, England
[3] Guys & St Thomas Fdn NHS Trust, St Thomas Hosp, Dept Occupat Therapy, London, England
[4] Kings Coll London, London WC2R 2LS, England
关键词
carpal tunnel syndrome; EMG; hand-held nerve conduction device; neurophysiologic testing; peripheral nerve lesions; SCREENING TOOL; NC-STAT;
D O I
10.1002/mus.23279
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Introduction: We assessed the clinical impact of replacing standard neurophysiologic testing with a hand-held device (Mediracer) for diagnosis of carpal tunnel syndrome (CTS). Methods: One hundred patients (200 hands) with suspected CTS were studied by blinded assessors [Hand-therapist (HT) 1 and Consultant Neurophysiologist] using the Mediracer, followed by standard neurophysiologic testing. To simulate testing by personnel without neurological training, Mediracer recordings were analyzed separately by an assessor who had not seen the patients (HT2). Results: Correlation of the CTS grades was 0.94 for the results obtained by HT1, and 0.87 for HT2. The sensitivity and specificity of the Mediracer was 0.85 and 0.9, respectively, by HT1, and 0.84 and 0.89 for HT2. Nine patients had conditions other than CTS, and 35 patients were judged to require further investigation. Conclusions: The Mediracer should only be used in patients with typical CTS symptoms and signs and no muscle wasting who have had careful neurological assessment. Muscle Nerve 45: 635-641, 2012
引用
收藏
页码:635 / 641
页数:7
相关论文
共 50 条
  • [1] Nerve conduction study in the asymptomatic hand of unilateral carpal tunnel syndrome
    Lee, SJ
    Hyun, JK
    Jeon, JY
    Park, WB
    Kim, BH
    MUSCLE & NERVE, 2005, 32 (03) : 387 - 388
  • [2] The optimum hand temperature to study nerve conduction in patients with carpal tunnel syndrome
    Gavanozi, Evangelia
    Veltsista, Dimitra
    Polychronopoulos, Panagiotis
    Chroni, Elisabeth
    JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2020, 51
  • [3] A handheld nerve conduction measuring device in carpal tunnel syndrome
    Tolonen, U.
    Kallio, M.
    Ryhanen, J.
    Raatikainen, T.
    Honkala, V.
    Lesonen, V.
    ACTA NEUROLOGICA SCANDINAVICA, 2007, 115 (06): : 390 - 397
  • [4] DIGITAL NERVE CONDUCTION IN CARPAL TUNNEL SYNDROME
    MCLEOD, JG
    AUSTRALASIAN ANNALS OF MEDICINE, 1967, 16 (02): : 186 - &
  • [5] MOTOR NERVE CONDUCTION IN THE CARPAL TUNNEL SYNDROME
    THOMAS, PK
    NEUROLOGY, 1960, 10 (12) : 1045 - 1050
  • [6] Reliability of Hand-Held Dynamometer Measurement for Thumb Palmar Abduction Strength in Carpal Tunnel Syndrome
    Doi, Kazuteru
    Yasui, Noritada
    Isomura, Yuya
    Tagawa, Masafumi
    Hattori, Yasunori
    Sakamoto, Sotetsu
    JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2025,
  • [7] The diagnostic utility of a new handheld nerve conduction device in carpal tunnel syndrome
    Ryhanen, J.
    Lesonen, V.
    Kallio, M.
    Kujala, S.
    Raatikainen, T.
    Hakalax, N.
    Tolonen, U.
    PROCEEDINGS OF THE 10TH CONGRESS OF THE INTERNATIONAL FEDERATION OF SOCIETIES FOR SURGERY OF THE HAND & 7TH CONGRESS OF THE INTERNATIONAL FEDERATION OF SOCIETIES FOR HAND THERAPY, 2007, : 355 - +
  • [8] Hand-Held Device
    Hadas, Rachel
    HUDSON REVIEW, 2019, 71 (04): : 791 - 794
  • [9] A new hand-held nerve conduction measuring device for the accurate diagnosis of diabetic peripheral neuropathy
    Greig, M.
    Shillo, P.
    Rao, G.
    Gandhi, R. A.
    Tesfaye, S.
    Selvarajah, D.
    DIABETIC MEDICINE, 2014, 31 : 144 - 144
  • [10] The true utility and indication for hand-held nerve conduction devices
    Ahmed, Aiesha
    Kothari, Milind J.
    MUSCLE & NERVE, 2012, 45 (05) : 621 - 622