Fiber-optic biochemical gas sensor (bio-sniffer) for sub-ppb monitoring of formaldehyde vapor

被引:22
|
作者
Kudo, Hiroyuki [1 ]
Wang, Xin [1 ,2 ]
Suzuki, Yuki [1 ]
Ye, Ming [1 ]
Yamashita, Toshifumi [1 ]
Gessei, Tomoko [1 ,2 ]
Miyajima, Kumiko [1 ]
Arakawa, Takahiro [1 ]
Mitsubayashi, Kohji [1 ]
机构
[1] Tokyo Med & Dent Univ, Inst Biomat & Bioengn, Dept Biomed Devices & Instrumentat, Chiyoda Ku, Tokyo 1010062, Japan
[2] Tokyo Metropolitan Ind Technol Res Inst, Kita Ku, Tokyo 1158586, Japan
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2012年 / 161卷 / 01期
基金
日本科学技术振兴机构; 日本学术振兴会;
关键词
UV LED; Gas sensor; Fiber-optic; NADH; Formaldehyde; Monitoring; Sub-ppb; NIO THIN-FILM; ETHANOL; DEHYDROGENASE; BIOSENSOR; CARBON; AIR; ACETALDEHYDE; MORTALITY; OXIDASE; ENZYMES;
D O I
10.1016/j.snb.2011.10.065
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A high sensitive fiber-optic biochemical gas sensor (bio-sniffer) for sub-parts-per-billion (sub-ppb) monitoring of formaldehyde (FA) vapor was constructed and tested. The bio-sniffer measures formaldehyde vapor as fluorescence of reduced nicotinamide adenine dinucleotide (NADH), which is the product of formaldehyde dehydrogenase (FALDH) reaction. An ultraviolet light emitting diode (UV-LED) with peak emission of 335 nm was employed as an excitation light source. Emission of the UV-LED was introduced to the optode through an optical fiber and fluorescence of NADH was picked up coaxially at the optode. In order to achieve sub-ppb sensitivity, two types of detectors (a spectrometer and a photomultiplier tube (PMT)) were tested. The reaction system was also improved. Aldehyde dehydrogenase (ALDH), which is more active with formaldehyde, was first tested instead of FALDH. The product of ALDH reaction was again oxidized by formate dehydrogenase (FDH) to enhance production of NADH. As a result of these improvements, real-time monitoring of FA was performed with a detection limit of 0.75 ppb, which is comparative to chromatographic methods. Such a high sensitive monitoring method is expected to improve the quality of indoor air management. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:486 / 492
页数:7
相关论文
共 20 条
  • [1] Direct Measurement of Gaseous Formaldehyde from Food with a Fiber-Optic Biochemical Gas Sensor (Bio-sniffer)
    Toma, Koji
    Miyajima, Kumiko
    Sawada, Shin-ichi
    Arakawa, Takahiro
    Kudo, Hiroyuki
    Akiyoshi, Kazunari
    Mitsubayashi, Kohji
    SENSORS AND MATERIALS, 2016, 28 (12) : 1265 - 1272
  • [2] Biochemical gas sensor (bio-sniffer) for ultrahigh-sensitive gaseous formaldehyde monitoring
    Kudo, Hiroyuki
    Suzuki, Yuki
    Gessei, Tomoko
    Takahashi, Daishi
    Arakawa, Takahiro
    Mitsubayashi, Kohji
    BIOSENSORS & BIOELECTRONICS, 2010, 26 (02): : 854 - 858
  • [3] Fiber-optic bio-sniffer (biochemical gas sensor) for high-selective monitoring of ethanol vapor using 335 nm UV-LED
    Kudo, Hiroyuki
    Sawai, Masayuki
    Suzuki, Yuki
    Wang, Xin
    Gessei, Tomoko
    Takahashi, Daishi
    Arakawa, Takahiro
    Mitsubayashi, Kohji
    SENSORS AND ACTUATORS B-CHEMICAL, 2010, 147 (02) : 676 - 680
  • [4] Fiber-Optic Bio-sniffer (Biochemical Gas Sensor) Using Reverse Reaction of Alcohol Dehydrogenase for Exhaled Acetaldehyde
    Iitani, Kenta
    Chien, Po-Jen
    Suzuki, Takuma
    Toma, Koji
    Arakawa, Takahiro
    Iwasaki, Yasuhiko
    Mitsubayashi, Kohji
    ACS SENSORS, 2018, 3 (02): : 425 - 431
  • [5] Optical bio-sniffer (biochemical gas sensor) for dimethyl sulfide vapor
    Mitsubayashi, K
    Hashimoto, Y
    Kon, T
    Minamide, T
    Taniguchi, M
    Saito, H
    Transducers '05, Digest of Technical Papers, Vols 1 and 2, 2005, : 2147 - 2150
  • [6] Fiber Optic Bio-sniffer (Biochemical Gas Sensor) using UV-LED light for Monitoring Ethanol Vapor with High Sensitivity & Selectivitiy
    Kudo, H.
    Miyajima, K.
    Takahashi, D.
    Arakawa, T.
    Saito, H.
    Mitsubayashi, K.
    Sawai, M.
    2009 IEEE SENSORS, VOLS 1-3, 2009, : 1956 - 1956
  • [7] Ultra-Sensitive Isopropanol Biochemical Gas Sensor (Bio-Sniffer) for Monitoring of Human Volatiles
    Chien, Po-Jen
    Suzuki, Takuma
    Ye, Ming
    Toma, Koji
    Arakawa, Takahiro
    Iwasaki, Yasuhiko
    Mitsubayashi, Kohji
    SENSORS, 2020, 20 (23) : 1 - 13
  • [8] Biochemical gas-sensor (bio-sniffer) for breath analysis after drinking
    Mitsubayashi, K
    Matsunaga, H
    Nishio, G
    Ogawa, M
    Saito, H
    SICE 2004 ANNUAL CONFERENCE, VOLS 1-3, 2004, : 97 - 100
  • [9] NADH-fluorometric biochemical gas sensor (bio-sniffer) for assessment of indoor air quality
    Kudo, Hiroyuki
    Yamashita, Toshifumi
    Ye, Ming
    Miyajima, Kumiko
    Arakawa, Takahiro
    Mitsubayashi, Kohji
    Gessei, Tomoko
    2012 IEEE SENSORS PROCEEDINGS, 2012, : 919 - 922
  • [10] NADH-fluorometric Biochemical Gas Sensor (Bio-Sniffer) for Evaluation of Indoor Air Quality
    Kudo, Hiroyuki
    Yamashita, Toshifumi
    Miyajima, Kumiko
    Arakawa, Takahiro
    Mitsubayashi, Kohji
    IEEE SENSORS JOURNAL, 2013, 13 (08) : 2828 - 2833