Improvement of foot pedals used during surgery based on new ergonomic guidelines

被引:41
|
作者
van Veelen, MA
Snijders, CJ
van Leeuwen, E
Goossens, RHM
Kazemier, G
机构
[1] Delft Univ Technol, Fac Ind Design Engn, NL-2628 CE Delft, Netherlands
[2] Erasmus Med Ctr Rotterdam, Dept Surg, NL-3000 DR Rotterdam, Netherlands
关键词
foot pedals; ergonomics; laparoscopy; posture; guidelines;
D O I
10.1007/s00464-002-9185-z
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: This study aims to create new ergonomic guidelines for the design of foot pedals used during surgery. Methods: Observations in the operating room, a questionnaire among 45 laparoscopic surgeons/residents, an ergonomic literature study, and clog measurements were used to assess the problems occurring during use and to compile new guidelines for foot pedals. Based on these guidelines a new foot pedal was designed and a prototype was manufactured. Results: During the surgical procedure 91% of the subjects occasionally loses contact with the foot pedal, which 56% experience as very annoying. All subjects think that the current foot pedals obstruct their freedom of movement; 75% occasionally hit the wrong switch and 53% experience physical discomfort in their legs and/or feet. Therefore, 93% of the subjects would like to control the diathermy in a different way. The new prototype of a foot pedal was evaluated in a pilot test and proved to be ergonomically better than the currently used foot pedals. Conclusion: The new guidelines for foot pedals result in an ergonomic improvement in their design.
引用
下载
收藏
页码:1086 / 1091
页数:6
相关论文
共 36 条
  • [1] Improvement of foot pedals used during surgery based on new ergonomic guidelines
    M. A. van Veelen
    C. J. Snijders
    E. van Leeuwen
    R. H. M. Goossens
    G. Kazemier
    Surgical Endoscopy And Other Interventional Techniques, 2003, 17 : 1086 - 1091
  • [2] Improvement of the laparoscopic needle holder based on new ergonomic guidelines
    M. A. van Veelen
    D. W. Meijer
    I. Uijttewaal
    R. H. M. Goossens
    C. J. Snijders
    G. Kazemier
    Surgical Endoscopy And Other Interventional Techniques, 2003, 17 : 699 - 703
  • [3] Improvement of the laparoscopic needle holder based on new ergonomic guidelines
    van Veelen, MA
    Meijer, DW
    Uijttewaal, I
    Goossens, RHM
    Snijders, CJ
    Kazemier, G
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2003, 17 (05): : 699 - 703
  • [4] Application of ergonomic guidelines during minimally invasive surgery: a questionnaire survey of 284 surgeons
    Wauben, L. S. G. L.
    van Veelen, M. A.
    Gossot, D.
    Goossens, R. H. M.
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2006, 20 (08): : 1268 - 1274
  • [5] Application of ergonomic guidelines during minimally invasive surgery: a questionnaire survey of 284 surgeons
    L. S. G. L. Wauben
    M. A. van Veelen
    D. Gossot
    R. H. M. Goossens
    Surgical Endoscopy And Other Interventional Techniques, 2006, 20 : 1268 - 1274
  • [6] Incorporation of Bovine-based Structural Bone Grafts Used in Reconstructive Foot Surgery
    Shibuya, Naohiro
    Jupiter, Daniel C.
    Clawson, Lacey D.
    La Fontaine, Javier
    JOURNAL OF FOOT & ANKLE SURGERY, 2012, 51 (01): : 30 - 33
  • [7] Ergonomic assessment of a new hand tool design for laparoscopic surgery based on surgeons' muscular activity
    Gonzalez, A. G.
    Barrios-Muriel, J.
    Romero-Sanchez, F.
    Salgado, D. R.
    Alonso, J.
    APPLIED ERGONOMICS, 2020, 88
  • [8] Medications Used During Improvement of Headache in Pediatric New Daily Persistent Headache
    McCracken, E.
    Goucher, O.
    Strelzik, J.
    Langdon, R.
    DiSabella, M.
    ANNALS OF NEUROLOGY, 2023, 94 : S81 - S81
  • [9] A New Approach to Quantifying Muscular Fatigue Using Wearable EMG Sensors during Surgery: An Ergonomic Case Study
    Merbah, Johan
    Care, Bertrand R.
    Gorce, Philippe
    Gadea, Francois
    Prince, Francois
    SENSORS, 2023, 23 (03)
  • [10] Dynamic Performance Improvement based on a New Parameter Estimation Method for IPMSM used for HEVs
    Wang, Aimeng
    Zhang, Li
    Dong, Shuhui