Validation of a smartphone embedded inertial measurement unit for measuring postural stability in older adults

被引:15
|
作者
De Groote, Friedl [1 ]
Vandevyvere, Stefanie [2 ]
Vanhevel, Florian [2 ]
De Xivry, Jean-Jacques Orban [1 ,3 ]
机构
[1] Katholieke Univ Leuven, Dept Movement Sci, B-3000 Leuven, Belgium
[2] Katholieke Univ Leuven, Fac Rehabil & Movement Sci, B-3000 Leuven, Belgium
[3] Katholieke Univ Leuven, Leuven Brain Inst, B-3000 Leuven, Belgium
关键词
Mobile technology; Standing balance; Force plate; FORCE-PLATE; FALL RISK; BALANCE; ACCELEROMETRY; POSTUROGRAPHY; RELIABILITY;
D O I
10.1016/j.gaitpost.2020.11.017
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Background: Identifying older adults with increased fall risk due to poor postural control on a large scale is only possible through omnipresent and low cost measuring devices such as the inertial measurement units (IMU) embedded in smartphones. However, the correlation between smartphone measures of postural stability and state-of-the-art force plate measures has never been assessed in a large sample allowing us to take into account age as a covariate. Research question: How reliably can postural stability be measured with a smartphone embedded IMU in comparison to a force plate? Methods: We assessed balance in 97 adults aged 50-90 years in four different conditions (eyes open, eyes closed, semi-tandem and dual-task) in the anterio-posterior and medio-lateral directions. We used six different parameters (root mean square and average absolute value of COP displacement, velocity and acceleration) for the force plate and two different parameters (root mean square and average absolute value of COM acceleration) for the smartphone. Results: Test-retest reliability was smaller for the smartphone than for the force plate (intra class correlation) but both devices could equally well detect differences between conditions (similar Cohen's d). Parameters from the smartphone and the force plate, with age regressed out, were moderately correlated (robust correlation coefficients of around 0.5). Significance: This study comprehensively documents test-retest reliability and effect sizes for stability measures obtained with a force plate and smartphone as well as correlations between force plate and smartphone measures based on a large sample of older adults. Our large sample size allowed us to reliably determine the strength of the correlations between force plate and smartphone measures. The most important practical implication of our results is that more repetitions or longer trials are required when using a smartphone instead of a force plate to assess balance.
引用
下载
收藏
页码:17 / 23
页数:7
相关论文
共 50 条
  • [1] Assessment of postural stability using inertial measurement unit on inclined surfaces in healthy adults
    Frames, Chris
    Soangra, Rahul
    Lockhart, Thurmon E.
    Biomedical Sciences Instrumentation, 2013, 49 : 234 - 242
  • [2] Construct Validity of a Wearable Inertial Measurement Unit (IMU) in Measuring Postural Sway and the Effect of Visual Deprivation in Healthy Older Adults
    Ferrari, Luca
    Bochicchio, Gianluca
    Bottari, Alberto
    Scarton, Alessandra
    Lucertini, Francesco
    Pogliaghi, Silvia
    Biosensors, 2024, 14 (11):
  • [3] Coherence and reliability of a wearable inertial measurement unit for measuring postural sway
    Hansson, Eva Ekvall
    Tornberg, Asa
    BMC RESEARCH NOTES, 2019, 12 (1)
  • [4] Coherence and reliability of a wearable inertial measurement unit for measuring postural sway
    Eva Ekvall Hansson
    Åsa Tornberg
    BMC Research Notes, 12
  • [5] Blood Pressure Measurement Based on the Camera and Inertial Measurement Unit of a Smartphone: Instrument Validation Study
    Yoon, Yong-Hoon
    Kim, Jongin
    Lee, Kwang Jin
    Cho, Dongrae
    Oh, Jin Kyung
    Kim, Minsu
    Roh, Jae-Hyung
    Park, Hyun Woong
    Lee, Jae-Hwan
    JMIR MHEALTH AND UHEALTH, 2023, 11
  • [6] Smartphone technology can measure postural stability and discriminate fall risk in older adults
    Hsieh, Katherine L.
    Roach, Kathleen L.
    Wajda, Douglas A.
    Sosnoff, Jacob J.
    GAIT & POSTURE, 2019, 67 : 160 - 165
  • [7] Fall Risk Assessments Based on Postural and Dynamic Stability Using Inertial Measurement Unit
    Liu, Jian
    Zhang, Xiaoyue
    Lockhart, Thurmon E.
    SAFETY AND HEALTH AT WORK, 2012, 3 (03) : 192 - 198
  • [8] Measuring postural stability with an inertial sensor: validity and sensitivity
    Neville, Christopher
    Ludlow, Caleb
    Rieger, Brian
    MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2015, 8 : 447 - 454
  • [9] Locomotive Syndrome Assessment in Older Adults Using a Single Inertial Measurement Unit
    Hosseini, Iman
    Ghahramani, Maryam
    2023 IEEE INTERNATIONAL SYMPOSIUM ON INERTIAL SENSORS AND SYSTEMS, INERTIAL, 2023,
  • [10] Gait Analysis Based on an Inertial Measurement Unit Sensor: Validation of Spatiotemporal Parameters Calculation in Healthy Young and Older Adults
    Sugiarto, Tommy
    Lin, Yi-Jia
    Chang, Chao-Chin
    Hsu, Wei-Chun
    2017 IEEE/SICE INTERNATIONAL SYMPOSIUM ON SYSTEM INTEGRATION (SII), 2017, : 517 - 522