Experimental verification and clinical implementation of a commercial Monte Carlo electron beam dose calculation algorithm

被引:25
|
作者
Fragosoa, Margarida [1 ]
Pillai, Sushakumari [1 ]
Solberg, Timothy D. [2 ]
Chetty, Indrin J. [1 ]
机构
[1] Univ Nebraska, Med Ctr, Omaha, NE 68198 USA
[2] Univ Texas SW Med Ctr Dallas, Dallas, TX 75390 USA
关键词
Monte Carlo; treatment planning algorithms; electron beam;
D O I
10.1118/1.2839098
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study describes the modeling and the experimental verification and clinical implementation of the alpha release of Pinnacle(3) Monte Carlo (MC) electron beam dose calculation algorithm for patient-specific treatment planning. The MC electron beam modeling was performed for beam energies ranging from 6 to 18 MeV from a Siemens (Primus) linear accelerator using standard-shaped electron applicators and 100 cm source-to-surface distance (SSD). The agreement between MC calculations and measurements was, on average, within 2% and 2 mm for all applicator sizes. However, differences of the order of 3%-4% were noted in the off-axis dose profiles for the largest applicator modeled and for all energies. Output factors were calculated for standard electron cones and square cutouts inserted in the 10x10 cm(2) applicator for different SSDs and were found to be within 4% of measured data. Experimental verification of the MC electron beam model was carried out using an ionization chamber and film in solid-water slab and anthropomorphic phantoms containing bone and lung materials. Agreement between calculated and measured dose distributions was within +/- 3%. Clinical comparison was performed in four patient treatment plans with lesions in highly irregular anatomies, such as the ear, face, and breast, where custom-designed bolus and field shaping blocks were used in the patient treatments. For comparison purposes, treatment planning was also performed using the conventional pencil beam (PB) algorithm with the Pinnacle(3) treatment planning system. Differences between MC and PB dose calculations for the patient treatment plans were significant, particularly in anatomies where the target was in close proximity to low density tissues, such as lung and air cavities. Concerning monitor unit calculations, the largest differences obtained between MC and PB algorithms were between 4.0% and 5.0% for two patients treated with oblique beams and involving highly irregular surfaces, i.e., breast and cheek. Clinical results are reported for overall uncertainty values (averaged over voxels with doses >50% dose(max)) ranging from 2% to 0.3% and calculations were performed using cubic voxels with side 0.3 cm. Timing values ranged from 2 min to 24.5 h, depending on the field size, beam energy, number, and thickness of computed tomography slices used to define the patient's anatomy for the overall uncertainty values mentioned above. (c) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:1028 / 1038
页数:11
相关论文
共 50 条
  • [1] CLINICAL IMPLEMENTATION OF A MACRO MONTE CARLO ELECTRON DOSE CALCULATION ALGORITHM
    Swinnen, A.
    Petillion, S.
    Van den Heuvel, F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S69 - S70
  • [2] Evaluation of a Commercial Monte Carlo Dose Calculation Algorithm for Electron Beam Treatment Planning
    Chamberland, E.
    Lachance, B.
    Beaulieu, L.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [3] Verification of a commercial implementation of the Macro-Monte-Carlo electron dose calculation algorithm using the virtual accelerator approach
    Tertel, Jakob
    Wulff, Joerg
    Karle, Heiko
    Zink, Klemens
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2010, 20 (01): : 51 - 60
  • [4] Evaluation of a commercial macro Monte Carlo electron dose calculation algorithm
    Popple, R
    Weinberg, R
    Antolak, J
    Brezovich, I
    Duan, J
    Pareek, P
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2016 - 2016
  • [5] Clinical implementation and validation of a commercial Monte Carlo dose calculation system
    Chuang, C
    Curran, B
    Verhey, L
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1453 - 1453
  • [6] Evaluation of an electron Monte Carlo dose calculation algorithm for electron beam radiotherapy
    Hu, A.
    Song, H.
    Chen, Z.
    Zhou, S.
    Yin, F.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2099 - 2099
  • [7] A virtual-acceterator-based verification of a Monte Carlo dose calculation algorithm for electron beam treatment planning in clinical situations
    Wieslander, Elinore
    Knoos, Tommy
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 82 (02) : 208 - 217
  • [8] Evaluation of a Commercial Monte Carlo Dose Calculation Algorithm for Electron Treatment Planning
    Huang, J.
    Dunkerley, D.
    Smilowitz, J.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E546 - E546
  • [9] Evaluation of a commercial Monte Carlo dose calculation algorithm for electron treatment planning
    Huang, Jessie Y.
    Dunkerley, David
    Smilowitz, Jennifer B.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (06): : 184 - 193
  • [10] Evaluation of a new commercial Monte Carlo dose calculation algorithm for electron beams
    Vandervoort, Eric J.
    Tchistiakova, Ekaterina
    La Russa, Daniel J.
    Cygler, Joanna E.
    [J]. MEDICAL PHYSICS, 2014, 41 (02) : 021711