Transverse connectors providing increased stability to the cervical spine rod-screw construct: an in vitro human cadaveric study Laboratory investigation

被引:12
|
作者
Shaw, Miranda N. [2 ]
Morel, Etienne P. [2 ]
Utter, P. Andrew
Gussous, Yazeed M. [2 ]
Ginoux, Lise [2 ]
Berglund, Lawrence J. [2 ]
Gay, Ralph E. [2 ,3 ]
Krauss, William E. [1 ]
机构
[1] Mayo Clin, Coll Med, Dept Neurol Surg, Rochester, MN 55905 USA
[2] Mayo Clin, Biomech Lab, Rochester, MN 55905 USA
[3] Mayo Clin, Dept Phys Med & Rehabil, Rochester, MN 55905 USA
关键词
cervical spine fusion; lateral mass fixation; transverse connector; biomechanics; LOAD-DISPLACEMENT CURVES; BIOMECHANICAL EVALUATION; CERVICOTHORACIC JUNCTION; IDIOPATHIC SCOLIOSIS; INTERNAL-FIXATION; PEDICLE SCREWS; TITANIUM WIRE; LUMBAR SPINE; POSTERIOR; INSTRUMENTATION;
D O I
10.3171/2011.1.SPINE10411
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. The object of this study was to determine if the addition of transverse connectors (TCs) to a rod-screw construct leads to increased stabilization of the cervical spine. Methods. Eleven human cadaveric cervical spines (C2-T1) were used to examine the effect of adding connectors to a C3-7 rod-screw construct in 3 models of instability: 1) C3-6 wide laminectomy, 2) wide laminectomy and 50% foraminotomy at C4-5 and C5-6, and 3) wide laminectomy with full medial to lateral foraminotomy. Following each destabilization procedure, specimens were tested with no TC, 1 TC between the C-5 screws, and 2 TCs between the C-4 and C-6 screws. Testing of the connectors was conducted in random order. Specimens were subjected to +/- 2 Nm of torque in flexion and extension, lateral bending, and axial rotation. Range of motion was determined for each experimental condition. Statistical comparisons were made between the destabilized and intact conditions, and between the addition of TCs and the absence of TCs. Results. The progressive destabilization procedures significantly increased motion. The addition of TCs did not significantly change motion in flexion and extension. Lateral bending was significantly decreased with 2 connectors, but not with 1 connector. The greatest effect was on axial rotation. In general, 2 TCs were more restrictive than 1 TC, and decreased motion 10% more than fixation alone. Conclusions. Regardless of the degree of cervical destabilization, 1 or 2 TCs decreased motion compared with rods and screws alone. Axial rotation was most affected. Transverse connectors effectively increase the rigidity of rod-screw constructs in the cervical spine. Severe cervical instability can be overcome with the use of 2 TCs, but in cases in which 2 cannot be used, 1 should be adequate and superior to none. (DOI: 10.3171/2011.1.SPINE10411)
引用
收藏
页码:719 / 725
页数:7
相关论文
共 6 条
  • [1] Biomechanical evaluation of a new modular rod-screw implant system for posterior instrumentation of the occipito-cervical spine: in-vitro comparison with two established implant systems
    M. Richter
    H.-J. Wilke
    P. Kluger
    S. Neller
    L. Claes
    W. Puhl
    European Spine Journal, 2000, 9 : 417 - 425
  • [2] Biomechanical evaluation of a novel atlas polyaxial transverse connecting screw system, an in vitro human cadaveric study
    Chen, Yun-Lin
    Lu, Wen-Jie
    Ma, Yu-Li
    Ma, Wei-Hu
    Jiang, Wei-Yu
    MEDICINE, 2023, 102 (47) : E36312
  • [3] Biomechanical evaluation of a new modular rod-screw implant system for posterior instrumentation of the occipito-cervical spine: in-vitro comparison with two established implant systems
    Richter, M
    Wilke, HJ
    Kluger, P
    Neller, S
    Claes, L
    Puhl, W
    EUROPEAN SPINE JOURNAL, 2000, 9 (05) : 417 - 425
  • [4] A comparative biomechanical study of a novel integrated plate spacer for stabilization of cervical spine: An in vitro human cadaveric model
    Majid, Kamran
    Chinthakunta, Suresh
    Muzumdar, Aditya
    Khalil, Saif
    CLINICAL BIOMECHANICS, 2012, 27 (06) : 532 - 536
  • [5] Initial stability of one-stage anterior debridement and cage implantation combined with anterior-lateral fixation by a dual screw-rod construct in the treatment of lumbosacral tuberculosis: a cadaveric biomechanical study
    Liu, Jiantao
    Gao, Yanzheng
    Gao, Zhengchao
    Niu, Binbin
    Lv, Dongbo
    Yang, Yin
    BMC MUSCULOSKELETAL DISORDERS, 2019, 20 (1)
  • [6] Initial stability of one-stage anterior debridement and cage implantation combined with anterior-lateral fixation by a dual screw-rod construct in the treatment of lumbosacral tuberculosis: a cadaveric biomechanical study
    Jiantao Liu
    Yanzheng Gao
    Zhengchao Gao
    Binbin Niu
    Dongbo Lv
    Yin Yang
    BMC Musculoskeletal Disorders, 20