Effective Masking Levels for Bone Conduction Auditory Steady State Responses in Infants and Adults With Normal Hearing

被引:5
|
作者
Hansen, Erin E. [1 ]
Small, Susan A. [1 ]
机构
[1] Univ British Columbia, Sch Audiol & Speech Sci, Vancouver, BC V6T 1Z3, Canada
来源
EAR AND HEARING | 2012年 / 33卷 / 02期
基金
加拿大自然科学与工程研究理事会;
关键词
BRAIN-STEM RESPONSES; TUBE MICROPHONE MEASURES; REAL-EAR; LONGITUDINAL CHANGES; YOUNG INFANTS; TONES; THRESHOLDS;
D O I
10.1097/AUD.0b013e31822f67f6
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: To obtain ear-specific bone conduction thresholds, masking of the nontest ear is often required. Bone conduction masking has not been formally investigated for infants assessed physiologically. The objective of this study was to determine effective masking levels (EMLs) for auditory steady state responses (ASSRs) elicited by bone-conducted stimuli in a group of normal-hearing infants and adults. Design: Participants were 15 infants younger than 6 mo and 15 adults, all with normal hearing. EML was defined as the lowest level of a binaural air-conducted masker that resulted in absent bone conduction ASSRs. Stimuli were single bone-conducted tones that were 100% amplitude modulated and 25% frequency modulated at 85 and 101 for 1000 and 4000 Hz, respectively. The stimuli were calibrated in dB HL (ANSI S3.6-1996) and expressed in dB HL or dB SL (dB relative to mean bone conduction ASSR thresholds reported in a previous study). The maskers were 1 and 4 kHz narrowband noise generated by a clinical audiometer. Unmasked and masked ASSRs were obtained for each participant. Real ear-to-coupler differences (RECDs) were also obtained for each participant and were used to convert masker dB SPL measured in the coupler to dB SPL in the individual ear canal. Results: Infant EMLs for ASSRs elicited to bone-conducted stimuli in dB HL were 6 to 7 dB higher and 8 to 10 dB lower for 1000 and 4000 Hz, respectively, compared with adults. When masker was adjusted for RECDs, infant EMLs were 12 dB higher at 1000 Hz and similar at 4000 Hz compared with adults. When the stimulus levels were corrected for the mean differences in ASSR bone conduction thresholds between infants and adults and the masker levels adjusted for RECDs, infants had lower EMLs at 1000 Hz and equal EMLs at 4000 Hz, in comparison to adults. Frequency-and level-dependent effects on ASSR amplitude due to masking were found and differed between infants and adults. Conclusions: Our findings indicate that there are frequency-and level-dependent infant-adult differences in EMLs for bone conduction ASSRs and confirm that a 1000 Hz stimulus is 12 dB more effective for infants compared with adults. The following infant preliminary masking levels for bone conduction stimuli are recommended: (i) 1000 Hz: 48 and 58 dB SPL at 15 and 25 dB HL, respectively, and (ii) 4000 Hz: 40 and 45 dB SPL at 25 and 35 dB HL, respectively.
引用
收藏
页码:269 / 278
页数:10
相关论文
共 50 条
  • [1] Effective Masking Levels for 500 and 2000 Hz Bone Conduction Auditory Steady State Responses in Infants and Adults With Normal Hearing
    Small, Susan A.
    Smyth, Aisling
    Leon, Griselle
    [J]. EAR AND HEARING, 2014, 35 (01): : 63 - 71
  • [2] Effective Masking Levels for Bone Conduction Auditory Brainstem Response Stimuli in Infants and Adults with Normal Hearing
    Lau, Ricky
    Small, Susan A.
    [J]. EAR AND HEARING, 2021, 42 (02): : 443 - 455
  • [3] Comparisons of Auditory Steady State Response and Behavioral Air Conduction and Bone Conduction Thresholds for Infants and Adults With Normal Hearing
    Casey, Kelly-Ann
    Small, Susan A.
    [J]. EAR AND HEARING, 2014, 35 (04): : 423 - 439
  • [4] Maturation of the Occlusion Effect: A Bone Conduction Auditory Steady State Response Study in Infants and Adults With Normal Hearing
    Small, Susan A.
    Hu, Ning
    [J]. EAR AND HEARING, 2011, 32 (06): : 708 - 719
  • [5] Auditory steady-state responses to bone conduction stimuli in children with hearing loss
    Swanepoel, De Wet
    Ebrahim, Shamim
    Friedland, Peter
    Swanepoel, Andre
    Pottas, Lidia
    [J]. INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2008, 72 (12) : 1861 - 1871
  • [6] Multiple auditory steady-state response thresholds to bone-conduction stimuli in young infants with normal hearing
    Small, Susan A.
    Stapells, David R.
    [J]. EAR AND HEARING, 2006, 27 (03): : 219 - 228
  • [7] Multiple auditory steady-state responses in children and adults with normal hearing, sensorineural hearing loss, or auditory neuropathy
    Attias, J
    Buller, N
    Rubel, Y
    Raveh, E
    [J]. ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2006, 115 (04): : 268 - 276
  • [8] FREQUENCY-SPECIFIC AUDITORY BRAIN-STEM RESPONSES - EFFECTIVE MASKING LEVELS AND RELATIONSHIP TO BEHAVIORAL THRESHOLDS IN NORMAL HEARING ADULTS
    PURDY, SC
    HOUGHTON, JM
    KEITH, WJ
    GREVILLE, KA
    [J]. AUDIOLOGY, 1989, 28 (02): : 82 - 91
  • [9] Multiple auditory steady state evoked responses in normal hearing children
    Douniadakis, D.
    Florou, V.
    Psarommatis, I.
    Tsakanikos, M.
    Karagiannidou, S.
    Tsilis, N.
    Apostolopoulos, N.
    [J]. PROCEEDINGS OF THE 8TH INTERNATIONAL CONGRESS OF MEDITERRANEAN SOCIETY OF OTOLOGY & AUDIOLOGY, 2006, : 7 - +
  • [10] Maturation of bone conduction multiple auditory steady-state responses
    Small, Susan Anne
    Stapells, David Richard
    [J]. INTERNATIONAL JOURNAL OF AUDIOLOGY, 2008, 47 (08) : 476 - 488