Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery

被引:136
|
作者
Cosgrove, VP [1 ]
Jahn, U [1 ]
Pfaender, M [1 ]
Bauer, S [1 ]
Budach, V [1 ]
Wurm, RE [1 ]
机构
[1] Free Univ Berlin, Klinikum Charite, Klin Strathlentherapie, D-10117 Berlin, Germany
关键词
micro multi-leaf collimator; MLC; conformal therapy; dosimetry; radiosurgery planning;
D O I
10.1016/S0167-8140(99)00020-1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A computer controlled micro multi-leaf collimator, m3 mMLC, has been commissioned for conformal, fixed-held radiosurgery applications. Measurements were made to characterise the basic dosimetric properties of the m3, such as leaf transmission, leakage and beam penumbra. In addition, the geometric and dosimetric accuracy of the m3 was verified when used in conjunction with a BrainSCAN v3.5 stereotactic planning system. Materials and methods: The m3 was detachably mounted to a Varian Clinac 2100C accelerator delivering 6 MV X-rays. Leaf transmission, leakage, penumbra and multiple, conformal fixed field dose distributions were measured using calibrated film in solid water. Beam data were collected using a diamond detector in a scanning water tank and planned dose distributions were verified using LIF TLDs and film. A small, shaped phantom was also constructed to confirm field shaping accuracy using portal images. Results: Mean transmission through the closed multi-leaves was 1.9 +/- 0.1% and leakage between leaves was 2.8 +/- 0.15%. Between opposing leaves abutting along the central beam-axis transmission was similar to 15 +/- 3%, but was reduced to a mean of 4.5 +/- 0.6% by moving the abutmen position 4.5 cm off-axis. Beam penumbrae were effectively constant as a function of increasing square field size and asymmetric fields and was seen to vary non-linearly when shaped to diagonal, straight edges. TMR, OAR and relative output beam data measurements of circular m3 fields were comparable to conventional, circular stereotactic collimators. Multiple, conformal field dose distributions were calculated with good spatial and dosimetric accuracy, with the planned 90% isodose curves agreeing with measurements to within 1-2 mm and to +/- 3% at isocentre. Portal films agreed with planned beams eye-view field shaping to within 1 mm. Conclusions: The m3 micro multi-leaf collimator is a stable, high precision field-shaping device suitable for small-field, radiosurgery applications. Dose distributions can be accurately calculated by a planning system using only a few beam data parameters. (C) 1999 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:325 / 336
页数:12
相关论文
共 50 条
  • [1] Commissioning and clinical use of a micro multi-leaf collimator for conformal radiosurgery
    Schlenger, L
    Bohsung, J
    Cosgrove, VP
    Groll, J
    Jahn, U
    Kaiser, A
    Pfaender, M
    Stuschke, M
    Budach, V
    Wurm, RE
    EUROPEAN JOURNAL OF CANCER, 1999, 35 : S94 - S94
  • [2] Dosimetric Characteristics of TrueBeam STx System with High Definition Multi-Leaf Collimator: Commissioning of Stereotactic Radiosurgery
    Chang, Z.
    Wu, Q.
    Yan, H.
    Yang, Y.
    Giles, W.
    Busselberg, C.
    Lipscomb, R.
    Yin, F.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Commissioning of the Elekta APEX dynamic micro multi-leaf collimator
    Ferrer, C.
    Plaza, R.
    Huertas, C.
    Aza, Z.
    Morera, R.
    Serrada, A.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1197 - S1197
  • [4] Physics characteristics and dose comparison for micro multi-leaf collimator and standard multi-leaf collimator in whole body conformal radiosurgery
    Yang, C
    Chen, H
    Gollamudi, S
    Raben, A
    MEDICAL PHYSICS, 2002, 29 (06) : 1212 - 1212
  • [5] Commissioning of mini-multi-leaf-collimator (MMLC) for stereotactic radiosurgery and radiotherapy
    Mardirossian, George
    Urie, Marcia
    Fitzgerald, Thomas J.
    Mayr, Nina
    Montebello, Joseph
    Lo, Yuan-Chynan
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2003, 11 (01) : 21 - 31
  • [6] Commissioning of multi-leaf collimator for fast neutron therapy
    Yudelev, M.
    Mogili, N.
    Hwang, T.
    Farr, J.
    Maughan, R.
    MEDICAL PHYSICS, 2006, 33 (06) : 2080 - 2080
  • [7] Dosimetric Impact of Systematic and Random Errors With Multi-Leaf Collimator Leaf Positioning in Intracranial Stereotactic Radiosurgery
    Han, C.
    Liu, J. R.
    Du, D.
    Liang, J.
    Qing, K.
    Watkins, W. T.
    Zhang, S.
    Liu, A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E545 - E546
  • [8] Commissioning of a micro-multileaf collimator for conformal stereotactic radiosurgery and radiotherapy
    Wurm, RE
    Cosgrove, VP
    Schlenger, L
    Kaiser, A
    Zeunert, G
    Dinges, S
    Bohsung, J
    Groll, J
    Pfaender, M
    Bauer, S
    Leonhardt, G
    Jahn, U
    Stuschke, A
    Budach, V
    CONTROVERSIES IN NEURO-ONCOLOGY, 1999, 33 : 64 - 77
  • [9] Characterization of a dynamic multi-leaf collimator for stereotactic radiotherapy applications
    Godwin, G. A.
    Simpson, J. B.
    Mugabe, K. V.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (14): : 4643 - 4654
  • [10] Dosimetric Comparison of Robotic and Linear Accelerator Multi-Leaf Collimator-Based Stereotactic Radiosurgery for Arteriovenous Malformation
    Kaliyaperumal, Venkatesan
    Abraham, Susan
    Veni, Maragatha
    Banerjee, Susovan
    Tamilselvan, S.
    Gupta, Deepak
    Dayanithi, K.
    Manigandan, D.
    Mishra, Saumyaranjan
    Bisht, Shyam Singh
    Kataria, Tejinder
    JOURNAL OF MEDICAL PHYSICS, 2021, 46 (01) : 16 - 25