Use of Image-free Navigation in Determination of Acetabular Cup Orientation: Analysis of Factors Affecting Precision

被引:8
|
作者
Fukui, Tomokazu [1 ]
Fukunishi, Shigeo [1 ]
Nishio, Shoji [1 ]
Shibanuma, Nao [2 ]
Yoshiya, Shinichi [1 ]
机构
[1] Hyogo Coll Med, Dept Orthopaed Surg, Nishinomiya, Hyogo 6638501, Japan
[2] Kobe Kaisei Hosp, Dept Orthopaed Surg, Kobe, Hyogo, Japan
关键词
TOTAL HIP-ARTHROPLASTY; COMPONENT; ACCURACY; ALIGNMENT; SYSTEM;
D O I
10.3928/01477447-20100510-61
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We have been using an image-free total hip arthroplasty (THA) navigation system (OrthoPilot; Aesculap, Tuttlingen, Germany) to ensure accurate and reproducible acetabular cup orientation. In this study, the accuracy of this system in the assessment of acetabular cup orientation was evaluated by comparing the intraoperative inclination and anteversion angles presented by the navigation system and the corresponding postoperative values obtained by computed tomography measurement. In the intraoperative accuracy analysis, we additionally examined the influence of factors such as body mass index and soft tissue thickness on assessment error. Intraoperative and postoperative results obtained from 115 consecutive navigated THAs were compared and analyzed. In both inclination and anteversion angles, good agreement was observed; a discrepancy of more than 5 degrees was observed in 1 and 3 cases, respectively. In the analysis of factors potentially affecting the accuracy of the intraoperative assessment, no correlation between each parameter and the intraoperative and postoperative discrepancy was demonstrated.
引用
收藏
页码:38 / 42
页数:5
相关论文
共 20 条
  • [1] Greater accuracy in positioning of the acetabular cup by using an image-free navigation system
    Kalteis, T
    Handel, M
    Herold, T
    Perlick, L
    Baethis, H
    Grifka, J
    INTERNATIONAL ORTHOPAEDICS, 2005, 29 (05) : 272 - 276
  • [2] Greater accuracy in positioning of the acetabular cup by using an image-free navigation system
    T. Kalteis
    M. Handel
    T. Herold
    L. Perlick
    H. Baethis
    J. Grifka
    International Orthopaedics, 2005, 29 : 272 - 276
  • [3] The accuracy of acetabular cup placement in primary total hip arthroplasty using an image-free navigation system
    Naito, Yohei
    Hasegawa, Masahiro
    Tone, Shine
    Wakabayashi, Hiroki
    Sudo, Akihiro
    BMC MUSCULOSKELETAL DISORDERS, 2021, 22 (01)
  • [4] The accuracy of acetabular cup placement in primary total hip arthroplasty using an image-free navigation system
    Yohei Naito
    Masahiro Hasegawa
    Shine Tone
    Hiroki Wakabayashi
    Akihiro Sudo
    BMC Musculoskeletal Disorders, 22
  • [5] Accuracy of an image-free cup navigation system -: an anatomical study
    Kalteis, T
    Beckmann, J
    Herold, T
    Zysk, S
    Bäthis, H
    Perlick, L
    Grifka, J
    BIOMEDIZINISCHE TECHNIK, 2004, 49 (09): : 257 - 262
  • [6] Comparison of the accuracies of computed tomography-based navigation and image-free navigation for acetabular cup insertion in total hip arthroplasty in the lateral decubitus position
    Hasegawa, Masahiro
    Tone, Shine
    Naito, Yohei
    Wakabayashi, Hiroki
    Sudo, Akihiro
    COMPUTER ASSISTED SURGERY, 2021, 26 (01) : 69 - 76
  • [7] Combined anteversion of the total hip arthroplasty implanted with image-free cup navigation and without stem navigation
    Fukunishi, Shigeo
    Fukui, Tomokazu
    Nishio, Shoji
    Fujihara, Yuki
    Okahisa, Shohei
    Yoshiya, Shinichi
    ORTHOPEDIC REVIEWS, 2012, 4 (04) : 146 - 149
  • [8] Computed Tomography Assessment of Image-free Navigation-assisted Cup Placement in THA in an Asian Population
    Kim, Tae-Hoon
    Lee, Suk-Ha
    Yang, Jae-Hyuk
    Oh, Kwang-Jun
    ORTHOPEDICS, 2012, 35 (10) : 13 - 17
  • [9] Three-dimensional analysis of image-free navigation system for total knee arthroplasty
    Tigani D.
    Rimondi E.
    Trentani P.
    Ansaloni M.
    Amendola L.
    Testi D.
    MUSCULOSKELETAL SURGERY, 2011, 95 (2) : 81 - 87
  • [10] CT-free image-guided navigation system for acetabular cup placement in routine clinical practice
    von Recum, J
    Wendl, K
    Korber, J
    Wentzensen, A
    Grützner, PA
    UNFALLCHIRURG, 2003, 106 (11): : 929 - 934