Evaluation of the accuracy of a mobile and a stationary system for three-dimensional facial scanning

被引:90
|
作者
Modabber, Ali [1 ]
Peters, Florian [1 ]
Kniha, Kristian [1 ]
Goloborodko, Evgeny [1 ]
Ghassemi, Alireza [1 ]
Lethaus, Bernd [1 ]
Hoelzle, Frank [1 ]
Moehlhenrich, Stephan Christian [1 ]
机构
[1] Univ Hosp RWTH Aachen, Sch Med, Dept Oral Maxillofacial & Plast Facial Surg, Pauwelstr 30, D-52074 Aachen, Germany
关键词
3D-scanner; Accuracy; Artec EVA; FaceScan3D; COOLING THERAPY METHODS; OBSERVER-BLIND; ORTHOGNATHIC SURGERY; IMAGING-SYSTEMS; 3D EVALUATION; PRECISION; FACE; STEREOPHOTOGRAMMETRY;
D O I
10.1016/j.jcms.2016.08.008
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Numerous three-dimensional (3D) facial scanners have emerged on the market; however, publications evaluating their accuracies are sparse. In this study, the accuracy of two 3D scanners used in facial scanning was evaluated. Materials and methods: A test specimen was attached at the right cheek and the forehead of 41 volunteers. These volunteers were scanned with Artec EVA (R) and FaceScan3D (R). The acquired data were aligned to a 3D model of the test specimen for comparing the mean error, original length and width and angles to the measured values. Results: The mean error in Best Fit alignment is significantly lower using Artec EVA (p < 0.001) for both test specimens. The deviation from the original length and width is significantly lower for the test specimens (p < 0.01) when measured with Artec EVA. The aberration of the angles measured between the front plane and the side plane is significantly lower when measured with Artec EVA (p < 0.001). Captured with Artec EVA the discrepancy between the original angle and the angle measured between the side planes to each other is significantly lower (p < 0.01). Conclusions: Scanning with Artec EVA leads to more accurate 3D models as compared to scanning with FaceScan3D. The exactness achieved by both scanners is comparable to other scanners mentioned in literature. (C) 2016 European Association for Cranio-Maxillo-Facial Surgery. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1719 / 1724
页数:6
相关论文
共 50 条
  • [1] Trueness of One Stationary and Two Mobile Systems for Three-Dimensional Facial Scanning
    Wang, Can
    Shi, Yi-Fei
    Xiong, Qian
    Xie, Pei-Jin
    Wu, Jun-Hua
    Liu, Wei-Cai
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2022, 35 (03) : 350 - 356
  • [2] Accuracy of Smartphone-Based Three-Dimensional Facial Scanning System: A Systematic Review
    Luo, Yuna
    Zhao, Minghao
    Lu, Jianjian
    AESTHETIC PLASTIC SURGERY, 2024, 48 (21) : 4500 - 4512
  • [3] Evaluation of facial blood flow using three-dimensional scanning
    Bock, A.
    Suschek, C. V.
    Oplaender, C.
    Hoelzle, F.
    Modabber, A.
    Pallua, N.
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2017, 55 (09): : 974 - 976
  • [4] Quantitative evaluation of three-dimensional facial scanners measurement accuracy for facial deformity
    Zhao Yi-jiao
    Xiong Yu-xue
    Sun Yu-chun
    Yang Hui-fang
    Lyu Pei-jun
    Wang Yong
    INTERNATIONAL CONFERENCE ON OPTICAL AND PHOTONIC ENGINEERING (ICOPEN 2015), 2015, 9524
  • [5] Three-Dimensional Accuracy of Facial Scan for Facial Deformities in Clinics: A New Evaluation Method for Facial Scanner Accuracy
    Zhao, Yi-jiao
    Xiong, Yu-xue
    Wang, Yong
    PLOS ONE, 2017, 12 (01):
  • [6] Three-Dimensional Facial Scanning at the Fingertips of Patients and Surgeons: Accuracy and Precision Testing of iPhone X Three-Dimensional Scanner
    Rudy, Hayeem L.
    Wake, Nicole
    Yee, Judy
    Garfein, Evan S.
    Tepper, Oren M.
    PLASTIC AND RECONSTRUCTIVE SURGERY, 2020, 146 (06) : 1407 - 1417
  • [7] Assessment of facial asymmetry in growing subjects with a three-dimensional laser scanning system
    Primozic, J.
    Perinetti, G.
    Zhurov, A.
    Richmond, S.
    Ovsenik, M.
    ORTHODONTICS & CRANIOFACIAL RESEARCH, 2012, 15 (04) : 237 - 244
  • [8] Validation of a three-dimensional facial scanning system based on structured light techniques
    Ma, Lili
    Xu, Tianmin
    Lin, Jiuxiang
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2009, 94 (03) : 290 - 298
  • [9] Accuracy and precision of integumental linear dimensions in a three-dimensional facial imaging system
    Kim, Soo-Hwan
    Jung, Woo-Young
    Seo, Yu-Jin
    Kim, Kyung-A
    Park, Ki-Ho
    Park, Young-Guk
    KOREAN JOURNAL OF ORTHODONTICS, 2015, 45 (03) : 105 - 112
  • [10] Evaluation of anthropometric accuracy and reliability using different three-dimensional scanning systems
    Fourie, Zacharias
    Damstra, Janalt
    Gerrits, Peter O.
    Ren, Yijin
    FORENSIC SCIENCE INTERNATIONAL, 2011, 207 (1-3) : 127 - 134