Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A phantom study

被引:18
|
作者
Dabli, Djamel [1 ]
Frandon, Julien [1 ]
Belaouni, Asmaa [1 ]
Akessoul, Philippe [1 ]
Addala, Takieddine [1 ]
Berny, Laure [1 ]
Beregi, Jean-Paul [1 ]
Greffer, Joel [1 ]
机构
[1] Univ Montpellier, Dept Med Imaging, CHU Nimes, Med Imaging Grp Nimes,EA 2994, Nimes, France
关键词
Dual-energy CT; Iodine quantification; Monochromatic images; Iterative reconstruction; ENERGY CT; COMPUTED-TOMOGRAPHY; PERFORMANCE; DIAGNOSIS; CELL; KV;
D O I
10.1016/j.diii.2021.08.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to assess the impact of advanced modeled iterative reconstruction (ADMIRE) algorithm and dose levels on the accuracy of Hounsfield unit (HU) measurement, image noise and contrast-to-noise ratio (CNR) in virtual monochromatic images (VMIs) with low iodine concentrations, and the accuracy of iodine quantification. Materials and methods: A CT phantom was scanned with dual-source CT using abdomen-pelvis examination parameters at four dose levels: 5, 8, 11 and 20 mGy. Images were reconstructed using filtered-back projection (FBP) and ADMIRE levels 3 and 5 (A3-A5). HU accuracy was assessed calculating the root-mean-square deviation (RMSDHU). Image noise and CNR were computed on VMIs at 40/50/60/70 keV for 4 iodine inserts with 0.5, 1, 2 and 5 mg/mL concentrations. Accuracy of iodine quantification was assessed by the RMSDiodine and iodine bias (IB). Results: The RMSDHU decreased significantly as the dose levels increased compared to 5 mGy, except for 8 mGy with A3 (P = 0.380) and with A5 level (P = 0.945). Noise increased by 63.0 +/- 3.0 (standard deviation [SD])% from 20 mGy to 5 mGy. Noise decreased significantly by -53.8 +/- 0.9 (SD) % with A5 compared to FBP. The CNR decreased by -43.1 +/- 6.5 (SD)% from 20 mGy to 5 mGy. It increased using ADMIRE, and as the ADMIRE levels increased. The RMSDiodine and IB decreased as the dose level increased, and this was similar for all reconstruction types. Conclusion: ADMIRE strongly improves image quality in VMIs and slightly improves HU accuracy but does not affect the accuracy of iodine quantification. (c) 2021 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:31 / 40
页数:10
相关论文
共 50 条
  • [1] Optimization of image quality and accuracy of low iodine concentration quantification as function of kVp pairs for abdominal imaging using dual-source CT: A phantom study
    Dabli, Djamel
    Frandon, Julien
    Hamard, Aymeric
    Belaouni, Asmaa
    Addala, Takieddine
    Beregi, Jean-Paul
    Greffier, Joel
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 88 : 285 - 292
  • [2] Deep learning image reconstruction to improve accuracy of iodine quantification and image quality in dual-energy CT of the abdomen: a phantom and clinical study
    Fukutomi, Akiyo
    Sofue, Keitaro
    Ueshima, Eisuke
    Negi, Noriyuki
    Ueno, Yoshiko
    Tsujita, Yushi
    Yabe, Shinji
    Yamaguchi, Takeru
    Shimada, Ryuji
    Kusaka, Akiko
    Hori, Masatoshi
    Murakami, Takamichi
    EUROPEAN RADIOLOGY, 2023, 33 (02) : 1388 - 1399
  • [3] Deep learning image reconstruction to improve accuracy of iodine quantification and image quality in dual-energy CT of the abdomen: a phantom and clinical study
    Akiyo Fukutomi
    Keitaro Sofue
    Eisuke Ueshima
    Noriyuki Negi
    Yoshiko Ueno
    Yushi Tsujita
    Shinji Yabe
    Takeru Yamaguchi
    Ryuji Shimada
    Akiko Kusaka
    Masatoshi Hori
    Takamichi Murakami
    European Radiology, 2023, 33 : 1388 - 1399
  • [4] Abdominal image quality and dose reduction with energy-integrating or photon-counting detectors dual-source CT: A phantom study
    Greffiera, Joel
    Dabli, Djamel
    Faby, Sebastian
    Pastor, Maxime
    Croisille, Cedric
    de Oliveira, Fabien
    Erath, Julien
    Beregi, Jean Paul
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2024, 105 (10) : 379 - 385
  • [5] Dual-source photon counting detector CT with a tin filter: a phantom study on iodine quantification performance
    Tao, Ashley
    Huang, Richard
    Tao, Shengzhen
    Michalak, Gregory J.
    McCollough, Cynthia H.
    Leng, Shuai
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (11):
  • [6] Systematic radiation dose optimization of abdominal dual-energy CT on a second-generation dual-source CT scanner: assessment of the accuracy of iodine uptake measurement and image quality in an in vitro and in vivo investigations
    Schindera, Sebastian T.
    Zaehringer, Caroline
    D'Errico, Luigia
    Schwartz, Fides
    Kekelidze, Maka
    Szucs-Farkas, Zsolt
    Benz, Matthias R.
    ABDOMINAL RADIOLOGY, 2017, 42 (10) : 2562 - 2570
  • [7] Systematic radiation dose optimization of abdominal dual-energy CT on a second-generation dual-source CT scanner: assessment of the accuracy of iodine uptake measurement and image quality in an in vitro and in vivo investigations
    Sebastian T. Schindera
    Caroline Zaehringer
    Luigia D’Errico
    Fides Schwartz
    Maka Kekelidze
    Zsolt Szucs-Farkas
    Matthias R. Benz
    Abdominal Radiology, 2017, 42 : 2562 - 2570
  • [8] Optimization of SPECT-CT Hybrid Imaging Using Iterative Image Reconstruction for Low-Dose CT: A Phantom Study
    Grosser, Oliver S.
    Kupitz, Dennis
    Ruf, Juri
    Czuczwara, Damian
    Steffen, Ingo G.
    Furth, Christian
    Thormann, Markus
    Loewenthal, David
    Ricke, Jens
    Amthauer, Holger
    PLOS ONE, 2015, 10 (09):
  • [9] Dose reduction and image quality in CT examinations using an iterative reconstruction algorithm: a phantom study
    Guariglia, S.
    Meliado, G.
    Zivelonghi, E.
    Pinali, L.
    Montemezzi, S.
    Cavedon, C.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2015, 1 (04):
  • [10] Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose
    Park, Eun-Ah
    Lee, Whal
    Kim, Kwang Woo
    Kim, Kwang Gi
    Thomas, Allmendinger
    Chung, Jin Wook
    Park, Jae Hyung
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (07): : 1775 - 1786