Evaluation of SD Bioline HIV/syphilis Duo rapid test kits in Nepal

被引:9
|
作者
Shakya, Geeta [1 ]
Singh, Dipendra Raman [2 ]
Ojha, Hemanta Chandra [2 ]
Ojha, Chet Raj [1 ]
Mishra, Shravan Kumar [1 ]
Malla, Karishma [3 ]
Chaudhary, Puspa [3 ]
Regmi, Kiran [4 ]
机构
[1] Natl Publ Hlth Lab, Kathmandu, Nepal
[2] Natl Ctr AIDS & STD Control, Kathmandu, Nepal
[3] Paropakar Matern Hosp, Kathmandu, Nepal
[4] Family Hlth Div, Kathmandu, Nepal
来源
BMC INFECTIOUS DISEASES | 2016年 / 16卷
关键词
Duo test; Sensitivity; Specificity; Kappa coefficient; GENITAL ULCERS; VIRAL LOAD; SYPHILIS; COINFECTION; PREVALENCE; ASSAY;
D O I
10.1186/s12879-016-1694-9
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Background: Accurate and prompt diagnosis of HIV and syphilis simultaneously has reinforcing effect on their control program because of their prevalent co-infection. Availability of a simple user-friendly two-pronged and affordable detection tools brings down the cost of health care. They are important in the antenatal clinics, with added opportunity for intervention and prevention of mother to child transmission. In cooperation with rapid test kit manufacturers, SD Bioline, NPHL and NCASC, an evaluation of commercially available HIV/syphilis Duo rapid test kit (SD Bioline) to assess its performance and operational characteristics was done in the present study. Method: A prospective laboratory-based cross sectional study was conducted at a large Women's Hospital. Ten thousand pregnant women, visiting the Hospital for antenatal care or for delivery, were enrolled in study. Tests were performed by the SD Bioline HIV/Syphilis Duo kit as well as national algorithm for HIV and syphilis diagnosis which were considered gold standard. Sensitivity, Specificity, positive predictive value and negative predictive value along with kappa coefficient were calculated for the kit under evaluation. Result: The sensitivity, specificity, Negative predictive value and Positive predictive value of the kit for HIV diagnosis were 100 % (95 % CI 83.18-100 %, 99.96-100 %, 83.18-100 %, and 99.96-100 %, respectively). Kappa value was found to be 1.0. Out of total cases, results of 9985 (99.85 %) cases were concordant with National algorithm for syphilis diagnosis. Thirteen (0.13 %) cases were found false positive while two were false negative. The sensitivity of the kit for syphilis diagnosis was found to be 95.45 % (95 % CI 84.86-98.74 %) and specificity was 99.87 % (95 % CI; 99.78-99.92 %). Positive predictive value was 76.36 % (95 % CI; 63.65-85.63 %) and Negative predictive value was 99.89 % (95 % CI; 99.39-99.99 %). Kappa value was found to be 0.85. Conclusion: The performance characteristics of SD Bioline HIV/Syphilis duo kit were found almost concordant with the kits being used for HIV and Syphilis diagnosis separately. Its implementation in antenatal clinics/VCTs could be an added opportunity for simultaneous diagnosis of HIV and syphilis.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Evaluation of SD Bioline HIV/syphilis Duo rapid test kits in Nepal
    Geeta Shakya
    Dipendra Raman Singh
    Hemanta Chandra Ojha
    Chet Raj Ojha
    Shravan Kumar Mishra
    Karishma Malla
    Puspa Chaudhary
    Kiran Regmi
    BMC Infectious Diseases, 16
  • [2] EVALUATION OF A DOUBLE RAPID TEST FOR SYPHILIS AND HIV: SD BIOLINE HIV/SYPHILIS DUO
    Chiappe, M. A.
    Lopez-Torres, L.
    Carcamo, C.
    Garcia, P.
    Peeling, R.
    SEXUALLY TRANSMITTED INFECTIONS, 2013, 89 : A363 - A363
  • [3] An evaluation of the SD Bioline HIV/syphilis duo test
    Holden, Jeffrey
    Goheen, Joshua
    Jett-Goheen, Mary
    Barnes, Mathilda
    Hsieh, Yu-Hsiang
    Gaydos, Charlotte A.
    INTERNATIONAL JOURNAL OF STD & AIDS, 2018, 29 (01) : 57 - 62
  • [4] CLINICAL EVALUATION OF A DUAL RAPID DIAGNOSTIC TEST FOR HIV AND SYPHILIS, SD BIOLINE HIV/SYPHILIS DUO, LIMA, PERU 2013
    Bristow, Claire C.
    Leon, Segundo
    Cordova, Lourdes Ramos
    Rivera, Silver K. Vargas
    Flores, Juan A.
    Caceres, Carlos F.
    Klausner, Jeffrey
    SEXUALLY TRANSMITTED DISEASES, 2014, 41 : S85 - S85
  • [5] The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study
    Shimelis, Techalew
    Tadesse, Endale
    BMJ OPEN, 2015, 5 (04):
  • [6] MULTI-SITE LABORATORY EVALUATION OF THE SD BIOLINE DUO HIV/SYPHILIS DUAL POINT-OF-CARE RAPID TEST
    Bristow, Claire C.
    Adu-Sarkodie, Yaw
    Ondondo, Raphael
    Bukusi, Elizabeth Anne
    Dagnra, Anoumou
    Oo, Khin Yi
    Khamsay, Chanthavysouk
    Huong, Le Thi
    Campuzano, Roberto Vazquez
    Klausner, Jeffrey
    SEXUALLY TRANSMITTED DISEASES, 2014, 41 : S79 - S79
  • [7] Evaluation of the SD BIOLINE HIV/syphilis Duo assay at a rural health center in Southwestern Uganda
    Omoding D.
    Katawera V.
    Siedner M.
    Boum Y.
    II
    BMC Research Notes, 7 (1)
  • [8] Prospective evaluation of the SD BIOLINE Dengue Duo rapid test during a dengue virus epidemic
    C. Simonnet
    A. Okandze
    S. Matheus
    F. Djossou
    M. Nacher
    A. Mahamat
    European Journal of Clinical Microbiology & Infectious Diseases, 2017, 36 : 2441 - 2447
  • [9] Prospective evaluation of the SD BIOLINE Dengue Duo rapid test during a dengue virus epidemic
    Simonnet, C.
    Okandze, A.
    Matheus, S.
    Djossou, F.
    Nacher, M.
    Mahamat, A.
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2017, 36 (12) : 2441 - 2447
  • [10] Evaluation of the SD BIOLINE dengue duo rapid test in the course of acute and convalescent dengue infections in a Mexican endemic region
    Sanchez-Vargas, Luis A.
    Sanchez-Marce, Elvis E.
    Vivanco-Cid, Hector
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2014, 78 (04) : 368 - 372