Peak fronting in reversed-phase high-performance liquid chromatography:: a study of the chromatographic behavior of oxycodone hydrochloride

被引:20
|
作者
Broglé, K [1 ]
Ornaf, RM [1 ]
Wu, D [1 ]
Palermo, PJ [1 ]
机构
[1] Purdue Pharma LP, Purdue Res Ctr, Ardsley, NY 10502 USA
关键词
peak fronting; oxycodone hydrochloride; reversed-phase high-performance liquid chromatography;
D O I
10.1016/S0731-7085(98)00290-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Severe peak asymmetry -fronting- was observed for oxycodone during elution at 30 degrees C from a C-18 HPLC column using a mobile phase consisting of 14.9% MeOH, 84.5% 0.05 M KH2PO4 (pH 3.0), 0.5% MTBE, and 0.1% TEA. Investigation using deuterium-labeled oxycodone and analysis by LC/MS showed that gem-diol and hemiketal adducts of oxycodone formed as a result of the equilibrium addition of water and methanol to the C-6 ketone on oxycodone. As a result of slow equilibrium kinetics at room temperature in aqueous solution, the gem-diol and methyl hemiketal eluted as an unresolved broad band in front of the oxycodone peak. Decreasing the column temperature to 0 degrees C decreased the rates of interconversion and allowed the resolution and separation of these species from each other and from oxycodone. Increasing the column temperature to 60 degrees C increased the rates of interconversion with the result that the three species eluted as a single, homogenous peak with greatly improved peak symmetry. (C) 1999 Elsevier Science B.V. All rights reserved.